FIELD SAFETY NOTICE KIMVENT* Closed Suction Systems EMEA5225 ATTENTION: Agnetha Helgesson MEDA AB TAMRO AB GOTEBORG IMPORTGATAN 20, PORT 2-4 HISINGS BACKA 0 Sweden Dear Valued Halyard Health Customer, What is the reason for this Voluntary Product Advisory Notice? Halyard Health (formerly known as Kimberly-Clark Health Care) has recently received reports indicating that the thumb valve of a very small percentage of KIMVENT* Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the “LOCKED” position. This situation has resulted in a partial air leak from the ventilator circuit triggering the ventilator alarm and replacement of the closed-suction device. Therefore, Halyard is notifying all customers who use this device to be aware of this potential issue and use the steps below to quickly correct the leak. There have been no reports of injuries associated with this condition to date. Which Products are impacted? This Product Advisory Notice applies to the KIMVENT* Closed Suction Systems identified in the list of potentially affected product codes and lot numbers (reference the ATTACHMENT). These are organized based on the main product categories. Please note that all potentially affected lots were shipped between 1st of April and 15th of July 2015. What is the purpose of this Product Advisory Notice? This notice is intended to inform all users of the KIMVENT* Closed Suction Systems of this potential valve leak and to provide instructions that immediately correct the leak. In the rare instance that a ventilator begins to alarm after attaching a KIMVENT* ClosedSuction System to the ventilator circuit or after suctioning, the cause might be associated with the thumb valve of the KIMVENT* System remaining partially open. To correct the leak without breaking the circuit, gently pull up on the thumb valve (i.e., in the Opposite direction than when depressing the thumb valve) and turn to the approximate 90 degree position as shown below (i.e., between the unlocked and locked positions). See figure. Halyard Belgium BVBA |Leonardo Da Vincilaan 1 - Building Aramis (4th floor) | 1930 Zaventem | Belgium T +32 (0)2 711 26 00 | F +32 (0)2 711 26 95 | [email protected] BTW/TVA/VAT BE 0553 802 001 | Citibank IBAN BE 03570133445584 | SWIFT/BIC CITIBEBX Ondernemingsnummer/Numéro d’enregistrement/Trade Registration: 0553.802.001 | RPR Brussel/RPM Bruxelles halyardhealth.nl | halyardhealth.fr | halyardhealth.de | halyardhealth.co.uk FIELD SAFETY NOTICE KIMVENT* Closed Suction Systems EMEA5225 If the leak continues and the ventilator alarm reactivates, replace the KIMVENT* Closed Suction System. Please report any occurrence of a valve leak to Halyard Health at the e-mail address [email protected]. Halyard Health is completing the investigation to prevent recurrence of this issue. If you require further assistance, please contact your Halyard Health Representative What should I do in response to this Product Advisory Notice? 1. Please review your distribution records to identify all customers who were shipped any of the impacted products. 2. Please notify each of your customers, who were shipped or who will be shipped any of the impacted products. 3. Please notify the competent authority on medical devices of your country, if applicable. Please note this notification is mandatory for countries of the European Union. 4. Please complete the attached Distributor Response Form and fax it at +32 (0)2 711 26 91 or e-mail to [email protected] within 5 business days of receipt of this notice. Thank you for your assistance and we apologize for any inconvenience this may cause. Sincerely, Olivier Zarza Quality and Regulatory Affairs Manager EMEA On behalf of Thomas Kozma, PhD Director, Regulatory Affairs Halyard Belgium BVBA |Leonardo Da Vincilaan 1 - Building Aramis (4th floor) | 1930 Zaventem | Belgium T +32 (0)2 711 26 00 | F +32 (0)2 711 26 95 | [email protected] BTW/TVA/VAT BE 0553 802 001 | Citibank IBAN BE 03570133445584 | SWIFT/BIC CITIBEBX Ondernemingsnummer/Numéro d’enregistrement/Trade Registration: 0553.802.001 | RPR Brussel/RPM Bruxelles halyardhealth.nl | halyardhealth.fr | halyardhealth.de | halyardhealth.co.uk FIELD SAFETY NOTICE KIMVENT* Closed Suction Systems EMEA5225 ATTACHMENT Lists of Potentially Affected KimVent* System Codes Catalog/product code Description Closed Suction System for Neonates / Pediatrics, 12 F, Y-Adapter 1912-5 195-5 197-5 198-5 Closed Suction System for Neonates / Pediatrics, 5 F, Y-Adapter Closed Suction System for Neonates / Pediatrics, 7 F, Y-Adapter Lot Code M5096T514 M5110T408 M5076T402 M5096T517 M5082T505 Closed Suction System for Neonates / Pediatrics, 8 F, Y-Adapter M5082T502 Closed Suction System for Adults, 12 F, T-Piece 215125-5 22103-5 M5103T601 Closed Suction System for Adults, 14 F, DSE M5085T604 22716-5 Turbo-Cleaning Closed Suction System for Adults, 12 F, DSE 240165-5 KimVent* Multi-Access Port Replacement Catheter for Adults, Endotracheal, 16 F M5089T307 Halyard Belgium BVBA |Leonardo Da Vincilaan 1 - Building Aramis (4th floor) | 1930 Zaventem | Belgium T +32 (0)2 711 26 00 | F +32 (0)2 711 26 95 | [email protected] BTW/TVA/VAT BE 0553 802 001 | Citibank IBAN BE 03570133445584 | SWIFT/BIC CITIBEBX Ondernemingsnummer/Numéro d’enregistrement/Trade Registration: 0553.802.001 | RPR Brussel/RPM Bruxelles halyardhealth.nl | halyardhealth.fr | halyardhealth.de | halyardhealth.co.uk M5089T807 FIELD SAFETY NOTICE KIMVENT* Closed Suction Systems EMEA5225 Distributor Response Form Please FAX a copy of this form to +32 (0)2 711 26 91, or send it via email at [email protected], within 5 business days of receipt of this notice. Facility: MEDA AB TAMRO AB GOTEBORG IMPORTGATAN 20, PORT 2-4 HISINGS BACKA 0 Sweden Our records indicate that you may have received certain impacted products identified in the chart below Catalog/product code 1912-5 Description Closed Suction System for Neonates / Pediatrics, 12 F, Y-Adapter 195-5 Closed Suction System for Neonates / Pediatrics, 5 F, Y-Adapter M5110T408 M5076T402 Closed Suction System for Neonates / Pediatrics, 7 F, Y-Adapter M5096T517 197-5 Lot Code M5096T514 M5082T505 198-5 Closed Suction System for Neonates / Pediatrics, 8 F, Y-Adapter M5082T502 Closed Suction System for Adults, 12 F, T-Piece 215125-5 22103-5 M5103T601 Closed Suction System for Adults, 14 F, DSE M5085T604 22716-5 Turbo-Cleaning Closed Suction System for Adults, 12 F, DSE 240165-5 KimVent* Multi-Access Port Replacement Catheter for Adults, Endotracheal, 16 F M5089T307 M5089T807 [ ] I certify that this facility has read and understood the information provided in the FSN [ ] I certify that this facility has notified the customers, who were shipped any of the impacted products and will notify the customers that will receive the impacted products I certify that the competent authority for medical devices of my country has been informed of this field safety notice: [ ] Yes [ ] N/A _____________________________________________ (Signature) (Date) _____________________________________________ (Name – print) _____________________________________________ (Title – print) Halyard Belgium BVBA |Leonardo Da Vincilaan 1 - Building Aramis (4th floor) | 1930 Zaventem | Belgium T +32 (0)2 711 26 00 | F +32 (0)2 711 26 95 | [email protected] BTW/TVA/VAT BE 0553 802 001 | Citibank IBAN BE 03570133445584 | SWIFT/BIC CITIBEBX Ondernemingsnummer/Numéro d’enregistrement/Trade Registration: 0553.802.001 | RPR Brussel/RPM Bruxelles halyardhealth.nl | halyardhealth.fr | halyardhealth.de | halyardhealth.co.uk FIELD SAFETY NOTICE KIMVENT* Closed Suction Systems EMEA5225 Halyard Belgium BVBA |Leonardo Da Vincilaan 1 - Building Aramis (4th floor) | 1930 Zaventem | Belgium T +32 (0)2 711 26 00 | F +32 (0)2 711 26 95 | [email protected] BTW/TVA/VAT BE 0553 802 001 | Citibank IBAN BE 03570133445584 | SWIFT/BIC CITIBEBX Ondernemingsnummer/Numéro d’enregistrement/Trade Registration: 0553.802.001 | RPR Brussel/RPM Bruxelles halyardhealth.nl | halyardhealth.fr | halyardhealth.de | halyardhealth.co.uk
© Copyright 2025 Paperzz