50097-820 Rev G

CER-1 & CER-2
Site Requirements
1
2
INTRODUCTION
The following document outlines the space requirements, water supply, drain and electrical specifications
required for the installation of the MEDIVATORS CER OPTIMA endoscope reprocessor models CER-1 and
CER-2 models. The CER-1 OPTIMA model is capable of reprocessing a single flexible endoscope, while
the CER-2 OPTIMA model can accommodate up to two endoscopes, allowing single or dual endoscope
reprocessing. This document is a guideline for installation specifications and requirements, in addition to any
facility/hospital requirements.
SPACE REQUIREMENTS
Chassis Dimensions
Height = 16 inches (41cm), 33 inches (84cm) with the lid fully open
Width = 22 inches (56cm), 26 inches (66cm) with barcode scanner mounted to CER
Depth = 20 inches (51cm), Allow a minimum of 4 additional inches (10.2cm) for water/drain/disinfectant hose
connections located in the back of the CER machine).
Empty Weight (approximated)
CER-1 OPTIMA model: 71 pounds (32kg)
CER-2 OPTIMA model: 77 pounds (35kg)
Cart Dimensions
Height = 33.0 inches (84cm)
Width = 21.5 inches (55cm)
Depth = 24.5 inches (62cm)
Active Vapor Management System Dimensions
Height = 32 inches (81cm)
Width = 23 inches (69 cm)
Depth = 24 inches (61cm)
Counter Top Installation
To install the endoscope reprocessor on a counter top, the counter top must be a minimum of 24 inches (61cm)
deep, (from front to back) and at a comfortable working height. Also ensure there is adequate clearance
between the open CER lid and any cabinets.
For the counter top, it is necessary to cut two 2 inch (5.1cm) holes in the back/rear of the counter top to
allow passage of the drain hose and the disinfectant chemical reservoir tubing. Two additional holes are also
required if the water inlet hose and electrical connection are located under the counter.
Cart Mounted Installation
The endoscope reprocessor can be placed on the cart, which is designed to provide easy access to the
disinfectant chemical reservoir and water filtration system.
3
WATER SUPPLY SPECIFICATIONS
Inlet water flow
Minimum of 2 to 3 gallons (7.6L to 11.3L) per minute
Water pressure
Minimum of 40 PSI (2.8 bar)
Maximum of 60 PSI (4.1 bar)

NOTE: PRESSURES OUTSIDE OF THIS RANGE WILL AFFECT THE OPERATION OF THE MACHINE. Consult
a qualified plumber for options to bring the water pressure within the 40-60 PSI range.
Water temperature
Cold water supply: Not to exceed 104°F (40°C)

NOTE: Endoscope manufacturers recommend that endoscopes not be exposed to temperatures above
130°F (55°C).
Water Hardness
Ensure water hardness is less than 12 gpg (200 ppm) for optimal performance.
De-ionized water (DI water), due to its corrosive nature, is not recommended for use in MEDIVATORS
reprocessors.
Reverse Osmosis (RO water) can be used however, not necessary since potable water can be used with the
MEDIVATORS reprocessors. If RO water is utilized, the pre-filter system must still be used.
Water Shut-off Valve
A shut-off valve must be installed before the water filtration system. It is important for the valve to be
mounted within the operator's reach to allow for easy and convenient daily use. The shut-off valve is used
for turning the water on at the beginning of the day, and shutting the water off at the end of the day.
Water Connection
All fittings on the reprocessor and the water filtration system are 3/4" male GHT (Garden Hose Thread).
All tubing necessary to connect the water filtration system to the reprocessor is included with the CER
OPTIMA® system. The water filtration system is installed in-line.
Local plumbing regulations vary regarding the installation of vacuum breakers for water line and back-flow
protection for the drain. Consult local plumbing regulations to ensure compliance.

2
NOTE: MEDIVATORS recommends checking local plumbing regulations regarding anti-siphon valves, vacuum
breakers, and plumbing connections, before attempting to install the endoscope reprocessor.
4
DRAIN SPECIFICATIONS
The drain fitting on the CER OPTIMA® reprocessor is 1.25 inch (32 mm) OD. It is recommended that the
reprocessor be installed with the 4-foot (1.2 m) flexible drain hose supplied with the CER OPTIMA system.
A drainpipe of no less than 1-1/2 inch (38 mm) diameter must be placed at least 1 foot (30.5 cm) below the
drain fitting of the reprocessor (having no loops), to ensure complete draining of fluids from the system. The
drain hose requires an air gap at the point it enters the drain. Consult local plumbing codes for size of air gap.
The facility drain must be capable of handling a discharge rate of 5 gallons per minute (19L per minute). A
facility drain having a dishwasher type "Y" drain is not an acceptable drain method and cannot be used. The
drain "Y" does not provide adequate flow rates nor air gap and must not be used. Refer to figure 1.
Figure 1
3
5
ELECTRICAL SPECIFICATIONS
CER OPTIMA® Reprocessor Electrical
120 VAC at 60 Hz, 3.0 amps, 1Ø
230 VAC at 50/60 Hz, 3.0 amps, 1Ø (230V models not available in North America)
CER OPTIMA Reprocessor Fuses (located in power entry module)
120VAC, Two 4.0 amp slow-blow fuses
230VAC, Two 2.0 amp slow-blow fuses
Power Cord
The CER OPTIMA reprocessor is supplied with one 10-foot (3.05 m) hospital grade removable power cord. A
minimum of one duplex (two outlets) outlet is required for the reprocessor and heated disinfectant reservoir. An
additional duplex outlet is required for the Active Vapor Management System and the Disinfectant
Transfer Pump.
Safety Classification
Ordinary protection
The reprocessor is not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or
nitrous oxide.


NOTE: For maximum safety, MEDIVATORS recommends the CER OPTIMA system only be used when
connected to an electrical outlet containing a Ground Fault Interrupter (GFI), or Residual Current
Device (RCD).
NOTE: MEDIVATORS recommends surge protectors be used for protection against electrical power spikes
and surges. Consult local electrical codes to ensure proper compliance.
Heated Disinfectant Reservoir (option) Electrical
120 VAC at 50/60 Hz, 3.5 amps, 1Ø
230 VAC at 50/60 Hz, 1.75 amps, 1Ø
Heated Disinfectant Reservoir Fuses (located in power entry module)
120VAC, 5.0 amp slow-blow fuses
230VAC, 2.0 amp slow-blow fuses
Disinfectant Transfer Pump (option) Electrical
120 VAC at 50/60 Hz, 0.5 amps, 1Ø
230 VAC at 50/60 Hz, 0.25 amps, 1Ø
Disinfectant Transfer Pump Fuses (located in power entry module)
120VAC, 1.5 amp slow-blow fuses
230VAC, 1.0 amp slow-blow fuses
6
VENTILATION REQUIREMENTS
During cycle operation and chemical heating process, chemical vapors are produced. It is a health and
safety requirement to have sufficient air changes, at least 10 to 12 times per hour, in the room to dissipate
the chemical vapor.
The flow of air should be considered by the customer as well. Air should flow away from the front of the unit
to the back, thus drawing fumes away from the operator's face. PPE should be used as required.
The unit may be supplied with an Active Vapor Management System to aid in chemical vapor management.
7
HIGH LEVEL DISINFECTANTS
The CER OPTIMA® reprocessor will monitor the minimum and maximum temperature of the high level
disinfectant to assure proper disinfections of the endoscope. The temperature limits are automatically
programmed when the disinfectant type is selected. If a disinfectant requires heating, this will require the use
of the heated disinfectant reservoir option. If a non- heated reservoir is utilized and the minimum required
disinfectant temperatures are not achieved, the CER OPTIMA reprocessor will halt and error. The cycle
cannot be completed without the proper disinfectant temperature. For further details, consult your local
MEDIVATORS Representative.
NORTH AMERICAN MINIMUM AND MAXIMUM ALARM PARAMETERS
DISINFECTANT TYPE
Minimum Temp alarm
HLD Contact Time
Number of Rinses
Maximum Temp alarm
RAPICIDE Disinfectant
35°C
5 min
2
40°
RAP OPA28H
25°C
5 min
2
40°
RAP OPA28U
20°C
10 min
2
40°
OPA
20°C
12 min
3
40°
GLUT
20°C
20 Min
2
40°
GLUT
25°C
20 Min
2
40°
GLUT
20°C
45 min
2
40°
®
INTERNATIONAL MINIMUM AND MAXIMUM ALARM PARAMETERS
DISINFECTANT TYPE
RAP OPA INT
Minimum Temp alarm
HLD Contact Time
Number of Rinses
Maximum Temp alarm
20°C
5 min
2
40°
MEDIVATORS®, RAPICIDE® and CER OPTIMA® are registered trademarks of Medivators Inc.
Medivators Website “Resource Center”
Go to: www.medivators.com, Select “Resource Center” and “User Library” for detailed user guides and hookup
information, report forms and logs, and product bulletins
MEDIVATORS Customer and Technical Support Toll Free: Phone: 1-800-444-4729 · FAX: 1-866-553-9444
Manufactured in the USA by:
Medivators Inc.
14605 28th Avenue North
Minneapolis, MN 55447 USA
Toll Free: +1.800.444.4729
Medivators BV
Sourethweg 11
6422PC Heerlen
The Netherlands
Tel: +31.45.5.471.471
www.medivators.com
© 2016 Medivators Inc. All rights reserved. 50097-820/G
Cantel Medical Asia/Pacific Pte. Ltd.
1A International Business Park
#05-01 Singapore 609933
Tel: +65.6227.9698
Cantel Medical Devices (China) Co. Ltd.
Unit 804-805, Innov Tower Block A,
Hongmei Road, Xuhui 200233 Shanghai
Tel: +86 21 60161380
Fax: +86 21 61210913