M-Series LabScope

Power
The circuit board, in the base of the unit, contains all the electrical functions. There are no user
repairable parts on the circuit board. If you suspect a faulty board, call LW Scientifics’ technical
service department.
M-Series LabScope
INPUT: AC90-240/50-6- HZ
OUTPUT: 12V/20W to a quartz halogen lamp (LW Scientific replacement part number: R-12V20W)
FUSE: One 2-amp fuse protects the circuit board from electrical overload. The fuse is located in the snap-out fuse panel
next to the on/off switch. There is also a spare fuse located inside the same panel. When replacing the fuse, always
install a new one of the same size and amperage.
Maintenance
Instruction Manual
1 Always cover your microscope with the dust cover when not in use.
Model pictured:
2 When cleaning the lenses, use lens paper or a cotton swab dipped in lens cleaning
M-Series LabScope
solution.
Not all features available on all models see back page for model specifications.
3 Excess oil should be cleaned off your 100x objective and stage at once. An alcohol pad is best for removing oil
from the stage and on the other metal parts, but is not recommended for use on the lenses. Use lens cleaning
solution and lens paper to clean off your objectives.
4 Dust in the nosepiece or ocular tubes should be blown out using filtered air. Canned air dusters work well for this job.
A
5 Whenever you remove an objective, we recommend that you place it back into the original plastic shipping vial until
ready to be placed back on the microscope.
C
6 To keep your microscope in top condition for years, LW Scientific recommends that you have the microscope
B
professionally serviced once a year.
Warning: The 40x objective is not sealed for oil immersion. Damage to the 40x objective due to oil
Specifications
Nosepiece
Reverse nosepiece
Quadruple hole multiple ball bearing (M2)
Quintuple hole multiple ball bearing (M5)
Head
Binocular (Seidentopf), Monocular or Trinocular
Inclined 30°
Rotates 360°
Diopter adjustment (Binocular or trinocular versions)
10X/20 and high-point eyepieces
Illumination
Moveable Abbe Condenser (NA 1.25)
Koehler Illumination standard
Iris Diaphragm
Filter holder in base with blue, green and yellow filters
Quartz halogen light (12v 20w JC-G4 bulb)
Input 90-240V auto-switching / 50-60Hz
35watt bulb upgrade available
Construction
Rugged alloy; acid and reagent resistant finish
Stage
Mechanical stage (160mm x 145mm)
– coaxial drive controls
Range of traverse is 77mm x 50mm
Objectives
Semi-Plan, Din Plan or Infinity
4X NA .10
10X NA .25
20X NA .40 (M5 only)
40XR NA .65
100XR NA 1.25 (oil immersion)
20X, 50X oil plan, and 60XR plan available
D E
P
H
I
J
K
L
M
N
O
P
F
O
N
I
H
G
J
M
Adjustment Controls
Eyepiece: Interpupillary distance adjustment 50-75mm
Stage Controls: Lateral and forward/reverse positioning;
Etched vernier scales
Coarse Adjustment: Range of 15mm
Fine Adjustment: Graduation of 2µm
Variable Light Adjustment: 0-20 watt output
A
B
C
D
E
F
G
K
Binocular Head
Eyepieces
Diopter Adjustment
Objectives
Slide Holder
Stage
X/Y Axis Controls
(located on left side)
Abbe Condenser
Iris Diaphragm
Base Condenser
On/Off Switch
Brightness Control
Fine Focus
Coarse Focus
Stage Stop
Friction Adjustment
L
Dimensions and Weight
Weight: 19.5 lbs/9kg
Height: 16.5"/42cm
Length: 10.6"/24cm
Width: 7.8"/16.5cm
Shipping weight: 24lbs/11kg
Introduction
The M-Series LabScope is our premium compound microscope, able to perform varied applications such as phase
contrast, darkfield and polarization, with semi-plan, plan and infinity optics (optional configurations). Equipped with
advanced optics and super-wide field, high-point eyepieces, the M-Series LabScope is the natural choice of researchers,
pathologists and professionals of every scientific discipline.
Objectives: The following numbers are
based on use with 10x/20 eyepieces.
50x Oil
Hard Carry Case
Polarization
Speed up differential
counting.
Custom-built for the
M-Series to protect
your scope.
Essential rheumatology
tool.
Epi Fluorescent
Phase Turret
50w Koehler
Semen Evaluation
100 watt
HBO light.
Phase with the
flip of a thumb.
You’re really going to
love the new Kohler!
Heated stage
with phase.
Size
4X
10X
20X
40XR
50XR
60XR
100XR
N.A.
0.10
0.25
0.40
0.65
0.95
0.85
1.25
Mag. Field of View
40X
5mm
100X
2mm
200X
1mm
400X
0.5mm
500X
0.4mm
600X
0.33mm
1000X
0.2mm
MKT-7.5.3-L-089 | Rev 0
P. 770.270.1394
F. 770.270.2389
nded
me
ecom
R
:
U
es
d
a
r
pg
Ask your
authorized LWS
dealer about
additional
accessories
865 Marathon Parkway Lawrenceville GA 30046
LW Scientific packs each M2/M5 Series LabScope with utmost care. Examine
Unpacking and Setup
the outer and inner containers for any visual damage. Retain all of the
packing material until you have examined and tested your new microscope.
If there is damage, please contact the shipping company, as our warranty
does not cover shipping damage. If you are uncertain who the shipper is, contact the distributor where you purchased
the microscope. Please retain all packaging material for future use. Carefully unpack your M2/M5 LabScope using the
following checklist for all the parts and accessories:
1 - Microscope body with Abbe condenser
2 - 10x eyepieces
1 - Binocular head (Seidentopf style)
3 - Filters (blue, yellow & green)
4 - Objectives (M2) all Din 4X, 10XR, 40XR, 100XR (oil)
5 - Objectives (M5) all Din 4X, 10XR, 20XR, 40XR, 100XR (oil)
2 - 2 amp fuses (both located in fuse-holder
near the on/off switch
1 - Operations Manual and Warranty Card
1 - Immersion oil
1 - Dust cover
Operation
Note: If your microscope has been exposed to cold weather, please allow time for all the parts
to come to room temperature before switching on the power. Excess cold can fog the lenses
and may cause the lamp to fail.
1 Once you have assembled all the parts, plug the power cord into the appropriate AC outlet.
2 Turn the light on using the black rocker switch on the back of the base. Next adjust the light intensity using the
adjustment knob located on the left side of the scope. Note: Rapid, repeated changes in light intensity will
dramatically shorten the life of the quartz halogen lamp.
3
In order to become familiar with the controls, choose a specimen slide with which you are familiar. For example, an
old hematology slide or a commercially prepared slide. Place the slide into the slide holder by pushing back on the
thumb guard to open the slide finger. The slide finger closes slowly to eliminate the possibility of chipping the corner of
your slide when it closes.
Move the slide to the center of the stage, by turning the stage and slide control knobs, just below the stage. These
4 knobs allow you to move the slide in the X-Y axis (forward/backward and left/right).
Assembly
5
1 The M2/M5 LabScope is packaged in two sections. The body of the microscope (including the objectives, filters, oil,
fuse and bulb) is in one large box and the head (including eyepieces and eye guards) is packaged in the smaller
box.
2
3
Remove the body of the microscope and place it on a sturdy, dust-free surface. Install the objectives in such a way
that when you turn the nosepiece clockwise, you are moving from the 4x, to 10x (to 20x-M5 only) to 40x, and finally
the 100x objective.
6 Once you are comfortably seated, look into the oculars and move the eyepiece tubes together or apart until you see
only one complete circle of light. You have now adjusted your interpupillary distance.
Using the 4x or 10x objectives and the coarse and fine adjustment knobs bring the field of view into focus. Now move
7 the 40x objective into place. You will feel a “clicking” action when the objective is seated properly. Again, focus the
best image. You will now be in the middle of the focus range. You may have to adjust the iris diaphragm (on the
condenser) for the best contrast.
Remove the microscope head from the Styrofoam carton and pull off the protective covers from the eyepiece tubes
4 and head mount. Insert the head mount into the upper arm of the body. Using the thumb screw, secure the head in
place. Note: Do not over-tighten.
1
8
Insert the 10x eyepieces. Attach the power cord.
Instructions for installing the optional trinocular tube:
2 Remove the metal cap on the top of the microscope head by inserting the Allen wrench (provided) into the screw on
opening.
tighten the tension control ring (located on the right side of the microscope between the coarse adjustment and the
body of the microscope). If the coarse focus is hard to turn, you may choose to loosen the friction adjustment.
Stage Stop: To help prevent the stage from hitting the objectives, this scope is equipped with an adjustable stage
10 stop. Rotate the 100X oil objective into place and put a slide into the slide holder. Slowly raise the stage, stopping
Tighten the screw on the side of the tube with the Allen wrench.
11
Lamp Replacement
Caution: Unplug the microscope and let it cool down if recently in use before
opening the base of the microscope. When handling a new quartz lamp, be sure
not to touch the new lamp with your fingers. Oil and sweat on the bulb may cause
it to explode or shatter as it heats up.
Turn the scope on its side. (We recommend that you place a cushion or soft cloth under the microscope to protect
against damage.) Locate the lamp compartment door; you will notice a black plastic knob near the front. Turn the knob
to open the door. The lamp compartment door may now be swung open to reveal the lamp plugged into its base.
Gently pull the old bulb out and discard.
Using a lint-free tissue or piece of cloth to hold the new lamp, install the bulb. Swing the lamp compartment door closed
and turn the black knob to lock. Carefully return the scope to the upright position and plug the microscope in the
appropriate AC outlet. The electronics on the M-Series LabScope will support up to a 35watt/ 12V bulb.
Diopter Adjustment: Since you are using a binocular microscope, you have to adjust for the normal difference in vision
between your two eyes. This is a simple but critical adjustment! Close your left eye and look into the right ocular with
your right eye. Adjust the fine focus to give you the best image. Now look at the ocular tube on the left. You will see
that the left ocular tube has a built-in adjustment ring. Now close your right eye and look with your left eye into the left
ocular. Using the diopter adjustment ring on the left ocular tube, adjust the focus until you see a clear, focused field.
9 Friction Adjustment: With repeated use and wear, the stage may drift out of focus. If this happens, you need only to
the side of the cap. Loosen the screw and remove the cap.
3 Insert the trinocular tube (eyepiece or c-mount, depending on how you ordered your microscope) into the
Open the iris diaphragm on the Abbe condenser (controlled by the black ring on the condenser). Insert the filter of
your choice on top of the condenser lens on the base. Note that many customers prefer to use the blue filter for
routine use.
when the slide makes contact with the objective. Now, turn the stage stop lever in a clockwise direction toward you.
The stage stop lever is located on the left side of the microscope between the coarse adjustment and the body of
the microscope.
Parfocality: All LW Scientific microscopes are manufactured to be parfocal – meaning that when you change
objectives or magnification, the specimen will remain very close to being in focus, with only a fine adjustment
needed. The M-Series LabScope is “preset” for parfocality.
Your M-Series scope has an iris in the sub-stage condenser, and another iris in the base Koehler condenser (if
12 equipped). Set the base iris to the widest open setting. The sub-stage iris should then be set, using the scale on the
side of the sub-stage condenser, to match the aperture of the objective for maximum resolution under each objective
power. For example, when using the 4x objective, the sub-stage iris should be set to 0.1, which is nearly closed. For the
10x objective, set the iris to .25. For the 40x objective, set the iris to .65. Under the 100x objective, the iris should be set
to 1.25, which is wide open.
You are now ready to use your M-Series LabScope – thank you for choosing LW Scientific!
P. 770.270.1394
F. 770.270.2389
865 Marathon Parkway Lawrenceville GA 30046