Models - Midmark

Applies to Models:
93805106 - 1 1/2 Liter
93805107 - 5 1/4 Liter
93800038 - 18 1/2 Liter
20220601 - 45 Liter
Induction Chamber
Replacement
Chambers Only
93805108 - 1 1/2 Liter
93805109 - 5 1/4 Liter
93800132 - 18 1/2 Liter
20220602- 45 Liter
Set of Three
93805139
1 1/2, 5 1/4, 18 1/2 Liter
Hose Kit
92800131 - For all Models
To install self locking cover...
Slide cover under locking bar.
Pull Locking Bar Up Gently
So Rivets Will Slide Under
AA1170
45 Liter Shown
© Midmark Corporation 2008
Style A
1
10583800
Rev.A
To install hoses...
A) Insert 15mm male fitting into 15mm female port.
B) Insert FS-2 outlet adapter into vaporizer outlet port.
C) Attach scavenging hose to 19mm OD/15mm ID Fitting.
Note: Scavenging hose not supplied.
Outlet
Inlet
FS2 Outlet Adapter
Regulated
Oxygen
Supply
Hose Kit
Scavenging Hose
Sold Separately
15mm
Male Fitting
Fitting
(19mm O.D.)
15mm
Female
Port
VA1171
2
© Midmark Corporation 2008
Note
The graphs show the approximate time in minutes it takes to fill the chamber
with any desired % concentration of anesthetic drug for a set gas flow.
1 1/2 Liter & 5 1/4 Liter
18 1/2 Liter & 45 Liter
APPROXIMATE TIME TO FILL CHAMBER WITH %
ANESTHETIC AS SET ON VAPORIZER
4.5
4.5
4
4
3.5
3.5
3
2.5
1 1/2 Liter Chamber
2
5 1/4 Liter Chamber
1.5
1
FLOW RATE (LPM)
FLOW RATE (LPM)
APPROXIMATE TIME TO FILL CHAMBER WITH %
ANESTHETIC AS SET ON VAPORIZER
3
2.5
18 1/2 Liter Chamber
45 Liter Chamber
2
1.5
1
0.5
0.5
0
0
1
2
3
4
5
6
7
8
9
10
11
12
0
0
TIME (MINUTES)
10
15
20
25
30
35
40
TIME (MINUTES)
Example: 1 1/2 L Chamber
Vaporizer Setting - 4%
Flow Rate - 0.5 Liters Per Minute
Time to obtain 4% Concentration in 1 1/2 L Chamber
= 3 Minutes (Approximate)
Example: 45 L Chamber
Vaporizer Setting - 4%
Flow Rate - 3 Liters Per Minute
Time to obtain 4% Concentration in 45 L Chamber
= 15 Minutes (Approximate)
Midmark Corporation
For contact information, go to:
www.midmark.com
© Midmark Corporation 2008
5
3