Nephrol Dial Transplant (2007) 22: 3553–3560 doi:10.1093/ndt/gfm436 Advance Access publication 23 August 2007 Original Article Resting energy expenditure and thermal balance during isothermic and thermoneutral haemodialysis—heat production does not explain increased body temperature during haemodialysis Jiřı́ Horáček1, Sylvie Dusilová Sulková2,3, Magdalena Fořtová4, František Lopot4, Marta Kalousová3, Luboš Sobotka2, Jiřı́ Chaloupka5, Vladimı́r Tesař6, Aleš Žák7 and Tomáš Zima3 1 Department of Internal Medicine II and 2Department of Gerontology and Metabolism, Charles University Prague, Faculty of Medicine and University Hospital Hradec Králové, Hradec Králové, 3Institute of Clinical Chemistry and Laboratory Diagnostics, 4Internal Department Strahov, Charles University Prague, 1st Faculty of Medicine and University Hospital Prague, Prague, 5Department of Occupational Health, Charles University Prague, Faculty of Medicine and University Hospital Hradec Králové, Hradec Králové, 6Department of Nephrology and 74th Department of Medicine, Charles University Prague, 1st Faculty of Medicine and University Hospital Prague, Prague, Czech Republic. Abstract Background. During routine haemodialysis (HD) body temperature increases, which contributes to haemodynamic instability. The relative roles of increased heat production and/or incomplete heat transfer are not fully elucidated. Concomitant measurement of heat production and heat transfer may help to assess the factors determining thermal balance during HD. Methods. Thirteen stable non-diabetic maintenance HD patients were investigated during two HD procedures (isothermic, dT ¼ 0, no change of body temperature; thermoneutral, dE ¼ 0, no energy transfer between blood and dialysate), using a blood temperature monitor (BTM) in active mode. Energy transfer, blood and dialysate temperature, and relative blood volume change (dBV) were continuously recorded, and resting energy expenditure (REE; Deltatrac Datex) was measured repeatedly during each procedure. Fourteen healthy persons served as controls for REE comparison. Results. In isothermic HD, median energy removal was 218 kJ/4 h HD (¼ heat flow 15.1 W). This cooling correlated with dBV induced by ultrafiltration (r ¼ 0.731, P < 0.01). There was no difference in dBV between isothermic (7.7%) and thermoneutral (8.1%) HD. Predialysis REE was 82.8 W/1.73 m2, not different from controls. No variation in REE during HD was observed, except a small and transient increase after a light meal (5 and 4%). In the time course of REE, no difference between the procedures was found. Correspondence and offprint requests to: Prof. Sylvie Dusilová Sulková, Department of Gerontology and Metabolism, Charles University Prague, Faculty of Medicine and University Hospital Hradec Králové, Email: [email protected] Conclusions. Our findings suggest that stable maintenance HD patients have REE not different from healthy controls, that HD procedure per se does not significantly increase REE and that neither isothermic nor thermoneutral regimen has any influence on metabolic rate. Therefore, body temperature elevation during routine HD may rather be due to decreased heat removal. With the use of BTM in active mode, body temperature can be kept stable (isothermic HD), which requires active cooling. This negative energy transfer is proportional to decrease in blood volume induced by ultrafiltration. Keywords: blood temperature monitor; isothermic haemodialysis; resting energy expenditure; thermal balance; thermoneutral haemodialysis; ultrafiltration Introduction There has been increasing interest in the thermoregulatory processes during haemodialysis (HD) in the last few years as thermal dysequilibrium is one of the major deviations from homeostasis during routine HD procedure, influencing greatly the haemodynamic stability in response to ultrafiltration [1–5]. Normally, the stability of body temperature is controlled by equilibrium between heat production and heat removal from the body. Heat elimination is mediated by perspiration, convection and radiation; they are capable of mutual compensation. None of these three components is directly measurable. On the other hand, heat production may be estimated using indirect calorimetry because it corresponds to energy expenditure. ß The Author [2007]. Published by Oxford University Press on behalf of ERA-EDTA. All rights reserved. For Permissions, please email: [email protected] 3554 HD procedure may affect the steady state between heat production and heat removal in several ways. First, there is a heat exchange between blood and dialysate in haemodialyser. Changes in heat dissipation from skin surface due to changes in skin perfusion may also be important. Finally, an increase in metabolic rate (i.e. increased heat production) during HD has been suggested. During routine HD, dialysis solution temperature is constant and no information about the real energy transfer is available. However, using a BTM device (Blood Temperature Monitor, Fresenius, Bad Homburg, Germany) the extracorporeal energy transfer during HD can be monitored, which offers a unique opportunity to quantify the heat loss. Moreover, the simultaneous registration of extracorporeal energy transfer together with the measurement of resting energy expenditure (REE) might provide more information about the thermoregulatory processes during HD. The BTM works in active or passive mode. During HD, two special sensors continuously measure the blood temperature both in arterial and venous extracorporeal lines and this process is highly standardized and precise. In passive mode, the energy transfer is only registered. In active mode, the dialysate temperature is continuously adjusted on the ‘biofeedback’ principle. This biofeedback allows to keep a pre-set body temperature (regimen T) or to reach a pre-set energy transfer (regimen E), respectively [1,3,5]. In this study, both energy production and energy transfer were studied during two distinct thermal regimens of HD (isothermic HD and thermoneutral HD). Isothermic HD (dT ¼ 0) keeps a stable body temperature. Thermoneutral HD (dE ¼ 0) means no thermal transfer in the dialyser. Our assumption was that if measured REE increased during both settings, then HD itself would be responsible for increased heat production and its accumulation during HD. If a change in REE occurred in one setting only, it would be related to the respective thermal balance, not to HD procedure itself (e.g. a change in REE in response to the change in body temperature). If REE remained stable in both thermal HD regimes, then thermal balance would be dependent on extracorporeal heat flow. We were also interested in possible relationship between thermal balance and relative blood volume changes (dBV) in response to ultrafiltration, as Rosales et al. [6] and Schneditz et al. [7] have previously described a close link between these variables. An accessory aim of the study was to compare the REE in our patients with a control group because of equivocal data on REE in dialysed patients in the literature. Subjects and methods Thirteen chronic HD patients (7 men and 6 women; aged 59.8 13.5 years, length of HD treatment 3.4 2.2 years) J. Horáček et al. Table 1. Baseline characteristics of haemodialysis patients (n ¼ 13) Variable Unit Body weight (BW) Body surface area Albumin Prealbumin Haematocrit kg m2 g/l g/l % Reference range Median IQR NA NA 32–45 0.20–0.36 37–47 (F) 42–52 (M) Urea mmol/l 2.0–6.7 (F) 2.8–8.0 (M) Ultrafiltration (UF) ml/HD NA UF as%BW % NA 78 1.96 36.0 0.30 32.9 72–87 1.82–2.03 34.5–38.0 0.25–0.41 29.6–34.4 23.0 20.2–28.4 2600 3.0 1550–2700 2.2–3.3 The patients were evaluated twice (before isothermic and thermoneutral HD) within 1 week and the average value of these two measurements was used for median and interquartile range (IQR) calculations. M, males; F, females. were studied prospectively (Table 1). The patients were clinically stable, with no signs of inflammation or heart failure. No patients with diabetes mellitus, or those treated with glucocorticoids, thyroid hormones or b-blockers were included. The control group of healthy subjects consisted of 4 men and 10 women with an average age of 41.3 20.5 years. All subjects were informed about the study procedure and its aim, and expressed their informed consent. Patients were observed during two HD procedures with different thermal balance, i.e. during isothermic (‘cool’) and thermoneutral (‘warm’) HD. Isothermic HD has been defined as dT ¼ 0, i.e. body temperature remained constant during HD because the excess heat produced during HD was convected out of the body (the energy balance of the procedure was negative). In contrast, thermoneutral HD has been defined as dE ¼ 0, i.e. no energy is lost in the extracorporal circuit and body temperature increases [1,2]. Both thermal regimens were continuously kept under control by BTM module (Fresenius, Bad Homburg, Germany) in active regimen, with HD monitor 4008S. Other HD characteristics remained identical in both regimens: low-flux steam sterilized polysulphone dialyzer 1.6 m2, two G15 dialysis needles, blood flow rate 300 ml/min, dialysate flow rate 500 ml/min, composition of dialysis solution: c(Naþ) 140 mmol/l, c(Kþ) 3 mmol/l, c(Ca2þ) 1.75 mmol/l and c(HCO 3 ) 32 mmol/l with no glucose, duration of HD 4 h, constant ultrafiltration rate and identical total amount of ultrafiltrate. The interval between measurements was one week (mid-week dialysis), afternoon shift. All patients had a well-functioning native arteriovenous fistula (proven by regular assessment of vascular access blood flow by thermodilution). REE was measured by indirect calorimetry (metabolic monitor Deltatrac Datex, Helsinki, Finland) using ventilated canopy. Patients were completely at rest during the whole HD procedure, lying supine in a quiet room with a temperature of 248C–258C. The measurement started at least 3 h after the last meal; however, no specific long-term food intake limitations were imposed to either control subjects or HD patients. The first measurement was performed before each HD, after 30 min of rest in supine position. It was then repeated 4 times: 10, 70, 110 and 215 min after the start of HD. The fourth measurement (110 min) followed shortly Resting energy expenditure and heat removal during haemodialysis 3555 after a light meal (two rolls and non-sweetened tea, 1000 kJ). Thus, REE was measured five times during each HD session. In each indirect calorimetry procedure, oxygen consumption (VO2) and CO2 production (VCO2) were registered for 10 min at 1 min intervals after 3 min of adaptation period. The average of these 10 measurements was taken as the result. Respiratory quotient (RQ ¼ VCO2/VO2) and resting energy expenditure (REE; Weir formula) were calculated by the device [8]. The measured REE values were compared with the values of basal metabolic rate (BMR) estimated from Harris–Benedict formula [9]. In HD patients, blood pressure (BP), heart rate (HR) and respiration rate (RR) were recorded before dialysis and at 30 min intervals during both procedures. Temperature in arterial blood line (Tart), venous blood line (Tven) and dialysate temperature (Tdial) as well as relative changes of intravascular volume (dBV; Critline; On-line Diagnostics, Riverdale, USA) [10] were recorded continuously throughout the study. Temperature in arterial blood line (Tart) measured by BTM is acceptable representative of mixed body temperature [1,5], provided the recirculation is measured and included into calculation, as it was in our practice. Pre-dialysis blood samples were drawn from inserted dialysis needle just before starting HD. During HD, samples were taken from arterial blood line in the 70th min of HD. Post-dialysis blood samples were taken using slow-flow method from arterial blood line. The samples were kept on ice for 30 min, and then centrifuged at 48C at 3000g. Thereafter, serum was kept at 208C until the laboratory assays were performed (max. 8 weeks). Serum urea and albumin concentrations were assayed using autoanalyser Kodak Ektachem 700XR and haematocrit levels using Critline device. For pre-albumin, immunoturbidimetric assay was used (Tinaquant-Prealbumin, Roche). To compare the magnitude of biological response to both procedures, pregnancy-associated plasma protein A (PAPP-A) was measured by TRACE (Time Resolved Amplified Cryptate Emission) technology based on nonradiating energy transfer. Commercial kit for PAPP-A determination (BRAHMS GmbH, Berlin, Germany) contains two different monoclonal antibodies—one is conjugated with europium cryptate and the other with fluorescent agent XL 665. The antigen (PAPP-A) present in serum samples is sandwiched between two conjugates. The fluorescent signal measured during the formation of the antigen–antibody complex by the KRYPTOR analyser (BRAHMS GmbH, Berlin, Germany) is proportional to the antigen concentration. Cyclic GMP (cGMP) as a marker of hydration was determined with competitive EIA using standard kits (Cayman Chemicals, USA). Concentration of non-esterified fatty acids (NEFA) was determined using the enzymatic-colorimetric method (NEFA, Randox Laboratories, UK); they were selected for their supportive information about metabolic response during HD. For statistical analysis, non-parametric tests were employed, as the assumptions for parametric tests were not consistently met. Data were expressed as medians (with 1st and 3rd quartiles). For comparisons between the control and HD groups, Mann–Whitney test was used. The two distinct thermodynamic regimens within the HD group were compared using Wilcoxon test for paired data. For variation of repeated measurements during HD, Friedman test was used, with subsequent Dunnett or Student–Newman–Keuls (SNK) test for significant difference between the individual measurements, as appropriate. Relations between variables were evaluated by non-parametric correlation analysis (Spearman r). SigmaStatÕ statistical software package, version 3.1 (Jandel Corp., San Rafael, USA), was used for calculations, and P < 0.05 was considered statistically significant. Results Thermal balance during thermoneutral and isothermic haemodialysis As expected, there was a significant difference in thermal balance between the HD procedures (Table 2). During thermoneutral HD, in accord with its definition, there was nearly no heat transfer between blood and dialysate, and continuous increase in arterial and venous line temperature was observed. During isothermic HD, dialysate temperature was continuously decreasing, which was paralleled by venous line temperature. The thermal balance of the procedure was clearly negative (218 kJ removed in 4 h; heat flow 15.1 W). In spite of this, there was a minor (though statistically significant) increase in arterial line temperature. While the difference in thermal balance and in venous line temperature was clearly apparent from the 30th min of HD, arterial line temperature (best reflecting core temperature) became significantly different only after the 210th min. Median difference in dialysate temperature till the end of both procedures was 1.98C with a starting dialysate temperature of 36.58C in both settings, and final values of 35.28C and 37.18C in isothermic and thermoneutral HD, respectively. Changes in intravascular volume (dBV) There was no difference between the changes of intravascular volume (dBV) (8.1 vs 7.7% in isothermic and thermoneutral HD, respectively, NS) (Table 2). In isothermic HD, significant correlation (r ¼ 0.731, P < 0.01) was found between dBV and total amount of heat removed (Figure 1). In thermoneutral HD, no clear-cut relationship between dBV and body temperature increase was found. Haemodynamic stability Both in isothermic and thermoneutral settings, blood pressure (systolic, diastolic and mean) remained stable and no symptomatic hypotension was observed. Also, there was no significant difference in the heart rate at any time point between the procedures, in spite of a minor (but significant) overall increase during thermoneutral HD (Table 3). 3556 J. Horáček et al. Table 2. Thermal parameters and relative decrease in blood volume in the course of isothermic (dT ¼ 0) and thermoneutral (dE ¼ 0) HD Variable Unit Regimen Haemodialysis Thermal kJ energy removal dE ¼ 0 8C dT ¼ 0 Tart dT ¼ 0 dE ¼ 0 Tven 8C dT ¼ 0 dE ¼ 0 Decrease % in blood volume dT ¼ 0 dE ¼ 0 Friedman P Start 30 min 60 min 90 min 120 min 150 min 180 min 210 min 240 min 0 0–0 0 0 36.7 36.6–36.8 36.8 36.4–37.0 36.2 36.0–36.6 36.4 36.3–36.7 0.0 0.0–0.0 0.0 0.0–0.0 14 5–19 4 1–7 36.8 36.6–36.9 36.9 36.5–37.2 36.6 36.3–36.7 37.0 36.6–37.4 1.7 0.9–1.9 1.9 1.0–2.1 31 5–41 3 1–5 36.9 36.7–37.0 36.9 36.6–37.2 36.2 35.7–36.7 36.9 36.5–37.2 1.9 1.2–4.4 2.0 1.7–3.7 57 4–75 3 0–5 36.9 36.7–37.0 36.8 36.6–37.3 36.0 35.6–37.2 36.8 36.6–37.3 3.2 2.3–3.8 3.8 2.9–4.6 81 3–110 3 2–4 36.9 36.7–37.0 37.1 36.6–37.3 35.8 35.5–36.3 37.1 36.5–37.3 4.4 3.1–7.0 5.1 3.9–7.9 115 17–147 2 1–5 36.9 36.7–37.1 37.1 36.7–37.5 36.0 35.3–36.7 37.2 36.8–37.4 4.3 2.3–7.2 5.1 3.8–6.5 143 60–194 3 1–6 36.9 36.8–37.2 37.3 36.9–37.5 35.7 35.1–36.5 37.1 36.9–37.4 5.3 2.2–8.0 5.8 4.3–8.5 173 89–238 3 1–5 36.9 36.8–37.2 37.4 36.9–37.5 35.9 35.3–36.4 37.4 36.8–37.5 5.4 2.8–9.0 6.6 5.6–8.0 218 89–284 2 0–5 36.8 36.7–37.1 37.3 36.9–37.5 35.6 35.3–36.2 37.2 36.7–37.6 8.1 3.8–10.9 7.7 6.5–10.1 <0.001 NS <0.001 <0.001 <0.05 <0.001 <0.001 <0.001 Decrease in blood volume is expressed as percentage of initial blood volume (measured by Critline device). Values are presented as median (above) with interquartile range (below). Variation of repeated measurements during HD was analysed by Friedman test (significance is given in the last column). If the variation was significant, Dunnett test (P < 0.05) was used to identify values significantly different from start. As expected from definitions of the regimens (dT ¼ 0 vs dE ¼ 0), thermal energy removal and the temperatures were clearly different between the procedures, reaching significance in 60 min (Wilcoxon test), except Tart where the difference became significant (P < 0.05) in 210 and 240 min only. Conversely, in blood volume changes there was no significant difference between the regimens at any time. Thermal energy removal, total amount of thermal energy transferred into extracorporeal circuit (measured by BTM module) within the given time; Tart, arterial blood line temperature; Tven, venous blood line temperature. Inflammatory marker PAPP-A increased after the start of dialysis, as expected, and returned to pre-treatment value. The sharp rise of NEFA in the 70th min was similar in both procedures, as was the subsequent decrease. No significant difference between isothermic and thermoneutral HD was found in any of these investigated parameters. Indirect calorimetry (REE and RQ) Fig. 1. Correlation of intravascular volume decrease (induced by ultrafiltration and measured by Critline device) and total thermal energy transferred into extracorporeal circuit (measured by BTM device) during 4 h of isothermic HD. Spearman (non-parametric) correlation: r ¼ 0.731; P < 0.01. Selected biochemical variables Serum urea concentrations were used for Kt/V calculations, and dialysis adequacy (eKt/V 1.3 or more) was confirmed. Blood concentrations of circulating volume marker, cGMP, metalloproteinase PAPP-A as marker of a complex biological response to extracorporeal procedure [11] and non-esterified fatty acids (NEFA) in the course of both procedures are summarized in Table 4. There was a significant decrease in cGMP level in both settings, and in the 70th min this decrease was relatively greater than that of intravascular volume. REE values remained stable in the course of both HD regimens (Table 5) except a small and transient (though significant) increase after the light meal (110 min). RQ was slowly but significantly declining during both HD regimens. However, this decline was abolished after the carbohydrate meal. Significant negative correlation (r ¼ 0.451, P < 0.05) was found between the change in RQ and in NEFA level during both HD settings. The values of pre-dialysis REE were generally stable within the same patient; median difference between the two measurements (i.e. before isothermic and thermoneutral HD) was 4.7% (4.1% when expressed as REE per 1.73 m2). Therefore, the mean of two pre-dialysis values in every patient was used for comparison with healthy controls. As summarized in Table 6, there was no difference in REE between HD patients and controls (in absolute values, or expressed in watts per standardized body surface area, or as percentage of BMR). Resting energy expenditure and heat removal during haemodialysis 3557 Table 3. Blood pressure (BP) and heart rate (HR) in the course of isothermic (dT ¼ 0) and thermoneutral (dE ¼ 0) HD Variable BP systolic Unit mmHg Regimen dT ¼ 0 dE ¼ 0 BP diastolic mmHg dT ¼ 0 dE ¼ 0 HR min1 dT ¼ 0 dE ¼ 0 Haemodialysis Friedman P Start 30 min 120 min 240 min 148 139–170 151 136–166 150 131–157 151 134–181 150 126–172 150 133–174 145 134–157 149 137–177 87 77–92 93 81–103 89 82–95 84 76–107 89 74–92 86 76–96 84 78–97 85 75–103 75 69–80 73 69–78 74 71–81 77 70–81 80 78–88 88 74–93 82 77–87 84 75–92 NS NS NS NS NS <0.001 Values are presented as median (above) with interquartile range (below). Variation of repeated measurements during HD was analysed by Friedman test (significance is given in the last column). If the variation was significant, Dunnett test (P < 0.05) was used to identify values significantly different from start. There were no statistically significant differences between the regimens (dT ¼ 0 vs dE ¼ 0) in any of the variables at any time of measurement (Wilcoxon test). Table 4. Selected biochemical variables in the course of isothermic (dT ¼ 0) and thermoneutral (dE ¼ 0) HD Variable CGMP Unit pmol/l Regimen dT ¼ 0 dE ¼ 0 PAPP-A mU/l dT ¼ 0 dE ¼ 0 NEFA mmol/l dT ¼ 0 dE ¼ 0 Haemodialysis before 70th min after 11.5 6.0–20.7 13.8 5.9–17.4 3.9 3.7–4.6 3.7 3.1–4.5 3.7 2.5–4.2 3.3 3.1–4.0 17.2 15.9–19.8 20.6 15.4–25.0 25.0 18.8–27.0 21.9 18.3–28.6 20.0 15.5–22.7 21.0 16.4–23.9 0.35 0.16–0.55 0.29 0.17–0.53 1.39 0.93–2.42 1.12 1.00–1.67 0.55 0.39–1.01 0.67 0.55–0.79 Friedman P SNK test P < 0.05 <0.05 1 vs 2,3 <0.01 1 vs 2,3 <0.01 2 vs 1,3 NS NS <0.001 1 vs 2 vs 3 <0.001 1 vs 2 vs 3 Values are corrected to blood volume changes and presented as median (above) with interquartile range (below). Variation of repeated measurements during HD was analysed by Friedman test. If the variation was significant, Student–Newman–Keuls (SNK) test was used to identify significant difference (P < 0.05) between the individual measurements (1 ¼ before HD, 2 ¼ in the 70th min, 3 ¼ after HD). There were no statistically significant differences between the regimens (dT ¼ 0 vs dE ¼ 0) in any of the variables at any time of measurement (Wilcoxon test). CGMP, cyclic guanosin monophosphate; PAPP-A, pregnancy-associated plasma protein A; NEFA, non-esterified fatty acids. Discussion In our study, heat production together with heat removal were being monitored in the course of two different thermal HD regimens. One of the compared regimens was isothermic HD that, with respect to thermal balance, makes HD as close as possible to physiological homeostasis because it prevents the increase in body temperature. The haemodynamic benefit of isothermic HD has been proven by Maggiore et al. [2] in a multicentre study and their conclusion has generally been accepted but still has remained underused [12]. In isothermic HD, the energy balance is negative. In two previous studies a quantitative relationship was found between ultrafiltration and energy transfer [6,7]. A similar conclusion can be drawn from our data: (cumulative) heat energy removal correlates well with (cumulative) decrease in intravascular volume (Figure 1). Furthermore, in our study ultrafiltration as well as blood volume decrease were considerably lower than in published studies, but this correlation was of 3558 J. Horáček et al. Table 5. Indirect calorimetry in the course of isothermic (dT ¼ 0) and thermoneutral (dE ¼ 0) haemodialysis Variable REE Unit W Regimen dT ¼ 0 dE ¼ 0 REE % baseline dT ¼ 0 dE ¼ 0 RQ 1 dT ¼ 0 dE ¼ 0 VCO2 ml/min dT ¼ 0 dE ¼ 0 VO2 ml/min dT ¼ 0 dE ¼ 0 Haemodialysis before 10 min 70 min 110 min 215 min 92.5 81.3–95.8 87.6 85.2–103.6 100 100–100 100 100–100 0.94 0.88–0.98 0.93 0.87–0.95 247 219–277 243 229–253 262 242–280 256 249–310 84.6 81.5–94.1 89.0 81.1–102.7 99 92–100 98 92–100 0.9 0.88–0.96 0.88 0.87–0.92 229 219–255 235 218–260 251 241–279 258 240–305 88.8 85.0–95.0 93.8 80.1–101.6 99 95–104 97 91–106 0.83 0.83–0.87 0.84 0.81–0.85 232 219–242 230 204–258 266 254–287 283 238–303 96.8 90.6–101.6 100.6 84.3–109.1 105 102–111 104 96–117 0.83 0.80–0.86 0.84 0.81–0.88 244 231–261 254 240–271 289 270–309 297 253–328 93.9 85.9–101.3 92.6 81.4–104.9 101 97–109 98 93–108 0.82 0.81–0.87 0.82 0.81–0.89 238 230–250 243 215–260 282 254–306 274 242–315 Friedman P SNK test P < 0.05 <0.01 4 vs 1,2,3,5 <0.01 4 vs 1,2,3,5 <0.01 4 vs 1,2,3,5 <0.01 4 vs 1,2,3,5 <0.001 1 vs 3,4,5 <0.01 1,2 vs 3,4,5 NS NS <0.01 3 vs 1,2,4,5 <0.01 4 vs 1,2,3,5 <0.05 2,3 vs 1,4,5 Values are presented as median (above) with interquartile range (below) Variation of repeated measurements during HD was analysed by Friedman test. If the variation was significant, Student–Newman–Keuls (SNK) test was used to identify significant difference (P < 0.05) between the individual measurements (1 ¼ before HD, 2 ¼ in the 10th min, 3 ¼ in the 70th min, 4 ¼ in the 110th min, i.e. after the light meal, 5 ¼ in the 215th min). There were no statistically significant differences between the regimens (dT ¼ 0 vs dE ¼ 0) in any of the variables at any time of measurement (Wilcoxon test). REE, resting energy expenditure; RQ, respiratory quotient (VCO2/VO2); VCO2, carbon dioxide production; VO2, oxygen consumption. Table 6. Indirect calorimetry—comparison of HD patients (n ¼ 13) and control group (n ¼ 14) Variable BMR REE REE REE RQ VCO2 VO2 Unit W W W/1.73 m2 % of BMR 1 ml/min ml/min Control HD Median IQR Median IQR 68.5 81.8 82.8 118 0.92 233 235 62.9–84.0 68.4–98.5 75.6–92.7 114–123 0.89–0.93 187–272 206–288 74.6 90.0 82.8 123 0.92 247 259 70.2–79.4 83.2–110.0 75.7–86.4 113–129 0.88–0.94 230–279 248–295 The patients in the HD group were measured twice (before isothermic and thermoneutral HD) within 1 week and the average value of these two measurements was used for median and interquartile range (IQR) calculations. There were no statistical differences between controls and HD patients in any of the variables (Mann–Whitney test). BMR, basal metabolic rate (calculated from Harris–Benedict formula); REE, resting energy expenditure (measured by indirect calorimetry); RQ, respiratory quotient (VCO2/VO2); VCO2, carbon dioxide production; VO2, oxygen consumption. similar strength. In our patients, the energy loss per 1% of body weight change corresponded to 5.6% of measured pre-dialysis REE, which is in perfect agreement with 6% estimation of Rosales et al. [6] despite considerably lower ultrafiltration volume in our patients. Also, our heat flow (15 W) is in accord with Schneditz et al. [7], who observed extracorporeal heat flow of 20 W (with a mean dBV of 15% in his patients). According to ‘volume hypothesis’ [1], the accumulation of heat in the body during conventional HD is a consequence of vasoconstriction, which is a regulatory haemodynamic response to decreasing intravascular volume due to ultrafiltration. Vasoconstriction leads to skin hypoperfusion, thus impairing heat dissipation. In order to prevent the rise in body temperature, the accumulated heat must be removed in another way, i.e. by heat transfer from blood into dialysate in dialyser. As vasoconstriction is presumably proportional to the loss of circulating volume, the necessary amount of heat removed is proportional to the loss of intravascular volume, too. Our findings are in accord with this ‘volume hypothesis’, and lend Resting energy expenditure and heat removal during haemodialysis 3559 further support to it in a novel setting, with smaller changes in intravascular volume, and no important changes in blood pressure. Quite recently, van der Sande et al. [13, 14] were first to show that in isothermic HD some thermal energy had to be removed (to keep body temperature constant) even in isovolaemic HD without any measured cutaneous perfusion changes. They concluded that ‘volume hypothesis’ does not provide full explanation for heat accumulation during HD, and more factors, so far poorly understood, may be involved. In our setting, ultrafiltration was always used, therefore we cannot comment on this. Twenty years ago Monteon et al. [15] demonstrated no important difference in energy expenditure between healthy subjects and stable patients undergoing maintenance HD. Their results were later supplemented with reports of increased REE precipitated by inflammation [16, 17] or accentuated hyperparathyroidism [18], but also of decreased expenditure in simple (non-inflammatory) malnutrition [19]; all these measurements were performed on non-dialysis days. Pre-dialysis REE in our patients did not vary on repeated measurement within 1 week, and it was not different from healthy controls (though our control group was not pair-matched). Our patients had no symptoms/signs of inflammation/infection or malnutrition. There are scarce data on time course of REE during HD and we have found no (full-text) report comparing different thermal HD regimens in this aspect. Ikizler et al. [20] described an increased REE in the course of HD, but the inflammatory status of patients was not reported, their dialysis solution contained 11 mmol/l glucose, and also in other aspects their method was different from ours. Also, studies comparing pre- and post-dialysis values of measured REE are scarce. An abstract by Lange et al. from 1995 was cited in [1] suggesting an average increase in REE by 8.6% (but with a high SD of 8.8%) during ‘cool’ HD, while during ‘warm’ HD it was 12.4% (SD 9.7%). These results are difficult to discuss as no full-text paper on these patients, with other important methodological data, has since appeared. Van der Sande et al. [13] found no change in REE before and after isothermic and thermoneutral HD (in both regimens with or without ultrafiltration). Our study gives further support to their measurements, providing original data on REE not only before and after, but also during the course of HD. However, this stability does not preclude mild changes in REE when more intensive cooling is applied, as observed by Rokyta et al. [21] in their critically ill patients. In our setting, HD was standard ‘low-flux’ procedure, using standard dialysis solution, where endotoxin concentrations were kept below the safety values recommended by European pharmacopoeia. There was no difference between isothermic and thermoneutral HD in serum PAPP-A levels, indicating that thermal balance does not influence the biological response to dialysis. Our secondary (though rather surprising) finding was continuously decreasing RQ in the course of both HD regimens. Ikizler et al. [20] described a similar trend, while van der Sande et al. [13] found no difference in RQ before vs after HD. Our finding may tentatively be explained by corresponding variations in NEFA concentrations and/or metabolism. Preferential NEFA oxidation decreases RQ, while glucose oxidation has the opposite effect. In the first hour of both HD regimens an increase in NEFA levels was found, followed by a decline but not to the baseline values. Theoretically, we might ascribe the initial rise of NEFA to heparin (2000 U administered at the beginning of HD), a well-known activator of lipoprotein lipase. More importantly, NEFA may also be mobilized by fasting during HD, possibly aggravated by glucose loss into dialysate. This hypothesis would be supported by their decline following a light (carbohydrate-containing) meal. While NEFA levels were decreasing after the light meal, the observed decline in RQ was stopped, and RQ then remained stable in spite of a temporary increase in oxygen consumption and REE. Also, there was a correlation between the changes in NEFA levels and in RQ during HD. We have found no data on NEFA dynamics during HD in the available literature. We are also not aware of any study describing the changes of vasoactive hormones during isothermic and thermoneutral HD. The sharp (and expected) decrease in cGMP, corresponding to the loss of circulating fluid volume, was not different between isothermic and thermoneutral HD, indicating no important difference in the fluid shifts between these procedures. In conclusion, our findings suggest that stablemaintenance HD patients have REE not different from healthy controls, that HD procedure per se does not significantly increase REE and that neither isothermic nor thermoneutral regimen has any influence on metabolic rate. Therefore, body temperature elevation during routine HD may rather be due to decreased heat removal. With the use of BTM in active mode, body temperature can be kept stable (isothermic HD), which requires active cooling. This negative energy transfer is proportional to decrease in blood volume induced by ultrafiltration. Acknowledgements. This study was supported by the research projects ‘MZO 0021620819’ and ‘MZO 00179906’, Czech Republic. We are grateful to Dr Jirina Soukupova for her valuable laboratory work, to Ing. Petr Moucka for his technical support and consultations and to Dr Jan Blaha for recruitment of patients. Conflict of interest statement. None declared. References 1. Lange H, Schwickardi M, Hermann W, Dahmen G, Issing J, Krämer M, Kuhlmann U. Effects of dialysate temperature on resting metabolic rate (RMR) during extracorporeal hemodialysis (HD) in ESRD [abstract]. Int J Artif Organs 1995; 18: 425 3560 2. Schneditz D. Temperature and thermal balance in hemodialysis. Semin Dial 2001; 14: 357–364 3. Maggiore Q, Pizzarelli F, Santoro A et al. The effects of control of thermal balance on vascular stability in hemodialysis patients: results of the European randomized clinical trial. Am J Kidney Dis 2002; 40: 280–290 4. Maggiore Q. Isothermic dialysis for hypotension-prone patients. Semin Dial 2002; 15: 187–190 5. Lopot F, Sulková S, Fořtová M. Temperature and thermal balance monitoring and control in dialysis. Hemodial Int 2003; 7: 177–183 6. Pergola PE, Habiba NM, Johnson JM. Body temperature regulation during hemodialysis in long-term patients: is it time to change dialysate temperature prescription? Am J Kidney Dis 2004; 44: 155–165 7. Rosales LM, Schneditz D, Morris AT, Rahmati S, Levin NW. Isothermic hemodialysis and ultrafiltration. Am J Kidney Dis 2000; 36: 353–361 8. Schneditz D, Rosales L, Kaufman A, Kaysen G, Levin NW. Heat accumulation with relative blood volume decrease. Am J Kidney Dis 2002; 40: 777–782 9. Takala J, Merilainen P. Handbook of Gas Exchange and Indirect Calorimetry. Datex Document 876710, Washington. 10. Harris JA, Benedict FG. Biometric Studies of Basal Metabolism in Man. Carnegie Institute of Washington 1919. Publication No. 279 11. Lopot F, Nejedlý B, Sulková S. Continuous blood volume monitoring and ultrafiltration control. Hemodial Int 2000; 4: 8–14 12. Kalousova M, Hodkova M, Dusilova-Sulkova S, Uhrova J, Tesar V, Zima T. Effect of hemodiafiltration on pregnancyassociated plasma protein A (PAPP-A) and related parameters. Renal Failure 2006; 28: 715–721 13. Selby NM, McIntyre CW. A systematic review of the clinical effects of reducing dialysate fluid temperature. Nephrol Dial Transplant 2006; 21: 1883–1898 J. Horáček et al. 14. van der Sande FM, Rosales LM, Brener Z et al. Effect of ultrafiltration on thermal variables, skin temperature, skin blood flow and energy expenditure during ultrapure hemodialysis. J Am Soc Nephrol 2005; 16: 1824–1831 15. van der Sande FM, Kooman J, Burema J et al. Effect of dialysate temperature on energy balance during hemodialysis: quantification of extracorporeal energy transfer. Am J Kidney Dis 1999; 33: 1115–1121 16. Monteon FJ, Laidlaw SA, Shaib JK, Kopple JD. Energy expenditure in patients with chronic renal failure. Kidney Int 1986; 30: 741–747 17. Neyra R, Chen KY, Sun M, Shyr Y, Hakim RM, Ikizler TA. Increased resting energy expenditure in patients with end-stage renal disease. J Parenter Enteral Nutr 2003; 27: 36–42 18. Utaka S, Avesani M, Draibe SA, Kamimura MA, Andreoni S, Cuppari L. Inflammation is associated with increased energy expenditure in patients with chronic kidney disease. Am J Clin Nutr 2005; 82: 801–805 19. Cuppari L, de Carvalho AB, Avesani M, Kamimura MA, Dos Santos LRR, Draibe SA. Increased resting energy expenditure in hemodialysis patients with severe hyperparathyroidism. J Am Soc Nephrol 2004; 15: 2933–2939 20. Kogirima M, Sakaguchi K, Nishino K, Ichikawa I, Hiramatsu F, Yamamoto S. Low resting energy expenditure in middle aged and elderly hemodialysis patients with poor nutritional status. J Med Invest 2006; 53: 34–41 21. Ikizler TA, Wingard RL, Sun M, Harvell J, Parker RA, Hakim R. Increased energy expenditure in hemodialysis patients. J Am Soc Nephrol 1996; 7: 2646–2653 22. Rokyta R, Matejovic M, Krouzecky A et al. Effects of continuous venovenous haemofiltration-induced cooling on global haemodynamics, splanchnic oxygen and energy balance in critically ill patients. Nephrol Dial Transplant 2004; 19: 623–630 Received for publication: 13.2.07 Accepted in revised form: 11.6.07
© Copyright 2026 Paperzz