Product Data Sheet MAGNETIC FLOW/ COIL TEST PIECES Magnetic Particle Inspection Equipment General Description Operating Instructions Our magnetic flow/coil test pieces contain artificial defects that simulate discontinuities when verifying MPI system performance. For the first use of your test piece, follow the procedure outlined below when the unit has just been calibrated, and there is fresh ink in the tank: Part number 026T002 RRP58004, TP4 5 holes, 200 x 25 x 25 mm 1. Ensure that the test piece is thoroughly degreased and free from corrosion and oxidation 2. Clamp the test piece between the contacts on the head and tailstock of your bench unit for magnetic flow, or position within the aperture of the coil 3. Apply magnetic ink while the current is flowing Part number 026T041 BS6072 - 1 hole, 380 x 20 x 20 mm 4. Establish the current required to make the hole nearest the outer surface visible on the outer surface 5. Further increase the current to establish indications from the other holes on the outer surface of the test piece 6. Note the current required to establish each hole indication. If your magnetising system is functioning correctly, each hole should be visible at each subsequent use of the test piece. www.eu.magnaflux.com Faraday Road, South Dorcan Industrial Estate, Swindon, SN3 5HE, UK T + 44 (0)1793 524566 E [email protected] Bahnhofstr 94-98, 73457 Essingen, Germany T +49 (0) 7365 81-0 E [email protected] Lagergrens gata 2, 652 26 Karlstad, Sweden T +46 54 29 39 50 E [email protected] Version 17.1