541 Continuous Cl2 Monitor - Categories On Nova Analytical Systems

541 SERIES
CHLORINE ALARM MONITOR
APPLICATIONS
Designed for the continuous monitoring of chlorine (Cl2) in
ambient air at water treatment plants, chlorine production
plants, commercial swimming pools, or any area in which
chlorine is handled or stored.
FEATURES
541D - Diffusion Style in
NEMA 12 Enclosure
(models shown with
optional strobe light)
• Stand-alone or remote sensor versions available
• Long-life electrochemical cell
• Fast response to the presence of chlorine in the air
• Instant warm-up, no wait
• Bright digital readout - no reading interpretation errors
• Easy use and calibrate
• Easy to maintain modular layout
541P - Pump Style in
NEMA 12 Enclosure
• 4-20mA output signal for recording
• NEMA 12 dust-tight enclosure
OPTIONS
• Alarm with LED indicator, relay contact, and
'Sonalert' audible alarm with 'Acknowledge' button
• Up to four remote sensors
• Strobe light for alarm indication
• NEMA 4 or 4X weather-proof enclosures
CALIBRATION
541DR1 - Alarm Unit with single remote
diffusion sensor
• On chlorine-free air or nitrogen for Zero
• On analyzed calibration gas of Cl2 in air for Span
NOVA ANALYTICAL SYSTEMS
www.nova-gas.com
www.nova-gas.com
DESCRIPTION
The Nova 541 Series chlorine alarm monitors are designed for the continuous detection of Cl2 in ambient air.
The Cl2 sensor is capable of excursions over 100 PPM with no damage. The Cl2 sensor is a disposable
electrochemical type with an expected life of between 3-4 years. The sensor needs no warm-up time and is
ready to go as soon as the monitor is turned on. Several versions are available as shown below.
MODELS
•
•
•
•
•
541D - one diffusion-type sensor in main cabinet
541DR1 - one diffusion sensor remote from main cabinet
541DR2 - two remote diffusion sensors
541DR3 - three remote diffusion sensors
541DR4 - four remote diffusion sensors
• 541P - built-in sample pump to draw in sample from
another location
Nova reserves the right to specification changes which
may occur with advances in design without prior notice.
SPECIFICATIONS
Description
Method of Detection:
Disposable electrochemical cell (3-4 years)
Ranges Available:
0-20 PPM standard (other ranges may be available)
Resolution:
1 PPM
Accuracy and Repeatability:
Better than 1% of full scale or 1 PPM, which ever is higher
Drift:
Less than 1% of full scale or 2 PPM per week, whichever is greater
Response Time (T-90):
Less than 15 seconds to 90% step change
Ambient Temperature Range:
32° to 120°F (0° to 40°C), may be operated at lower temperature with auxiliary heater
Linearity:
± 1% of full scale or 1 PPM, whichever is higher
Size and Weight:
14" H x 12" W x 8" D (35.6 x 30.5 x 20.3 cm) @ 18lbs. (8.2kg)
Power:
115VAC 60Hz (220VAC 50Hz available)
Output Options:
0-1VDC or 4-20mA
Alarms:
Optional High and/or Low alarm with SPDT contacts with 10A 115VAC rating, audible
alarm with reset button, and indicator light for each alarm level; optional Low Flow
alarm and strobe light also available
UNIQUE APPLICATIONS
All Nova analyzers are built using proven technologies and techniques. If this product does not suit your
application, please contact Nova at 1-800-295-3771. In many cases, we are able to build an analyzer
specific to your needs.
NOVA ANALYTICAL SYSTEMS
A UNIT OF TENOVA GOODFELLOW INC.
IN USA:
1925 Pine Avenue • Niagara Falls, NY • 14301
Tel: 1-800-295-3771 • 716.285.0418 • Fax: 716.282.2937
IN CANADA:
270 Sherman Avenue North • Hamilton, ON • L8L 6N5
Tel: 905.545.2003 • Fax: 905.545.4248
email: [email protected]
[email protected]
www.nova-gas.com