Working Report 2014-28 Sampling and Characterisation of Groundwater Colloids in ONKALO at Olkiluoto, Finland, 2013 Sari Luste Mikael Takala Pentti Manninen Ramboll Finland Oy June 2014 Working Reports contain information on work in progress or pending completion. SAMPLING AND CHARACTERISATION OF GROUNDWATER COLLOIDS IN ONKALO AT OLKILUOTO, FINLAND, 2013 ABSTRACT The purpose of this study was to estimate the concentration of groundwater colloids and to compare the results with the previous ones. The water samples were collected from groundwater stations ONK-PVA1, ONK-PVA5 and ONK-PVA10 in August 2013. The colloid concentrations were determined from scanning electron microscopy (SEM) micrographs taken from the filters in which the groundwater was run through. The particle calculations were performed with computer software ImageJ 1.47. In addition, field flow fractionation (FFF) measurements were performed from the unfiltered water samples. The colloid concentration (Ø 0 – 1 µm) determined by the single particle analysis of SEM micrographs in ONK-PVA1 was 300 µg/l while the colloid concentration in ONK-PVA5 was 30 µg/l and ONK-PVA10 40 µg/l. FFF measurements supported the results of single particle analyses with a difference that an extra peak was found from ONK-PVA1 sample. The peak, which showed no evidence in single particle analyses, was suspected to contain humic substances. Keywords: colloids, ONKALO, single particle analysis. OLKILUODON ONKALON POHJAVEDEN KOLLOIDIEN NÄYTTEENOTTO JA KARAKTERISOINTI 2013 TIIVISTELMÄ Työn tarkoitus oli arvioida pohjavedessä esiintyvien kolloidien määrää ja verrata tuloksia aiempiin tuloksiin. Näytteet haettiin pohjavesiasemilta ONK-PVA1, ONKPVA5 ja ONK-PVA10 elokuussa 2013. Kolloidien pitoisuus määritettiin pyyhkäisyelektronimikroskoopilla otetuista mikrografeista suodattimilta, joiden läpi pohjavettä laskettiin. Partikkelilaskenta suoritettiin ohjelmaa ImageJ 1.47 apuna käyttäen. Sen lisäksi käytettiin kenttävirtausfraktiointimenetelmää (FFF) suodattamattomille vesinäytteille. Kolloidien konsentraatio (halkaisija 0 – 1 µm) SEM kuvista tehtyjen laskentojen perusteella pisteen ONK-PVA1 vedessä oli 300 µg/l. Konsentraatiot pisteissä ONKPVA5 ja ONK-PVA10 olivat 30 µg/l ja 40 µg/l. FFF menetelmä tuki tuloksia sillä erolla, että FFF tuloksissa havaittiin näytteen ONK-PVA1 osalta ylimääräinen piikki. Kyseiselle piikille ei SEM kuvista tehdyissä laskennoissa ollut vastaavaa havaintoa ja sen arveltiin esittävän humusyhdisteitä. Avainsanat: kolloidit, ONKALO, single particle analysis. LIST OF SYMBOLS AND ABBREVIATIONS DLS Dynamic light scattering FFF Field-flow fractionation HDPE High density polyethylene LC Liquid chromatography PSS Polystyrene sulphonate PTFE Polytetrafluoroethylene RI Refractive index RPD Relative percentage difference RSD Relative standard deviation SEM Scanning electron microscopy SLS Static light scattering UV Ultraviolet 1 TABLE OF CONTENTS ABSTRACT TIIVISTELMÄ LIST OF SYMBOLS AND ABBREVIATIONS 1 INTRODUCTION .................................................................................................... 2 2 EXPERIMENTAL .................................................................................................... 3 2.1 Preparations .................................................................................................... 3 2.2 Sampling ......................................................................................................... 3 2.3 Single particle analysis ................................................................................... 5 2.4 FFF measurements ......................................................................................... 7 3 RESULTS AND DISCUSSION ............................................................................... 8 3.1 Single particle analysis ................................................................................... 8 3.1.1 ONK-PVA1 results ...................................................................................... 8 3.1.2 ONK-PVA5 results ...................................................................................... 9 3.1.3 ONK-PVA10 results .................................................................................. 10 3.1.4 Discussion ................................................................................................. 11 3.1.5 Reliability evaluation ................................................................................. 11 3.1.6 Other observations .................................................................................... 14 3.1.7 Comparison with the previous results ....................................................... 14 3.2 FFF measurements ....................................................................................... 15 4 RECOMMENDATIONS......................................................................................... 18 5 CONCLUSIONS.................................................................................................... 19 REFERENCES ............................................................................................................. 20 2 1 INTRODUCTION ONKALO is the underground rock characterisation facility that is currently being constructed by Posiva for spent nuclear fuel disposal research purposes. For long-term safety evaluation, characterisation of groundwater colloids is essential owing to the fact that groundwater colloids may transport radionuclides released from a failed spent fuel canister. Colloids are generally referred to as particles that do not appear to be sedimented even after a long period of time. The sizes of colloid particles are usually from 1 nm to 1 000 nm. Colloids in natural groundwaters consist of inorganic material (e.g. clay, oxides and carbonates) and organic material (mainly humic and fulvic acids). Owing to their small size, nano-micrometer particles may be transported along fractures in the bedrock. The transportation is controlled by groundwater movement, attachmentdetachment kinetics and sedimentation rates. Colloid formation and stability is dependent on the surrounding hydrogeochemical conditions e.g. pH, EC and redox. Generally colloids tend to form agglomerates in saline conditions and therefore the colloid concentrations are lower in saline waters than in fresh waters. The water samples were collected from groundwater stations ONK-PVA1, ONK-PVA5 and ONK-PVA10. The experiments were done for the first time in ONK-PVA5 and ONK-PVA10. The colloids of interest are colloids that reflect the in-situ conditions. During the sampling the in-situ conditions are tried to maintain at the sampling locations by avoiding sudden changes in flow rates and keeping the flow rates steady. The colloid concentrations were determined by single particle counting of scanning electron microscopy (SEM) micrographs taken from the filters. The particle calculations and size distribution determinations were performed with the assistance of ImageJ, an image analysis program. In addition, field-flow fractionation (FFF) was used in colloid particle distribution determinations. FFF is an elution method which have similar application possibilities as liquid chromatography (LC) (Baalousha, 2011). The purpose of this study was to analyse the colloid content and to compare the results with the previous ones. Enhancements in the procedure and new supporting analytical methods were expected to increase the reliability of the results. 3 2 EXPERIMENTAL 2.1 Preparations The colloid samples were collected on a Nuclepore polycarbonate 0.05 µm filter. The sampling assembly had three filter holders with sample containers. The sample containers were tared before sample collection. The sample lines were (polytetrafluoroethylene) PTFE ¼ in. (inner diameter) lines and the valves as well as the fittings were Swagelock’s (PTFE). The sample containers were made of plastic (high density polyethylene, HDPE). The filter holders were made of polycarbonate. The filters and filter holders were cleaned by pre-treatment with ultra clean Millipore water with an ultrasonic mixer to remove any particles from the filter surface (Nuclepore 0.05 µm) and filter holders before sampling. Further, before the sampling the filtering lines were flushed with the site water. The filter holders were assembled at the laboratory of Ramboll Analytics and sealed before the sampling to avoid any contamination by aerosol particles. 2.2 Sampling The sampling of colloid samples was performed on 20th and 21st of August 2013. The samples were taken from the groundwater stations ONK-PVA1 (depth about 14.6 m b.s.l), ONK-PVA5 (depth about 228.7 m b.s.l) and ONK-PVA10 (depth about 336.5 m b.s.l). Figure 1 illustrates the locations of the groundwater stations in ONKALO. The water yields in ONK-PVA1, ONK-PVA5 and ONK-PVA10 were 700, 110 and 50 ml/min. The parallel samples were taken simultaneously. Nuclepore 0.05 µm filters were used to capture the colloids from the sample stream. The filtration of parallel water samples took around 3 hours at ONK-PVA1, 1.5 hours at ONK-PVA5 and 2 hours at ONK-PVA10. Departing from the previous studies, the filter holders were vacuumdried after the sampling and sealed until the SEM-analyses. The water run through the filters was weighed (Table 1). No analyses were made from the filtered water. 4 Figure 1. The groundwater stations in ONKALO, Olkiluoto (figure made by Pöyry Finland Oy). 5 Table 1. Volume (ml) of the filtered water. Volume of water run through filter (ml) 277.4 1 017.9* 175.5 1 004.3 1 001.2 888.4 1 018.1 1 022.7 1 012.6 Sample ONK-PVA1 a ONK-PVA1 b ONK-PVA1 c ONK-PVA5 a ONK-PVA5 b ONK-PVA5 c ONK-PVA10 a ONK-PVA10 b ONK-PVA10 c *Discarded since the filter was probably leaking. One filter sample from ONK-PVA1 was discarded, because the filter holder was probably leaking. The filtered volumes were taken into account in result calculations. Samples were taken for FFF-measurements as well. The sampling sites were the same and the amount of non-filtered water samples were around 1 l. The sampling of each sample at ONK-PVA1 and ONK-PVA5 took about 10 min while the sampling of one sample took around 30 min at ONK-PVA10. 2.3 Single particle analysis The particle concentrations were calculated by counting single particles from the SEM micrographs. When taking the SEM micrographs, particles in the filters were found to be more homogenously divided compared to the previous studies. Thus, it can be noted that the drying phase was beneficial. The particle calculation was done with computer software ImageJ 1.47, which has been originally developed for microbiological studies for bacteria enumeration. In the previous study (Takala et al., 2011) software proved to be useful for the calculation of colloidal particles from SEM images. The program appeared to work well also for colloid enumeration from SEM micrographs. The size distribution (number of particles in size group/l) was produced for selected filter membranes. The distribution was calculated by the following equation 1: d [coll ] dØ dN dØ s i -1 s V (1) where Σ corresponds to the sum of the various micrographs examined, dN/dØ is the number of detected particles in size group dØ (µm), s (µm2) is the active surface area of the filter, si (µm2) is the surface area of the micrograph, and V is the filtered volume (l). The mass concentration (mg/l) for different size intervals was determined by the following equation 2: d [coll ] Ø 3 d [coll ] dØ 6 dØ (2) 6 where ρ is the particle density (mg/mm3) and Ø is the particle diameter (µm). The density of the particles was assumed to be 2 (mg/mm3) and the shape was assumed to be spherical. The particles on the micrographs are rarely spherical, and so the mass concentration is overestimated. A correction factor was not used. The micrographs were taken randomly from the filter surface with magnification factors of 1 000, 4 000 and 8 000. SEM micrographs were taken from the filters at Mikrofokus Oy in Helsinki by Simo Lehti. The number of the micrographs taken is shown in Table 2. Table 2. The samples and number of micrographs. Sample Magnification Number of micrographs ONK-PVA1 1 000 12 (three parallel filters, one discarded*) 4 000 24 8 000 24 ONK-PVA5 1 000 12 (three parallel filters) 4 000 24 8 000 24 ONK-PVA10 1 000 12 (three parallel filters) 4 000 24 8 000 24 * three parallel filters, of which sample “b” was discarded; double pictures from sample “c”. The computer aided particle calculation process initiated with a step of making a binary image of the SEM micrograph. The binary was needed for the software being able to detect the particles. The calculation boundaries were set so that particles smaller than 5x5 pixel were not calculated. The upper boundary was set to 100x100 pixel. The circularity factor was set to 0.01 – 1, where 1 is a perfectly round particle and when the value gets closer to 0 it indicates increasingly elongated shape. The circularity factor is calculated as following (equation 3): 4 (3) The image sizes were 1 000x750 pixels and resolution was 72 dpi. The scales of the images for each magnification are presented in the following table 3. Table 3. Scaling pixels to µm in different magnifications. Magnification Scale: 1 pixel to µm 1 000 1 pixel = 0.124 µm 4 000 1 pixel = 0.031 µm 8 000 1 pixel = 0.016 µm 7 ImageJ divides the particles by particle area. The diameter of the particles was thus calculated by equation 4. Particles, with diameter less than 2 µm were explored. 2 (4) 2.4 FFF measurements The FFF measurements were conducted at the University of Helsinki by Matti Jussila. The analyses were performed by Postnova AFFF equipment using regenerated cellulose membrane and 10 mM phosphate buffer as an eluent (pH 7). The tests were performed by three different molecular weight series (MW cut off): 30 000 g/mol, 5 000 g/mol and 1 000 g/mol. The membrane with 5 000 g/mol cut off was selected. The used detector was ultraviolet (UV; 254 nm) detector. Refractive index (RI) and dynamic light scattering (DLS) detectors were not suitable for these studies. With static light scattering (SLS) –detector visible peaks were found but they suffered from interferences. Peaks were, however, similar than with UV detector. Thus, results obtained with UV detector were selected for further examination. Since the low concentration, high sample volumes were needed to be used. The sample volume was 48 ml and it was injected to the channel with the 0.2 ml/min flow rate during 4 h. Injection was followed by a 30 min focusing phase and finally the sample run. The flow rate for the detector was 0.5 ml/min. Polystyrene sulphonate (PSS) standards were used for estimation of molecular weights. PSS standards were 15 000 g/mol, 30 000 g/mol and 1 000 000 g/mol. 8 3 RESULTS AND DISCUSSION 3.1 Single particle analysis According to the SEM micrographs and the analyst, the particles were rather equally distributed in all filters. This might be due to the filter drying phase readily after the sampling. Since the water was removed, it did not flush and move the content settled in the filters, as was suspected to have occurred in the previous studies. The results are presented and discussed for each sampling point in the following sections. 3.1.1 ONK-PVA1 results SEM micrographs were similar with all magnifications. An example of a micrograph (ONK-PVA1, 8 000 -fold magnification) and the binary figure of the micrograph are presented in the figure 2. Figure 2. SEM micrograph from ONK-PVA1 filter with magnification 8000 and its binary figure. The colloid particle concentration and the colloid mass concentration distributions are presented in the figure 3. The particle total concentration was 2.0 · 109 pt/l with mode particle size 0.2 µm. The mass concentration was calculated assuming the particles to be spherical and the particle density to be 2 mg/ml. The total mass concentration was 1.3 mg/l. The particle and mass concentration of particles with a diameter equal or less than 1 µm was 1.7 · 109 pt/l and 0.3 mg/l. 9 0,12 5,00E+08 4,50E+08 0,1 4,00E+08 3,50E+08 0,08 0,06 2,50E+08 mg/l pt/l 3,00E+08 2,00E+08 0,04 1,50E+08 1,00E+08 0,02 5,00E+07 0 0,00E+00 0 0,5 1 1,5 2 d (µm) Particle concentration (pt/l) Mass concentration (mg/l) Figure 3. Colloid particle (pt/l) and mass (mg/l) concentration distributions of ONKPVA1 water. 3.1.2 ONK-PVA5 results SEM micrographs were similar with all magnifications. An example of a micrograph (ONK-PVA5, 8 000 -fold magnification) and the binary figure of the micrograph are presented in the figure 4. Figure 4. SEM micrograph from ONK-PVA5 filter with magnification 8000 and its binary figure. The colloid particle concentration and the colloid mass concentration distributions are presented in the figure 5. The particle total concentration was 3.3 · 108 pt/l with mode particle size 0.2 µm. The mass concentration was calculated assuming the particles to be spherical and the particle density to be 2 mg/ml. The total mass concentration was 0.070 mg/l. The particle and mass concentration of particles with a diameter equal or less than 1 µm was 3.2 · 108 pt/l and 0.030 mg/l. 10 1,40E+08 0,008 1,20E+08 0,007 0,006 0,005 8,00E+07 0,004 6,00E+07 mg/l pt/l 1,00E+08 0,003 4,00E+07 0,002 2,00E+07 0,001 0,00E+00 0 0 0,5 1 1,5 2 d (µm) Particle concentration (pt/l) Mass concentration (mg/l) Figure 5. Colloid particle (pt/l) and mass (mg/l) concentration distributions of ONKPVA5 water. 3.1.3 ONK-PVA10 results SEM micrographs were similar with all magnifications. An example of a micrograph (ONK-PVA10, 8 000 -fold magnification) and the binary figure of the micrograph are presented in the figure 6. Figure 6. SEM micrograph from ONK-PVA10 filter with magnification 8000 and its binary figure. The colloid particle concentration and the colloid mass concentration distributions are presented in the figure 7. The particle total concentration was 3.8 · 108 pt/l with mode particle size 0.2 µm. The mass concentration was calculated assuming the particles to be spherical and the particle density to be 2 mg/ml. The total mass concentration was 0.11 mg/l. The particle and mass concentration of particles with a diameter equal or less than 1 µm was 3.6 · 108 pt/l and 0.040 mg/l. 11 1,20E+08 0,01 0,009 1,00E+08 0,008 0,007 8,00E+07 0,005 6,00E+07 mg/l pt/l 0,006 0,004 4,00E+07 0,003 0,002 2,00E+07 0,001 0 0,00E+00 0 0,5 1 1,5 2 d (µm) Particle concentration (pt/l) Mass concentration (mg/l) Figure 7. Colloid particle (pt/l) and mass (mg/l) concentration distributions of ONKPVA10 water. 3.1.4 Discussion The colloid particle concentration distributions were more or less equal within all sampling points. The amount of colloids was somewhat higher in ONK-PVA1 water compared to ONK-PVA5 and ONK-PVA10 water, which had similar concentrations. The colloid mass concentration distribution was different in ONK-PVA1 water compared to the ONK-PVA5 and ONK-PVA10 water. With particle size above one micrometer in diameter the colloid mass concentration in ONK-PVA1 water increased. Instead, in case of ONK-PVA5 and ONK-PVA10 water, the mass concentration was the highest somewhat below one micrometer in diameter and decreased with larger particles. Some differences were found also in visual appearance of SEM micrographs (figure 2 vs. figures 4 and 6). The peak in colloid particle concentration distribution between 0.6 – 1 µm was assumed to be erroneous and possibly being caused by the selected magnifications in SEM analysis. It is possible that the 1 000 and 4 000 –fold magnifications had too wide gap diminishing some of the particles in the range. Without this gap, the particle concentration distribution might have been more uniform with one peak only (maximum in 0.2 µm). 3.1.5 Reliability evaluation The reliability of the colloid concentrations measured with different magnification was attempted to be evaluated by calculating the expected number of colloids in a specified size group based on the observed number of colloids in the specified size group from the image with a small magnification factor. The expected number of colloids in a specified size group for a higher magnification factor can be calculated by the following equation (equation 5): 12 , (5) Where nimgi is the number of images taken with magnification factor ximgi and ncoll,imgi is the total number of colloids in the specified size group from images img i. img i are the images with the higher magnification factor (f.ex 4 000) and img j are the images with the lower magnification factor (f.ex 1 000). The relative percentage difference (RPD) can be calculated based on the expected number of colloids in the specified size group and the observed number of colloids as presented in the following equation. | | (6) The expected numbers of colloids for size groups from 0.7 – 0.8, 0.8 – 0.9 and 0.9 – 1.0 µm were calculated from SEM images taken from filter ONK-PVA5 (sample a). The results are summarized in table 4. The detection limit for magnification factor 1 000 is 0.7 µm, that is why the evaluation was not performed for smaller size groups. Table 4. Summary of calculated differences from expected number of colloid and observed number of colloids from SEM images taken from ONK-PVA5 (parallel samples a, b and c). 1 000 x 166 876 652 377 µ 42 219 163 94 4 000 x 215 106 85 35 RPD 418% 52% 48% 63% µ 10 55 41 24 8 000 x 43 26 19 14 RPD 314% 53% 53% 41% x is the observed number of colloids and µ is the expected number of colloids, calculated with equation 5. The images taken with magnification factor 1000 were used as reference. Based on the results the RPD ranges from 63% to 41%, except for size group 0.7 – 0.8, where the observed number of colloids is actually higher in the images with magnification factor 4 000 and 8 000 than in the images taken with a 1 000 magnification. This is probably due to image background noise which disturbs the detection of colloids that are just above the detection size, resulting in a lower count of colloids in this size group. The colloids are not evenly distributed onto the filter surface. The heterogeneity caused by the uneven distribution increases as the magnification factor increases i.e. the sample size reduces. To attempt to understand the uncertainty to the analysis caused by the distributional heterogeneity, the calculated colloid amount of size group 0.5 to 0.6 µm was calculated from individual images taken from filters ONK-PVA5(a), ONKPVA5(b) and ONK-PVA5(c). From each filter eight images were taken. The average number of colloids and the relative standard deviation (RSD) were calculated. This relative standard deviation represents the short scale (less that 1 cm) distributional heterogeneity as the images are collected from the same filter sample. The results are 13 summarized in table 5. The long scale distributional heterogeneity (5 to 10 cm) is further evaluated by comparing results from filters ONK-PVA5 (a), (b) and (c). Table 5. Summary of colloids (0.5 to 0.6 µm) and relative standard deviations (RSD) from images taken from filters ONK-PVA 5 (samples a, b and c). Magnification factor 4 000. Filter (ONK-PVA 5-) Average RSD (a) 42 34 27 9 37 20 44 22 29 41% (b) 1 2 0 0 1 1 0 4 1.1 121% (c) 16 18 18 37 23 6 1 2 15 80% The short scale distributional heterogeneity ranges from 41% to 121%. The long scale distributional heterogeneity can be calculated from the RSD of results from filters by counting the square root of the RSD square sums. The long scale distributional heterogeneity is 150%. This evaluation demonstrates that it is essential to not only analyze several images from a filter but also take the images evenly from around the filter. The Poisson distribution can be used to evaluate the sampling variability based on the number of particles detected using equation 7. √ (7) The Poisson distribution does not take into account the distributional heterogeneity, therefore the variation estimate obtained from equation 7 should be considered as a minimum variability. In figure 8 is illustrated the error calculated with equation 7 and the observed variability from filters ONK-PVA5 (a) to (c) magnifications 4 000 and 8 000. Figure 8. Dashed line is the RSD estimate using the Poisson model. The green and blue markers are RSD:s calculated from ONK-PVA5 filter images (Size 0.5 – 0.6 µm, x4000 and x8000). 14 The total variability of the colloid concentration analysis is difficult or even impossible to quantify with the available data. The variability is a function of among others the colloid concentration, sample size, analysis error and physical and temporal changes in the sample stream. The total portion of the sample analysed is proportional to the total size of the SEM images and the filter size. The total area of the SEM images is ca. 0.1 mm2 and the total surface area of the three filters is 4 770 mm2. This means that only 0.002% of the filtered volume is analysed. 3.1.6 Other observations During SEM analysis it was observed, that two samples from the station ONK-PVA5 contained some microbes. The microbes were present in filtered material. Also some salt crystals were observed. Neither microbes nor salt crystals affected the SEM analyses. 3.1.7 Comparison with the previous results The colloid particle concentrations and mass concentrations during years 2006 to 2013 are presented in the table 4. The results from 2006 are not comparable, owing to the differences in the sampling procedures and interpretations of results. Concentrations in ONK-PVA1 were the highest at 2013. Concentrations have showed significant variation which might be due to the unevenly spread particles in the filter. In 2013 studies, the particles were found to be evenly distributed compared to the previous years. Table 6. Colloid concentrations (Ø 0 – 1 µm) in 2006 – 2013. Date ONK-PVA1 18 April 2006 3 Oct. 2007 29 Oct. 2008 21 Dec. 2009 13 Oct. 2011 20.-21. Aug. 2013 ONK-PVA3 19 Nov. 2007 29 Oct. 2008 12 Oct. 2011 ONK-PVA5 20.-21. Aug. 2013 ONK-PVA10 20.-21. Aug. 2013 Sodium fluorescein [coll] pt/l [coll] µg/l ~20 µg/l <1 µg/l <1 µg/l <1 µg/l <1 µg/l <1 µg/l 2.7 · 109 1.6 · 104 3.5 · 108 1.7 · 106 6.9 · 107 1.7 · 109 200 0.001 200 0.5 6 300 ~20 µg/l <1 – 2 µg/l <1 µg/l 8.2 · 107 7.8 · 108 7.4 · 107 10 700 7 <1 µg/l 3.2 · 108 30 <1 µg/l 3.6 · 108 40 The sodium fluorescein concentrations were below detection limits in the all groundwater stations indicating that water used during drilling was not present. 15 3.2 FFF measurements The results of the preliminary FFF measurements showed significant differences between the samples from ONK-PVA1 compared to samples from ONK-PVA5 and ONK-PV10. In ONK-PVA1 samples, two peaks were found, whereas in ONK-PVA5 and ONK-PVA10 samples only one peak was visible. There was no corresponding observation in single particle analysis from SEM micrographs. Concentrations were estimated to be somewhat higher in the ONK-PVA1 sample compared to other samples. Even though the injected sample volume was kept constant, it was assumed that the amount of sample in the injection channel varied. Thus, the responses varied also. The standards are presented at the figure 9 and the results at the figure 10. In FFF measurements it should be noted that in general the retention time is dependent of the particle size (i.e. the higher the particle size the longer the retention time). Figure 9. FFF analyses from PSS standards. 16 ONK‐PVA1 UV absorbance PVA1a PVA1b PVA1c 0 10 20 30 40 50 Retention time (min) ONK‐PVA5 UV absorbance PVA5a PVA5b PVA5c 0 10 20 30 Retention time (min) 40 50 ONK‐PVA10 UV absorbance PVA10a PVA10b PVA10c 0 10 20 30 40 50 Retention time (min) Figure 10. Results of the FFF analyses of ONK-PVA1, ONK-PVA5 and ONK-PVA10 water. 17 According to the analyses, the peak found from the ONK-PVA1 samples in the retention time of 4 to 5 min could represent humic substances. The pore size of the used membrane is not suitable for retaining humics. However, since the electrostatic repulsion, a portion of the humic substances were suspected to retain in the channel in case of ONK-PVA1. The effect is somewhat dependent of the salt concentration, which, according to previous studies, has been lower in ONK-PVA1 water compared to ONKPVA5 and ONK-PVA10 water (Penttinen et al., 2011). With increasing salt concentration the particle size of humics decreases. Small particles tend to agglomerate and be removed from the water by settling in the bottom. This might be the reason that the peak conceivably representing humics was not found in the ONK-PVA5 and ONKPVA10 samples. The presumed content might be responsible for the visual difference found in SEM micrographs between ONK-PVA1 vs. ONK-PVA5 and ONK-PVA10. The latter peak in FFF analysis found from each sample (around 17 min) could represent the colloids, with mode particle size of 0.2 µm. According to PSS standards, the molecular size would be >1 000 000 g/mol. Generally speaking, it can be estimated that 0.1 µm corresponds roughly 1 000 000 g/mol, even though there is no direct correlation between the size and mass. When analysing the results it should be noted, that the molecular weight is not a linear function of retention time. No small peaks were observed on the right side of the peak from FFF analyses suggesting that there might be a sort of error in analysing the SEM micrographs or particle calculations. Even so, the shape of especially ONK-PVA1 curve seems to be elongated to the right in FFF results. 18 4 RECOMMENDATIONS In SEM analysis, the micrograph magnifications need to be selected so, that there is no significant gaps in the range of noticeable particles. In the future, SEM samples will be analysed at more regular magnification intervals (e.g. 2000x) to help bridge the gap between the 1000 and 4000 magnifications. In the future FFF studies, the samples are suggested to be preconcentrated e.g. with cross flow filtration. This would result in significantly shorter analysis time and the method development would be much faster. In addition, if the concentrations would become high enough and there would be knowledge of the features of the molecules/particles, also SLS or DLS detectors could be employed. 19 5 CONCLUSIONS Colloids (Ø 0 – 1 µm) were calculated to be present at concentrations of 300 µg/l in ONK-PVA1, 30 µg/l in ONK-PVA5 and 40 µg/l in ONK-PVA10. The mode particle size was 0.2 µm, as in the previous study. Compared to previous studies the colloids were found to be more uniformly distributed in the samples (filters) and thus the results were considered to be somewhat more representative. Nevertheless, the area for particle calculation is very low compared to the total area of the filter. FFF studies from the unfiltered groundwater showed similar particle distribution as the single particle analysis. However, there was an exception in ONK-PVA1 sample in which an extra peak was found representing material of small particle sizes (possibly humic substances). According to both methods, the amount of colloids was defined to be higher in ONKPVA1 compared to ONK-PVA5 and ONK-PVA10. Also indications of differences in the colloid content were found in ONK-PVA1 vs. ONK-PVA5 and ONK-PVA10 samples. 20 REFERENCES M. Baalousha, B. Stolpe, J. Lead, (2011); Flow field flow fractionation for the analysis and characterization of natural colloids and manufactured nanoparticles in environmental systems: a critical review, Journal of Chromatography A, 1218, 40784103. C. Degueldre, H. R. Pfeiffer, W. Alexander, B. Wernli, and R. Breutsch, (1996/I); Colloid properties in granitic groundwater systems, I: Sampling and characterisation, Applied Geochemistry, Vol 11, pp, 677-695, Elsevier Science Ltd. ImageJ, Image Processing and Analysis in Java. http://rsbweb.nih.gov/ij/ (6.9.2013). E. Järvinen, P. Manninen, M. Takala, and S. Vilhunen, (2011); Sampling and Characterisation of Groundwater Colloids in ONKALO at Olkiluoto, Finland, 20092010. Posiva Working Report 2011-29. T. Penttinen, S. Partamies, A-M. Lahdenperä, T. Lamminmäki, A. Lehtinen and S. Sireeni (2011); Results of Monitoring at Olkiluoto in 2010, Hydrogeochemistry, Posiva Working report 2011-44. M. Takala, S. Ojala, E. Järvinen, M. Manninen, (2012); Sampling and Characterisation of Groundwater Colloids at ONKALO, Olkiluoto Finland, 2011. Posiva Working Report 2012-61. M. Takala, E. Järvinen, and S. Vilhunen, (2009); Sampling and Characterisation of Groundwater Colloids in ONKALO at Olkiluoto, Finland, 2008. Posiva Working Report 2009-108. M. Takala, and P. Manninen, (2008); Sampling and Characterisation of Groundwater Colloids at ONKALO, Olkiluoto Finland, 2007. Posiva Working Report 2008-32. M. Takala, and P. Manninen, (2007); Sampling and Characterisation of Groundwater Colloids at ONKALO, Olkiluoto Finland, Posiva Working Report 2006-98.
© Copyright 2026 Paperzz