10X6 Chambers - Technical Specifications ACCU

10X6 Chambers - Technical Specifications
for use with: Accu-Gold+ / Accu-Dose+
ACCU-GOLD+
ACCU-DOSE+
10X6-6
10X6-180
General Purpose In Beam Chamber
Leakage and Low Level Measurements Chamber
A well documented wide dynamic range chamber with many
dose and rate applications. Also recommended for measuring
exposure time in Auto Dose mode.
Designed for leakage measurements. Cross-section of 100
cm2 and volume of 180 cm3. Also for very low dose to image
receptor.
10X6-6M
10X6-1800
Dedicated Mammography Chamber
A world standard for mammography, ready for any
mammographic tube track-filter combination. Extraordinary flat
energy response over 10 - 40 keV has been documented in
technical papers and makes corrections unnecessary. Also
recommended for measuring exposure time in Auto Dose mode.
Radiation Protection Chamber
For low-level radiation measurements such as shielding
leakage, irradiator and environmental. Unlike typical survey
meters, the 1800 cm3 volume chamber offers improved
accuracy over a wider dynamic range.
10X6-0.18
10X6-10
General Purpose In Beam Chamber
This 10cc multi-purpose ion chamber is suitable for measuring
R&F as well fluoro. It’s small size (5cm diameter x 1cm thick),
excellent energy response and x-ray transparency makes it ideal
for those difficult measurements where space is at a minimum.
High Dose Rate Chamber
For in-beam measurements of high-intensity gamma radiation.
Gamma irradiators and beam type irradiators. The fully
guarded chamber is mounted at the end of a 3 m low noise
triax cable.
10X6-0.6
10X6-60/60E
Service and Image Intensifier Chamber
The dynamic range and thin profile is ideal for Input Dose at the
Image Intensifier, High dose rates encountered in Fluoroscopy
and Cine, spot film devices & other special procedures.
Additionally, the -60E (extended) chamber's increased sensitivity
at lower energies turns the chamber into a "Universal" detector,
covering mammography through R&F and beyond.
10X6-3CT
Computed Tomography Dose Index (CTDI)
and DWP or DLP Chamber
Although designed specifically for CT X-ray beam
measurements, either free-in-air or mounted in a head or body
phantom, it can be used for DWP and DLP applications such as
Dental x-ray measurements, due to the chambers excellent
energy and partial volume response as well as uniformity along
its entire 10 cm active length.
High Dose Rate Chamber
This high dose rate chamber provides an excellent response at
therapy and other high energy, high dose rate applications. The
fully guarded chamber is mounted at the end of a 12 m low
noise triax cable. Can be used with or without build up cap
depending upon the application.
10X6-0.6CT
Modern Wide Beam Multi-Slice
CT Chamber
0.6cc thimble chamber as described in the AAPM Report No.
111 “Comprehensive Methodology for the Evaluation of
Radiation Dose in X-ray Computed Tomography.” Ideal for
dose measurements in modern wide beam multi-slice CT.
Calibrated using X-rays @ 150 kVp, Phantom adapter
included.
426 West Duarte Road Monrovia, CA 91016 USA T: (626) 357-7921 F: (626) 357-8863 www.radcal.com
© 7/2016
10X6 Chambers - Technical Specifications
ACCU-GOLD+
for use with: Accu-Gold+ / Accu-Dose+
ACCU-DOSE+
SPECIFICATIONS / TECHNICAL DATA:
CHAMBERS
10x6-6
10x6-6M
10x6-101
10X6-60/60E
Min Rate
2 µR/s
20 nGy/s
2 µR/s
20 nGy/s
2 µR/s
20 nGy/s
Max Rate
17 R/s
149 mGy/s
10 R/s
88 mGy/s
17 R/s
149 mGy/s
2 R/s
Min Dose
10 µR
100 nGy
10 µR
100 nGy
10 µR
100 nGy
1 µR
59 kR
516 Gy
59 kR
516 Gy
59 kR
516 Gy
5.9 kR
Max Dose
Cine
Specifications
0.1 µR/f - >1 R/f
1 nGy/f - >10 mGy/f
Calibration
Accuracy
200 nR/s 2.0 nGy/s
10x6-3CT *
2 µR/s
20 nGy/s
19 mGy/s
40 R/s
10 nGy
20 µR
52 Gy
118 kR
350 mGy/s
0.6 R/s
4.9 mGy/s
200 nGy
200 nR
2 nGy
20 nR
200 pGy
1 kGy
2.0 kR
17 Gy
196 R
1.7 Gy
±4% using X-rays
±4% using X-rays
@ 60kVp and 2.8 mm @ 30kVp and 0.50
AL HVL
mm AL HVL
±5% using X-rays @
150 kVp & 10.2 mm
Al hvl
-60 ±4%
using
X-rays
@ 150kVp
and 10.2
mm AL
HVL
Exposure Rate
Dependance
±5%, 0.4 mR/s to 80 ±5%, 0.02 R/min to
R/s, up to 500 R/s for 600 R/min
50 us pulses
±5%, 0.4 mR/s to 80 ±5%, 2 mR/min to
R/s, up to 500 R/s for 199 R/min
50 us pulses
Energy
Dependance
±5%, 30 keV to 1.33
MeV (with build-up
material)
Active Length / Area
Construction
27.6 mm
Polycarbonate walls
and electrode
conductive graphite
interior coating; 6 cm3
active volume; 0.05kg
±5%, 10 keV to 40
keV
N/A
mg/cm2
0.7
metalized
polyester window;
polyacetal exterior;
6cm3 active volume;
0.08kg
±5%, 1.5 mm Al to
15 mm Al hvl
-60 ±5%
20 keV to
1.33 MeV
(with
build-up
material)
N/A
-60E ±4% 190
±4%mGy/s
using X-rays
@ 150kVp and 10.2
mm AL HVL
190
mGy/s
±2%,
2mR/s to 40
R/s
mGy/s
-60E ±5% 190
±5%,
3 to 20 mm
0.2 mm Al AL HVL
HVL to
1.33 MeV
(with
build-up
material)
N/A
N/A
Polycarbonate walls and
electrode. Conductive
graphite exterior
coating. 10.3 cm3 active
volume. 1.5m low noise
triax cable. 0.05kg
Polycarbonate walls;
conductive graphite
exterior coating; 60
cm3 active volume,
0.5 m low-noise triax
cable; 0.13kg
190 mGy/s
100 mm
190
mGy/s
C552
air-equivalent
walls and electrode:
polyacetal exterior cap;
3 cm3 active volume;
1.5m, low noise triax
cable; 0.11kg
Calibration Accuracy ± 4 %, Energy Dependence ± 5 %. Plug-and-play. * Uniformity Along Length & Partial Volume Exposure
1Specifications apply when used with Accu-series control unit.
** Complies with CFR 21
10x6-1800
1 nGy/s
0.01 µR/f ->100 mR/f 190 mGy/s
N/A
0.1 nGy/f ->1.0 mGy/f
using
X-rays
@ 50kVp
and 0.88
mm AL
HVL
10x6-180
100 nR/s
0.1 μR/f ->1 R/f
1 nGy/fr -> 10 mGy/f
N/A
All specifications subject to change.
5 nR/s
10x6-0.18
10X6-0.6/0.6CT
50 pGy/s
50 µR/s
500 nGy/s
20 µR/s
200 nGy/s
18 mR/s 0.2 mGy/s
720 R/s
6.31 Gy/s
133 R/s
1.17 Gy/s
200 µR
2 µGy
100 µR
1 µGy
2 MR
17 kGy
589 kR
5 kGy
N/A
N/A
±4% @ 60Co
N/A
0.6CT
±4% using
x-rays @
150 kVp
and
10.2mm Al
HVL
±4% using X-rays
@ 150kVp and 10.2
mm AL HVL
±4% using X-rays
@ 150kVp and 10.2
mm AL HVL
±5%, 20 mR/hr to
2000 R/hr
+0%, -5%, 0.1 mR/hr ±2%, 3 mR/s to
to 20R/hr, -10% to 65 720 R/s
R/hr
±5%, 30 keV to 1.33
MeV (with build up
material)
±5%, 33 keV to 1.33
MeV (with build up
material)
±5%, 45 keV to 1.33
MeV
100 cm 2 **
165 mm
8.1 mm
19.7 mm
Polycarbonate walls
and electrode;
conductive
graphite exterior
coating; 180 cm3
active volume; 0.11kg
Polycarbonate walls
and electrode;
conductive graphite
exterior coating;
1800 cm3 active
volume; 0.54 kg
C552 air-equivalent
walls and electrode;
polyacetal exterior
cap, 0.18 cm3 active
volume, 3m triax
cable
C552 air-equivalent
walls and electrode;
polyacetal exterior
cap, 0.6 cm3 active
volume, 0.6 12m triax
0.6CT 3m triax cable
0.6
±4% @
Co
60
±2%, 10 mR/s to
100 R/s
0.6
0.6CT
±5%, 40 ±5% 3 to
keV to
20 mm Al
1.33 MeV HVL
(with build
up cap)
±5%, to within 0.25 cm of chamber ends for a constant volume slice. Active length of 10 cm.
426 West Duarte Road Monrovia, CA 91016 USA T: (626) 357-7921 F: (626) 357-8863 www.radcal.com
© 7/2016