How to check your pH electrode slope 1) Select the millivolt (mV) mode of your pH meter 2) Using a pH electrode, obtain mV readings of two fresh calibration buffers (i.e., pH 4.0 and pH 7.0 are best) 3) Determine the net mV change 4) Determine the net mV change per pH unit change and compare using the chart at right >102 102% 101% 100% 99% 98% 97% 96% 95% 94% 93% 92% <92 1pH unit >59.3 59.3 58.8 58.2 57.6 57.0 56.4 55.8 55.3 57.0 54.1 53.5 <53.5 at 25°C 3pH units >178.0 178.0 176.3 174.5 172.8 171.0 169.3 167.5 165.8 171.0 162.3 160.5 <160.5 1pH unit 3pH units >60.3 60.3 59.8 59.2 58.6 58.0 57.4 56.8 56.2 58.0 55.0 54.4 <54.4 >181.0 181.0 179.3 177.5 175.7 173.9 172.2 170.4 168.6 173.9 165.1 163.3 <163.3 Status Poor OK Very Good Ideal Very Good Very Good Very Good Very Good Very Good OK OK OK Poor Care and Maintenance of Electrodes • Calibrate regularly with fresh buffers • Rinse electrode tip well between readings • On refillable electrodes be sure to check your fill level and open the fill hole before calibration or measurement • Never rub the bulb to dry it; a static charge can develop! • Store your electrode with fill-hole closed in electrode storage solution (Never store in deionized water!) • Special electrode cleaning solutions (13-642-583) help with tough samples that coat your electrode bulb—proteins, bacteria, oils, and more • Record temperature along with your pH value For example, at 25°C: pH 4 = 170.5mV, pH 7 = -3.4mV: Net change = 170.5mV (-3.4mV) = 173.9mV. Since pH 4 and pH 7 are 3 pH units apart, using the chart, 98% = Very Good electrochemistry 101 with Dr. Seymour Blue Need Help? Call 00 •Organic Oil/Grease Films—liquid dish soap and warm water are best. Isopropyl alcohol can be used on most electrodes if additional cleaning is needed. NOTE: Solvents such as acetone will damage plastic body electrodes! •Inorganic Salt Deposits—soak probe in 0.1M HCl as needed •Proteins—Use a protein-cleaning solution (13-642-584) If you have a refillable probe, drain the chamber and refill with correct reference filling solution. Rinse the probe tip with distilled or deionized water and condition the probe in storage solution before using again. w Te 0 7 l: 1 6 -800-76 w w m a . o f x: i c 6 shersci. 6 1-8 1 1 0 0- 9 2 6 for your accumet specialist! Buffers should be at the same temperature as your sample. pH Scale 0 0 1 Battery Battery Acid Acid 2 Lemon LemonJuice Juice 1 Increasing Acidity Neutral pH Values of Buffers at Various Temperatures F 1.67 1.67 1.67 1.67 1.68 1.68 1.69 1.71 1.72 1.74 1.77 1.79 4.00 4.00 4.00 4.00 4.01 4.02 4.04 4.06 4.09 4.13 4.16 4.21 6.98 6.95 6.92 6.87 6.86 6.85 6.84 6.83 6.84 6.85 6.86 6.88 7.11 7.08 7.06 7.01 7.00 6.98 6.97 6.97 6.97 6.99 7.03 7.08 9.46 9.40 9.33 9.23 9.18 9.14 9.07 9.01 8.96 8.92 8.89 8.85 10.32 10.25 10.18 10.06 10.01 9.97 9.89 9.83 9.79 9.78 9.78 9.80 3 3 Vinegar Vinegar Orange Juice 4 Increasing 4 Acidity 55 (per NIST) Standardized Buffers 0°C 5°C 10°C 20°C 25°C 30°C 40°C 50°C 60°C 70°C 80°C 90°C 2 Acid Rain Cleaning Probes Cleaning a dirty, clogged or coated electrode can restore proper electrode performance and prolong the useful life of the electrode. Here are some cleaning suggestions: 888-358-4706 Did you know? 12.79 12.73 12.67 12.52 12.46 12.36 12.17 11.96 11.73 11.47 11.19 10.89 Neutral 66 77 Milk Milk Distilled Water 88 Increasing 99 Alkalinity 10 10 Increasing Alkalinity 11 11 12 12 13 13 14 14 Baking Soda Soda Baking Sea Water Sea Water Milk of of Magnesia Magnesia Milk Ammonia Ammonia Lye Lye Acid Rain ©2009 Thermo Fisher Scientific Inc. All rights reserved. Litho in U.S.A DC/MA 7/09 09_2129 BN0720094 Slope at 20°C
© Copyright 2026 Paperzz