FACS Laboratory [email protected] Training Guide http://london-research-institute.org.uk/technologies/facs BD LSRs Operator Training LRI FACS lab 1 Components 1.1 Basic parts overview……………………………………………………………………………………….pg3-6 Overview of LSRs…………………………………………………….……………………3 Filling and emptying the tanks…………………………………..…………..……5 Starting up the LSR…………………………………………………………….………..6 Cleaning and Shutting down…………………………………………………………6 1.2 Fluidics…………………………………………………………………………………………………………….pg7-11 Basic fluidic mechanics in flow cytometry ……………………………………7 The sample injection port (SIP) …………………………………………………..9 Troubleshooting…………………………………………………………………………….10 o Unblocking a probe……………………………………………………………10 o De-bubbling filters……………………………….……………………………11 1.3 Optics……………………………………………………………………………………………………………...pg12-31 Optics Overview.……………………………………………………………………………12 Detection optics.……………………………………………………………………………13 1.4 Digital electronics………………………………………………………………………………………………pg32-34 Digital flow cytometry and the LSRs…………………………………………….32 Digital pulse processing..………………………………………………………………32 2 Fluorochromes & Multi-colour Flow 2.1 Commonly used fluorochromes………………………………………………………………………..pg35-39 2.2 Fluorochrome selection……………………………………………………………………………………..39-41 Viability dyes……………………………………………….………………………………….39 Fixable LIVE/DEAD amine reactive dyes……………………………………….40 Availability………………………………………………………………………………………40 Physical Characteristics………………………………………………………………….41 2.3 Combining fluorochromes………………………………………………………………………………….pg41-43 Spectral overlap………………………………………………………………………………41 Problem pairs.…………………………………………………………………………………41 Example combinations……………………………………………………………………42 2.4 Compensation…………………………………………………………………………………………………….pg44-49 Controls.…………………………………………………………………………………………44 Compensation beads.……………………………………………………………………45 The bi-exponential display.……………………………………………………………47 Compensation - the median fluorescence technique..…………………37 Copying spectral overlap.………………………………………………………………49 Automatic compensation………………………………………………………………..49 3 BD FACSDivaTM : Software Basics 3.1 The Browser, Worksheet, Inspector and Acquisition windows………………………..pg50-52 3.2 Example experiment………………………………………………………………………………………….pg52-73 Working with plots, regions and gates.…………………………………………55 Baseline PMT voltages……………………………………………………………………62 Compensation…………………………………………………………………………………63 o Automatic compensation……………………………………………..…….68 Exporting data from FACSDiva as listmode (.fcs) files…………………72 Working with templates…………………………………………………………………73 4 Quality Control 4.1 Routine Checks..…………………………………………………………………………………………………pg74-75 4.2 Laser Delay, Area Scaling and Windows Extension………………………………………….pg75-76 5 Sample preparation 5.1 Harvesting cells………………………………………………………………………………………………….pg77 5.2 Resuspending cells for analysis…………………………………………………………………………pg78 6 Glossary.………………………………………………………………………………………………………………pg79-84 7 Troubleshooting Guide .……………………………………………………………………………………pg85 1.1 Basic Parts Overview In the FACS Lab, all cytometers are normally switched on at 9am. When booking to start before or at 9, or on weekends, please be aware that the LSRs have multiple laser lines that need to be given a minimum of 30 minutes to warm up before running samples. Overview of LSRs From the control panel you can perform the operations RUN, STANDBY and PRIME, and make some sample flow rate adjustments. The machine should be on STANDBY with a tube of water when you come to it. Check that the sheath tank is filled to the line and the waste tank is empty. If the sheath tank runs dry, air will enter the system and can be difficult to remove; you will be unable to acquire any data if this happens. * Important! * Before starting an experiment, check the levels of sheath fluid and waste. If you are using the machine for a long time check the tanks regularly. We cannot stress enough how critical it is that you are capable of maintaining the tanks and purging air from the system. If you are unsure of how to do so, please ask a member of the FACS Lab, to avoid ruining an important experiment, and possibly the next user’s experiment too! LSRIIA and LSRIIB overview 1. Power switch 2. Sheath tank 3. Waste tank 4. Control panel 5. Sample Injection Port (SIP) 6. Sheath Filter 7. Sheath line air purge (roll-valve) Fig. 1.1.1 Location of important parts on LSRII. 3 LSRFortessa A, B and C overview 1. Power switch 2. Sheath tank 3. Waste tank 4. Control panel 5. Sample Injection Port (SIP) 6. Sheath filter 7. Sheath line air purge (roll-valve) Fig. 1.1.2 Location of important parts on LSRFortessa 4 Filling and emptying the tanks A B Fig. 1.1.3 A De-pressurising sheath tank by pulling up on valve (black circle). B Lid of waste tank showing waste line (orange tube) and alarm (black wire). First check the machine is in STANDBY (standby button on control panel is orange): *Filling the sheath tank* Fig. 1.1.3 A de-pressurising tank Unclip the sheath pressure line (red circle - green tube on top of tank). De-pressurise the tank (black circle - pull metal release valve up). Unscrew & remove lid. Fill with PBS to groove in tank - do not overfill. PBS can be located in bottles on the shelf in the corridor. Replace the lid and re-connect the pressure line. De-bubble the LSR – it is good practice to always do so when you re-fill the tank. Please refer to the Troubleshooting section in section 1.2 for instructions. *Emptying the waste* Fig. 1.1.3 B waste tank lid Unclip the orange line (waste line) and disconnect the alarm (black wire - twist connection) from the lid of the waste tank. Unscrew the lid: add 100g Virkon powder (use measuring cup in Virkon container) to the tank. Empty the waste into the sink by LSRIIA (Virkon is kept under the sink). Replace lid, re-connect waste line and alarm. 5 Starting up the LSR Switch on the LSR and start the computer workstation. There is no password to log into Windows (hit “Enter”). Priming the machine Before running any samples it is good practice to prime the fluidics to remove any bubbles from the flow cell. Additionally, before you run your sample (or while you are waiting for the lasers to warm up), it is a good idea to check the machine is clean by simply running fresh dH20 in a new tube. If move arm to side during Prime, remember to return it after! When running dH20, have threshold at 15,000 and FSC at voltage used for cells. The event rate should be mostly 0 events/second but up to 100evt over 30 seconds due to bubbles, noise. If you have time, it does not hurt to run some cleaning agents for a short period, followed by dH20. After these steps and allowing the lasers to warm up, the machine is ready to run your samples - this will be covered later. *Priming the LSR* Remove the tube of dH20 from the sample injection port (SIP). With SIP arm to the side, press PRIME; this empties out the flow cell and refills it to remove bubbles. LSR returns to STANDBY when done. Repeat 1-2 times. Cleaning and shutting down When you are done running your samples, it is imperative to clean the LSR sufficiently to prevent clogging of the SIP and remove dyes which may remain in the tubing: *Cleaning procedure* Run tube of detergent for 5 min on HI (BD FACS Rinse). Run tube of bleach (BD FACS Clean) for 3min on HI if DNA dyes/viability dye such as DAPI, Hoechst, PI etc. have been used; if not - proceed to next step. Run tube of dH20 for 5min. Check the machine is clean with a fresh tube of dH20. If machine is not clean, repeat Clean & Rinse steps. If the machine has been left dirty, running cleaning agents for a longer length of time will usually help; otherwise the machine may require a long clean which is routinely performed by the FACS Lab staff. The last user of the day should switch off the cytometer after the cleaning. IMPORTANT! Read also Troubleshooting, in Section 1.2 - purging air from the machine. 6 1.2 Fluidics A pressurised fluidics system delivers the sample to the flow cell where interrogation by the lasers takes place, using the process of hydrodynamic focusing to help create a tight, single cell flow to allow us to look at each cell one by one. This is achieved by laminar flow, higher sheath to core velocities and by the shape of the flow cell. Flow cell Lasers Sample Lo/Med/Hi controls Vacuum Waste Sheath Pressure line Fig. 1.2.1 Basic diagram of flow cytometer fluidics system Basic Fluid Mechanics in Flow Cytometry In the cuvette type cytometers (e.g. Caliburs, LSRs), cells are introduced into the middle of the flow cell, injected into the centre of a smoothly flowing stream (the sheath fluid). Since both the core stream (consisting of your injected sample) and the sheath stream are flowing smoothly, laminar flow is observed where the two will maintain their relative positions and do not mix. We want to analyse one cell at a time, so a narrow core carrying single cells to the laser interrogation point(s) is critical. One of the ways this is achieved is by using a cuvette where the sides taper inwards towards the area of laser interrogation. 7 Note that fluid flow at the exit from the region of cuvette constriction into the narrower region where laser interrogation occurs can be described as slug flow (having constant velocity across the fluid cross section). This continues for a while, but quickly returns to a parabolic laminar flow profile (velocity at the edges is slower than at the centre). Fig. 1.2.2 Fluid dynamics in the flow cell. (Shapiro’s Practical Flow Cytometry, 4th Ed.) It is important that laser interrogation occurs before the parabolic profile is reestablished so that cells will have the same velocity - and therefore take the same time to pass through the laser beam, regardless of the position within the core. In other words, variations in distribution will not have an impact on illumination time, giving better instrument precision. (Further detail can be found in Howard Shapiro’s Practical Flow Cytometry 4th Ed). Which speed to run your sample- HI, MED, LO? Although we cannot adjust the pressure of the sheath, we can change the pressure applied to the sample, to inject more/less sample per unit of time. This is useful when the sample is very dilute or to save time when looking for very rare events (i.e. need to run through a large number of cells). Increasing the sample pressure will result in a widening of the core carrying your particles of interest, as shown in Fig. 1.2.3. This affects how uniformly particles flow past the interrogation points and will affect measurements such as DNA (which are best run in LO) where precision is important. Increasing sample pressure will result in an increase in DNA histogram CV. For qualitative measurements such as immuno-phenotyping, there may be some loss of resolution but running in MED/HI is generally not a problem, when using log scale. 8 Fig. 1.2.3 Core stream diameter increase is observed when sample pressure is high (right) compared to low sample pressure (left). Samples should be sufficiently concentrated to enable a reasonable run time, especially if you need to run your samples in LO. Correct sample preparation (see section 5) is also vital to prevent blocking the probe: there is no PAUSE function in DIVA, one must stop recording, clear the blockage, then APPEND to the saved file. It is better to avoid having clumps in the first place. The sample injection port (SIP) The SIP comprises the sample injection tube and support arm. The sample injection tube is encased in a sleeve. When the support arm is open, the sample injection tube backflushes sheath fluid. This is then removed by a vacuum pump (activated when the arm is open) up the outer sleeve to the waste tank, helping to clean the injection tube between samples. Fig. 1.2.4 Sample injection port component diagram and photograph. 9 *Important!* The machine starts taking up your sample as soon as you have installed the tube on the SIP, NOT when you press Acquire. Also, leaving a sample on the injection tube with the support arm open will result in loss of sample! Be quick with the support arm when putting tubes in place. Troubleshooting Unblocking a probe Properly prepared samples should not block the machine. Samples should be carefully checked for clumps that can block the probe (filter and/or pipette sample if ANY clumps visible). Check first that the sheath tank is not empty: Remove your sample and Prime several times. If this does not work, run Clean followed by Rinse for 5-10 minutes. Run dH20 for 2-3 minutes before putting your sample back on. Fig. 1.2.5 Unblocking the sample probe using a syringe. For a persistent clog, it is sometimes necessary to apply a gentle vacuum by syringe there is a large syringe with rubber tubing in the lab for this purpose. Please ask a member of staff to demonstrate. Make sure that when you prime the machine, there is no tube in place (first, the flow cell empties- flushing out the probe, then the flow cell re-fills). Keeping a tube of Rinse/Clean/dH20 on as you do this is counter-productive, as any debris flushed out will also be taken back up again. *Problematic samples with clumps* Please be aware that if you can see the clump, it is too big! 10 De-bubbling filters If the machine has been allowed to run dry you may need to purge air from the filter on the sheath tank and sheath line, and also the flow cell (using Prime). Air can be hard to remove; it is better to prevent this from happening in the first place by checking the tanks! *Knowing how to troubleshoot is important to avoid running into trouble if you will be using the machine after hours* You may be left unable to run the rest of your experiment, or cause subsequent users to be unable to run theirs. This problem may present itself as unexpected dots on your plots (caused by air bubbles, includes both randomly distributed events and low scatter “debris-like” signals) accompanied by the LSR not showing events from your tube. De-bubbling Filters A B Fig. 1.2.6 A Purging air from sheath filter: Tap filter gently and check for bubbles. Carefully roll valve and PBS is released along with bubbles. Tilting the tank/filter as you do this in order to encourage the bubbles to move towards the purge valve can also be useful, please ask a member of staff to demonstrate. B Purging air from the sheath line: roll valve (PBS is released, along with bubbles). The cytometer should be on standby while de-bubbling filter/line. 11 1.3 Optics Optics Overview Excitation: Light sources: lasers LSRIIA - Blue (488nm), Violet (405nm), UV (355nm), Red (633nm) LSRIIB - Blue (488nm), Violet (405nm), Yellow-Green (561nm), Red (633nm) Fortessa A - Blue (488nm), Violet (405nm), Yellow-Green (561nm), Red (633nm), UV (355nm) Fortessa B - Blue (488nm), Violet (405nm), Yellow-Green (561nm), Red (639nm), UV (355nm) Fortessa C - Blue (488nm), Violet (405nm), Yellow-Green (561nm), Red (639nm), UV (355nm Filters/mirrors: Route the laser beam to the flow cell to interrogate the sample. Collection: Scattered light and any emitted fluorescence are directed by fibre optics to the appropriate block of photomultiplier tubes (PMTs or ‘detectors’) specific for each laser. Filters/mirrors: Used to guide the light to the appropriate PMT in the block. Detectors: Receive the scattered light and fluorescence signals. From these photons, an amplified electrical signal is generated that we observe through the Diva software. Fig. 1.3.1 Example diagram of LSRII optical bench 12 The optical bench of the LSRs see a departure from the layout seen in the FACS Scan, Calibur and Vantage. Instead of passing the light through several mirrors to the correct PMT, the fluorescence from each laser (spatially separated) is directed to the relevant octagonal or triagonal arrangement of PMTs. This results in more efficient collection of the signals. Detection Optics The filters used in the cytometers are optimized for their respective detectors. They can be changed but their transmission characteristics should be known. Gloves should be worn when handling the filters and dichroic mirrors, and they should be held by their edges to avoid leaving fingerprints on the transmission surfaces. The specifications of the filters can be found on the outside edge (by which they should be handled) in small print. A short pass filter (SP), for example 450SP, transmits wavelengths shorter than its specification- i.e. 450nm. Long pass (LP) filters are similarly named, for example 610LP, which transmits light longer than 610nm. Band pass (BP) filters transmit within a given range and are named accordingly. For example, a “710/50BP” is a filter that transmits 710nm±25nm (685nm-735nm), “530/30BP” transmits 515-545nm. Dichroic mirrors can also have a short or long pass function, transmitting longer/shorter than the specification. These are also known as beam-splitters as the rest of the light is reflected. Dichroics are particularly useful for directing light of specific wavelengths down the appropriate route within the optical bench. In front of each PMT is a dichroic long pass (DCLP) mirror. At the first PMT, the longest wavelength to be analysed passes through and everything shorter is reflected to the next PMT. At the second PMT, there is a similar arrangement with DCLP allowing the next longest wavelength to pass and reflecting the rest; and so forth down the PMTs. This set up loses less light than the older collection optics (about 10% is lost passing through the dichroic- generally, mirrors are more efficient at reflecting than filters are at transmitting light). 13 Fig.1.3.2 Filter processing of light to allow specified wavelengths to be transmitted. The standard configuration of filters is listed inside the hood of the LSRII; it is a good idea to check that the person before you has not changed the filters without returning them to the standard set-up. Please ask a member of the FACS Lab to demonstrate how to change optical filters before attempting to do so yourself. LSRIIA PMT Position LSRIIA Diva Parameter Blue octagon, A 780/60 blue Longpass Filter 735DCLP Blue octagon, B 695/40 blue 680DCLP Blue octagon, C 660/20 blue 635DCLP Blue octagon, D 610/20 blue 595DCLP Blue octagon, E 575/26 blue 550DCLP Blue octagon, F 530/30 blue 505DCLP Blue octagon, G SSC - FSC FSC - Violet trigon, A 530/30 violet 505DCLP Violet trigon, B 450/50 violet UV trigon, A 525/50 UV 450DCLP UV trigon, B 440/40 UV - Red trigon, A Red trigon, B 780/60 red 730/45 red 755DCLP Red trigon, C 660/20 red - DCLP = Dichroic Long Pass mirror. Fig. 1.3.3 Standard optical filter configuration in LSRIIA. 14 685DCLP Blue octagon, A LSRIIB Diva Parameter 695/40 blue Blue octagon, B 575/26 blue 550DCLP Blue octagon, C 530/30 blue 505DCLP Blue octagon, D SSC - Blue octagon, E - - Blue octagon, F - - Blue octagon, G - - FSC FSC - Violet trigon, A 660/20 violet 635DCLP Violet trigon, B 605/40 violet 595DCLP Violet trigon, C 560/40 violet 550DCLP Violet trigon, D 510/20 violet 505DCLP Violet trigon, E 450/50 violet - Violet trigon, F - - Y/Green octagon, A 780/60 yellow 735DCLP Y/Green octagon, B 705/70 yellow 690DCLP Y/Green octagon, C 660/20 yellow 635DCLP Y/Green octagon, D 620/40 yellow 600DCLP Y/Green octagon, E 585/15 yellow 570DCLP Y/Green octagon, F - - Red trigon, A 780/60 red 735DCLP Red trigon, B 710/50 red 685DCLP Red trigon, C 660/20 red - LSRIIB PMT Position DCLP = Dichroic Long Pass mirror. Fig. 1.3.4 Standard optical filter configuration in LSRIIB. 15 Longpass Filter 685DCLP Fortessa A PMT Position Blue octagon, A Fortessa Diva Parameter 780/60 blue Longpass Filter 750DCLP Blue octagon, B 695/40 blue 685DCLP Blue octagon, C 610/20 blue 600DCLP Blue octagon, D 580/30 blue 550DCLP Blue octagon, E 530/30 blue 505DCLP Blue octagon, F SSC - FSC FSC - Violet octagon, A 530/30 violet 475DCLP Violet octagon, B 450/50 violet - UV trigon, A 530/30 UV 505DCLP UV trigon, B 450/50 UV - Red trigon, A 780/60 red 750DCLP Red trigon, B 730/45 red 710DCLP Red trigon, C 670/14 red - Y/Green octagon, A 780/60 yellow 750DCLP Y/Green octagon, B 710/50 yellow 685DCLP Y/Green octagon, C 670/30 yellow 635LP Y/Green octagon, D 610/20 yellow 600DCLP Y/Green octagon, E 582/15 yellow - DCLP = Dichroic Long Pass mirror. Fig. 1.3.5 Standard optical filter configuration in LSRFortessa A. 16 Fortessa B and C PMT Fortessa Diva Parameter Longpass Filter Blue octagon, A 710/50 blue 685DCLP Blue octagon, B 530/30 blue 505DCLP Blue octagon, F SSC - FSC FSC - Violet octagon, A 670/30 violet 630DCLP Violet octagon, B 610/20 violet 600DCLP Violet octagon, C 586/15 violet 570DCLP Violet octagon, D 540/30 violet 535DCLP Violet octagon, E 525/50 violet 505DCLP Violet octagon, F 450/50 violet - UV trigon, A 450/50 UV 505DCLP UV trigon, B 530/30 UV - Red trigon, A 780/60 red 750DCLP Red trigon, B 730/45 red 690DCLP Red trigon, C 670/14 red - Yellow/Green octagon, A 780/60 yellow 750DCLP Yellow/Green octagon, B 710/50 yellow 685DCLP Yellow/Green octagon, C 670/30 yellow 635LP Yellow/Green octagon, D 610/20 yellow 600DCLP Yellow/Green octagon, E 586/15 yellow - Position DCLP = Dichroic Long Pass mirror. Fig. 1.3.6 Standard optical filter configuration in LSRFortessa B and C. 17 LSRII (A) Blue Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.7 18 LSRII (A) Trigons: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.8 19 LSRII (B) Blue Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.9 20 LSRII (B) Yellow-Green Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.10 21 LSRII (B) Violet Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.11 22 LSRII (B) Red Trigon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.12 23 LSRFortessa A Blue Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.13 24 LSRFortessa A Yellow/Green Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.14 25 LSRFortessa A Violet Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.15 26 LSRFortessa A Trigons: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.16 27 LSRFortessa B and C Blue Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.17 28 LSRFortessa B and C Yellow/Green Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.18 29 RFortessa B and C Violet Laser Octagon: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.19 30 LSRFortessa B and CTrigons: filter optics and collection pathway BP = Band Pass filter DCLP = Dichroic Long Pass mirror Figure 1.3.20 31 1.4 Digital Electronics Digital flow cytometry and the LSRs The LSRs are BD’s top of the range benchtop analyser, offering flexibility to multi-colour applications. They are faster: the electronics are capable of digitising signals 10 million times per second; there is increased processing capability due to reduced electronic “dead time”. They have better resolution: improved ADC (increased bit size data) and electronics giving better channel resolution and detection sensitivity. We can have a full compensation matrix: inter-laser compensation is possible and can be applied or changed on/offline. More information can be obtained and digital pulse processing of all parameters is possible. Digital pulse processing In basic terms, analog signals have a continuous distribution (i.e. they can take any value within a defined range) and digital signals are a discrete distribution (i.e. can only have certain binned values the number of which is determined by the digitisation). Ultimately, all cytometers are digital but they differ in where digitisation takes place. In any case the more discrete channels that a distribution can be divided into, the better the stored data relates to original data. Analog: PMT Compensation LIN/LOG Transformation Digitise (by ADC) Digital: PMT Digitise (by ADC) Compensation LIN/LOG Transformation Fig. 1.4.1 Order of data processing in analog vs. digital systems. In the LSRs, signals from the detector (PMT) are continuously digitised by an Analog to Digital Converter (ADC). The ADC captures a snapshot of the electric voltage from the PMT and represents it as a digital number that can be sent to a computer. By capturing the voltage millions of times per second (millions of time “slices”), you can get a very good approximation to the original light signal. The ADC in the LSRs for height is 14-bit, this gives 1014 i.e. 16,384 channels available (16, 384 “levels” that the signal may be assigned). Each measurement represents the height of the pulse at that time. The area ADC gives 18-bit data i.e. 1018 or 262,144 channels. If we add together all the time slices of a pulse, we get the area, i.e. the total amount of fluorescence coming from the cell. 32 If we attach an oscilloscope to a detector on a flow cytometer, we can see the voltage pulse that is generated: Photon of light Photomultiplier Tube (PMT) Voltage pulse Fig. 1.4.2 Photon conversion by PMT to a voltage pulse. We can measure the height of the total pulse by looking at the output of the ADC (channel number read-out): 16,384 Pulse Height Photon of light Photomultiplier Tube (PMT) 0 Fig. 1.4.3 PMT generated pulse: pulse height readout from ADC. The area of the pulse is determined by adding the height values for each time slice of the pulse (which is determined by the speed of the ADC, which is 10MHz i.e. 10 million per second or 10 per microsecond). The area is a better representation of the total amount of fluorescence: Photon of light Photomultiplier Tube (PMT) Pulse Area (between 0 - 262,144) Fig. 1.4.4 PMT generated pulse: pulse integral readout from ADC. 33 The width parameter can be thought of as being directly related to the time that a particle of interest takes to flow past through the laser beam. Fig. 1.4.5 Time taken to pass through laser is proportional to width parameter. However, in the LSR, this is a calculated rather than a measured parameter. Because the width can also be taken as the number of time slices sampled, this is calculated as Area/Height x 64K. Because fluorescence signals from the PMTs are digitised immediately in digital cytometers, there is the advantage that noise introduced by other electronic circuitry is reduced, and cytometer dead time is reduced. The signals are sampled over tiny time slices and each given a value; digitised values are retained in buffer memory until the area integral and pulse width measurements are completed. The digital values are then passed to other electronics such as gating boards, before interfacing with the computer workstation (Diva software). Note: the data is stored as linear, if you wanted to perform log calculations, the computer “looks up” the correct value using a “Look Up Table”. The fcs 3 files contain the uncompensated values, (but the compensation matrix is also saved) meaning that compensation can be augmented post collection (remember that compensation on saved analog files cannot be taken off it is possible to add compensation using analysis software but too much compensation is unsalvageable). The width is a processed parameter on the LSR; H and A are both measured but we generally use the A parameter on LSR because it represents total fluorescence coming from the cell. 34 2.1 Commonly Used Fluorochromes Laser 488nm Blue 633nm Red 407nm Violet 355nm UV Collection Filter (bandpass) Diva Parameter Name 530/30 530/30 blue 575/26 575/26 blue 610/20 610/20 blue 660/20 660/20 blue 695/40 695/40 blue 780/60 780/60 blue PE-Cy7 660/20 660/20 red APC, Alexa 633, Alexa 647, TO-PRO3, Cy5, DRAQ5, DRAQ7, Dye Cycle Ruby 730/45 730/45 red APC-Cy5.5, 780/60 780/60 red Fluorochromes FITC, Alexa 488, eGFP, CFSE, Dye Cycle Green PE, Cy3, Sytox-Orange, Dye Cycle Orange PI, PE-Texas Red, PE-Alexa Fluor 610 PE-Cy5 (also “Tricolour” or “Cychrome”), 7AAD PerCP, PerCP-Cy5.5, DRAQ5, Dye Cycle Ruby APC-Cy7, DRAQ7, Dye Cycle Ruby DRAQ7, Alexa Fluor 750 Alexa 405, CFP, Pacific Blue, DAPI, Hoechst (blue), Celltrace Violet, Dye Cycle Violet, BV421, VPD450, Calcein Violet 450, Fixable Viability Dye eFluor 450, Cell Proliferation Dye eFluor 450 Qdot 525, Pacific Orange, Krome Orange 450/50 450/50 violet 530/30 530/30 violet 440/40 440/40 UV Celltrace Violet, DAPI, Hoechst (blue) 525/50 525/50 UV PI Fig. 2.1.1 This table gives examples of the fluorochromes most commonly used on LSRIIA, showing which laser they are excited by and which filter should be used to collect the signal. All of these fluorochromes can be used with the standard LSRII A filter arrangement as shown in section 1.3. 35 Laser 488nm Blue 633nm Red 407nm Violet 561nm YellowGreen Collection Filter (bandpass) Diva Parameter Name 530/30 530/30 blue 575/26 575/26 blue 695/40 695/40 blue 660/20 660/20 red 710/50 710/50 red 780/60 780/60 red 450/50 450/50 violet 510/20 510/20 violet 560/40 560/40 violet Qdot 565, Pacific Orange, Krome Orange, BV570 605/40 605/40 violet Qdot 585, Qdot 605, BV605 660/40 660/40 violet Qdot 655, BV650 585/15 585/15 yellow PE, Alexa Fluor 555 620/40 620/40 yellow 660/20 660/20 yellow PE-Texas Red, M-Cherry, M-RFPMars PE-Cy5 (also “Tricolour” or “Cychrome”), 7AAD, E2-Crimson, M-Plum 705/70 705/70 yellow 780/60 780/60 yellow Fluorochromes FITC, Alexa 488, eGFP, CFSE, Dye Cycle Green PI, Cy3, Sytox-Orange, Dye Cycle Orange PerCP, PerCP-Cy5.5, PerCP-eFluor710, DRAQ5, Dye Cycle Ruby APC, Alexa 633, Alexa 647, TO-PRO-3, Cy5, DRAQ5, DRAQ7, Dye Cycle Ruby APC-Cy5.5, Alexa 700, Cycle Ruby DRAQ7, Dye APC-Cy7, DRAQ7, Alexa Fluor 750 Alexa 405, CFP, Pacific Blue, DAPI, Hoechst (blue), Celltrace Violet, Dye Cycle Violet, BV421, VPD450, Calcein Violet 450, Fixable Viability Dye eFluor 450, Cell Proliferation Dye eFluor 450 Qdot 525, BV510, BD Horizon V500, DRAQ7 PE-Cy7, DRAQ7 PE-Cy5.5, Fig. 2.1.2 This table gives examples of the fluorochromes most commonly used on LSRIIB, showing which laser they are excited by and which filter should be used to collect the signal. All of these fluorochromes can be used with the standard LSRIIB filter arrangement as shown in section 1.3. 36 Laser 488nm Blue 561nm YellowGreen 355nm UV 633nm Red 405nm Violet Collection Filter (bandpass) Diva Parameter Name 530/30 530/30 blue 580/30 575/26 blue 610/20 610/20 blue 695/40 695/40 blue PerCP, PerCP-Cy5.5, PerCP-eFluor710, DRAQ5, Dye Cycle Ruby 780/60 780/60 blue PE-Cy7 582/15 582/15 yellow PE, Alexa Fluor 555 610/20 610/20 yellow 670/30 670/30 yellow 710/50 710/50 yellow PI, PE-Texas Red, M-Cherry, M-RFPMars PE-Cy5 (also “tricolour” or “cychrome”), 7AAD, E2-Crimson, M-Plum PE-Cy5.5, DRAQ7 780/60 780/60 yellow PE-Cy7, 450/50 450/50 UV DAPI, Hoechst (blue), Celltrace Violet 530/30 530/30 UV 670/14 670/14 red PI APC, Alexa 633, Alexa 647, TO-PRO-3, 730/45 730/45 red APC-Cy5.5, 780/60 780/60 red 450/50 450/50 violet 525/50 525/50 violet APC-Cy7, Alexa Fluor 750, DRAQ7 Alexa 405, CFP, Pacific Blue, DAPI, Hoechst (blue), Celltrace Violet, Dye Cycle Violet, BV421, VPD450, Calcein Violet 450, Fixable Viability Dye eFluor 450, Cell Proliferation Dye eFluor 450, VioBlue Qdot 525, BV510, BD Horizon V500, Fixable Viability Dye eFluor 506, VioGreen Fluorochromes FITC, Alexa 488, eGFP, CFSE, Dye Cycle Green PE, Cy3, Sytox-Orange, Dye Cycle Orange PI, PE-Texas Red DRAQ7 Cy5, DRAQ5, DRAQ7, Dye Cycle Ruby DRAQ7, Dye Cycle Ruby Fig. 2.1.3 This table gives examples of the fluorochromes most commonly used on LSRFortessa A, showing which laser they are excited by and which filter should be used to collect the signal. All of these fluorochromes can be used with the standard LSRFortessa A filter arrangement as shown in section 1.3. 37 Laser 488nm Blue 561nm YellowGreen 355nm UV 639nm Red 405nm Violet Collection Filter (bandpass) Diva Parameter Name 530/30 530/30 blue 710/50 710/50 blue 586/15 586/15 yellow PE, Alexa Fluor 555 610/20 610/20 yellow 670/30 670/30 yellow PI, PE-Texas Red, M-Cherry, M-RFPMars PE-Cy5 (also “Tricolour” or “Cychrome”), 7AAD, E2-Crimson, M-Plum 710/50 710/50 yellow 780/60 780/60 yellow 450/50 450/50 UV DAPI, Hoechst (blue), Celltrace Violet 530/30 530/30 UV 670/14 670/14 red PI APC, Alexa 633, Alexa 647, TO-PRO-3, Cy5, DRAQ5, DRAQ7, Dye Cycle Ruby 730/45 730/45 red APC-Cy5.5, 780/60 780/60 red 450/50 450/50 violet 525/50 525/50 violet 540/30 540/30 violet 586/15 586/15 violet BV570 610/20 610/20 violet BV605 670/30 670/30 violet BV650 Fluorochromes FITC, Alexa 488, eGFP, CFSE, Dye Cycle Green PerCP, PerCP-Cy5.5, PerCP-eFluor710, DRAQ5, Dye Cycle Ruby DRAQ7 PE-Cy7, DRAQ7 PE-Cy5.5, DRAQ7, Dye Cycle Ruby, APC-Cy7, Alexa Fluor 750, DRAQ7, Fixable Viability Dye eFluor 780 Alexa 405, CFP, Pacific Blue, DAPI, Hoechst (blue), Celltrace Violet, Dye Cycle Violet, BV421, VPD450, Calcein Violet 450, Fixable Viability Dye eFluor 450, Cell Proliferation Dye eFluor 450, VioBlue Qdot 525, BV510, BD Horizon V500, Fixable Viability Dye eFluor 506, VioGreen Fig. 2.1.4 This table gives examples of the fluorochromes most commonly used on LSRFortessa B and C, showing which laser they are excited by and which filter should be used to collect the signal. All of these fluorochromes can be used with the standard LSRFortessa B filter arrangement as shown in section 1.3. 38 *Which LSR? * The configurations of the four LSRs are different (see section 1.3); this may mean that one machine is more appropriate than the other for a specific experiment. These applications can only be run on a specific LSR: -LSRIIA or LSRFortessas: Hoechst side population and Indo-1 calcium flux. -LSRIIB or LSRFortessas: RFP and variants. These applications may be better using the yellow laser on LSRB or LSRFortessas: -Any experiment using PE and PE tandems (there is less autofluorescence and better S/N ratio). -Experiments combining GFP or CFSE with PE. Note: For PerCP or PerCP variants (PerCP-Cy5.5 for example), use 488 (blue) laser excitation NOT the 561nm (yellow). 2.2 Fluorochrome Selection There are many factors that will affect the choice of fluorochromes for an experiment. Please speak to a member of the FACS Lab when designing a new experiment and before purchasing expensive reagents! Viability dyes The presence of dead and dying cells in a sample can have a detrimental effect on the analysis and interpretation of fluorescence data. Dead cells may lose markers present on/in live cells and can become autofluorescent, i.e. giving a positive fluorescent signal although no fluorochrome is present. They may also non-specifically take up antibodies, generating artefacts that are difficult to exclude when you wish to analyse your data. These problems are easily overcome by using a viability dye, sometimes referred to as a live/dead dye. These dyes are excluded by the intact membranes of live cells, but can enter dead and dying cells. Therefore, gating on the population negative for the viability dye easily excludes dead cells. 39 Dye Propidium iodide (PI) 7AAD TO-PRO-3 Excited by: Blue Laser/Yellow laser Blue Laser /Yellow laser Red Laser Collected in (LSRIIA/B - LSRFortessa A/B): 610/20bp / 575/26bp 660/20bp / 610/20bp 660/20bp DAPI DRAQ7 UV laser / Violet laser Red laser 440/40bp / 450/40bp / 450/50bp 670/14bp / 660/20bp / 710/50 bp / 730/45bp /780/60bp Fig. 2.2: Examples of commonly used viability dyes. *Tips and Tricks* The dyes mentioned above can only be used in unfixed samples. In fixed samples the dye would enter all cells. The most popular viability dye for the LSRs is DAPI. This is because it is excited by the violet and UV lasers and rarely overlaps with another fluorochrome in the same experiment. Fixable LIVE/DEAD amine reactive dyes If you wish to be able to fix your sample, you may try using the amine-reactive dyes from Molecular Probes, Invitrogen. In membrane-compromised cells, the dye has access to free amines both in the cell interior and on the cell surface; in viable cells, the dye can react only with cell-surface amines, generating lower levels of fluorescence. Up to 50fold difference in intensity can be obtained between live and dead cells, and the staining is preserved following formaldehyde fixation. Availability This may be the major factor limiting your choice of fluorochrome combination. Although a selection of different directly conjugated fluorochromes are available for commonly used antibodies such as CD4, this may not be the case with less common antibodies. In many cases it is necessary to use a two-step staining process involving primary and secondary antibodies and in this case great care must be taken to avoid non-specific staining and cross-reaction between antibodies. Please ask a member of the FACS Lab if you need further advice on this. 40 Physical Characteristics Physical characteristics of fluorochromes may influence their selection, for example, the size of different fluorochromes varies considerably and when doing intracellular staining it may be necessary to select a smaller fluorochrome. The brightness (quantum efficiency) of different fluorochromes also varies. For multi-colour staining it is advisable to use the brightest fluorochromes to identify the dimmest or least expressed markers. 2.3 Combining fluorochromes Spectral overlap With all flow cytometers, problems can be encountered when using multiple fluorochromes because the emission of each fluorochrome can “spill over” into the detectors of others. This problem increases with the number of fluorochromes used and is therefore especially relevant to the LSRs. Fig. 2.3.1 Emission spectrum of two commonly used fluorochromes. FITC (green line) is detected with a 530/30 filter and PE (purple line) is detected with a 575/26 filter. These filters are chosen to capture the maximum emission of their respective fluorochromes. However, some FITC emission can be seen spilling over into the 575/26 spectral region and dim PE emission can be seen in the 530/30 and 660/20 (PE-Cy5) regions. Problem pairs Because of the spectral overlap, certain combinations of fluorochromes can be problematic, in many cases this will depend on how bright the fluorochromes are. For example, combining a very bright GFP with dim PE could be difficult, but a dimmer GFP 41 signal may not cause problems. Other combinations will always be potentially difficult, an example of this is given in Fig. 2.3.2 below. Fig. 2.3.2 Spectral overlap of PE-Cy5 (excitation dark green line, emission light green line) and APC (excitation dark purple line, emission light purple line). PE-Cy5 is optimally excited by the 488 laser and APC by the red laser, however some excitation of the Cy5 component of PE-Cy5 can occur with the red laser. Both fluorochromes emit in the 660/20nm region so it can be very difficult to separate the red excited PE-Cy5 signal from the APC; take care in checking the voltages for APC and PE-Cy5 to obtain a good balance of signals and still be able to compensate PE-Cy5 out of the APC channel. Example combinations Fig. 2.3.3 2-colour staining, using FITC (blue line) and APC (purple line). The emission of these two fluorochromes is well separated resulting in very little spectral overlap. If using live (unfixed) cells DAPI should be added to this combination as a viability dye. 42 Fig. 2.3.4 3-colour staining, using FITC, APC and PE-Cy7 (yellow line). The emission of PE-Cy7 is higher than that of both FITC and APC so there is little spectral overlap. If using live (unfixed) cells DAPI should be added to this combination as a viability dye. *Useful websites: * The following websites all enable you to view the excitation and emission of various fluorochromes and compare different fluorochromes to determine the likelihood and severity of spectral overlap. http://www.bdbiosciences.com/spectra/ http://probes.invitrogen.com/resources/spectraviewer/ http://www.ebioscience.com/resources/fluorplan-spectra-viewer.htm 43 2.4 Compensation Compensation is used to control the effects of spectral overlap. The aim of compensation is to account for spectral overlap of fluorochromes into adjacent detectors. This is very important for the reasons explained in section 2.3 above. The LSR has a number of features that help to improve the ease and accuracy of compensation. Having the correct controls is also very important. For further explanation on the principles of Compensation, please speak to any member of the FACS Lab. Controls For each of the fluorochromes used in the test samples, a single colour positive control is needed for compensation. Each control should ideally be made up of a population of negative or unstained cells and a population of single colour positive cells. Both the positive and negative cells should have the same level of background autofluorescence, ideally this will mean ensuring that they are the same type of cell, e.g. lymphocytes. Ideally the positive and negative populations in the controls should both be large, well defined and well separated. However, if you are looking for a rare population or have a very limited number of cells this might not be possible. There are three ways to overcome this problem: 1. Use different cells for your compensation controls - e.g. if the test samples are primary tissue in limited supply, find a cell line that expresses the markers of interest and use these for the compensation. 2. Use a different antibody conjugated to the same fluorochrome – if a marker of interest is rare or possibly absent in the control cells, it is better to use a different antibody, directed against a more common marker, but carrying the same fluorochrome as will be used in the test. CD45 antibodies are often used for this purpose when studying rare lymphocyte markers. It is important to note however that the exact same fluorochrome should be used, e.g. GFP and FITC may both emit a signal detected in the 530/30 channel, but they are not the same fluorochrome. Tandems are trickier due to variations in chemical conjugation: if a tandem conjugate is being used, e.g. PE-Cy5, the same batch of tandem should be used. 44 However, it is safer still to use a cell line highly expressing the marker and label with the same antibody you are planning to use. 3. Use compensation beads: If the marker of interest is rare or absent on/in the control cells, you can still use the antibody if you substitute the cells for “comp beads”. Comp beads are polystyrene microparticles, available as Anti-Rat or Antimouse Ig, kappa. They bind to rat or mouse light chain Ig antibodies (check which ones are right for your antibodies). Beads with no binding capacity are also available and can be used as a negative control to determine background fluorescence, ensuring that clear positive and negative populations are present in all compensation controls. Compensation Beads To prepare the controls the antibodies to be used are incubated with the beads as follows: Combine 100µl staining buffer + one drop negative beads + one drop of Comp beads (Anti-mouse or Anti-rat) + antibody (at the concentration used in the experiment). Incubate for 30min (RT in the dark). Wash with 2ml of staining buffer and resuspend compensation controls in 500µl staining buffer and run them on the flow cytometer. Cells can also be added to the compensation control tubes to check for autofluorescence from your cells and to help establish your baseline voltages. A B Fig. 2.4.1 A Scatter plot showing a large scatter gate that includes the Comp beads and cells. B APC-Alexa750+ve beads can clearly be distinguished from unstained particles and used for compensation (plot showing only scatter gated events). 45 The Bi-exponential Display The bi-exponential display is a feature of digital cytometers. It allows you to see below 0 on a log axis. This means that any events that would normally appear crushed on the axis will be visible; the bi-exponential display does not change the data, it merely transforms it to present it in a visually understandable way. This makes it much easier to do compensation, and is particularly useful for seeing if a sample has been over compensated, as shown below. A C B D E Fig. 2.4.2 A shows an uncompensated sample on the standard LSR 5 log display, B shows the same uncompensated sample viewed using the bi-exponential display. C shows the sample once compensation has been applied “by eye” without using the biexponential display, the centre of the positive population appears to be roughly in line with the centre of the negative population. D shows what the compensation used in C looks like when viewed with the bi-exponential display. It can clearly be seen that the sample is actually over compensated, with the positive events below the negatives. E shows correct compensation, as viewed on the bi-exponential display. 46 The bi-exponential display function can be selected for individual axes and plots. To switch it on: select the plot, then in the Instrument window check the boxes for the relevant axes. Note: The scale of the axis will automatically adjust to give the best display of all current events. During sample acquisition this can lead to large blank areas appearing in the negative region, however this should be corrected when all the collected data is displayed for analysis. You may find it best to turn off the bi-exponential display when acquiring data. Compensation – median fluorescence technique It is recommended that you use the statistics functions in Diva to calculate the exact amount of compensation required. This is done by drawing gates around the positive and negative populations and comparing the median fluorescence of the two. For example, if you are compensating FITC out of PE, the median fluorescence of the double negative + - population will be X. The FITC PE cells have no PE and should therefore give the same median PE signal as the double negative (X). However, because of the spectral overlap + - of the FITC signal into the PE channel, the FITC PE population has an increased median PE fluorescence of Y. To correctly compensate this sample you simply increase the amount of compensation applied until Y=X. You just have to imagine a line in the centre of the positive population that needs to be lined up with the centre of the negative population. You will be taken through the process of compensation in section 3. *Tips and Tricks* If you are using any fluorochrome pairs that could be difficult to compensate, e.g. PECy5 with APC, it is a good idea to run these first and do a rough check on the compensation by eye before saving all the samples. This prevents you from having to re-record all your controls if you need to change the voltages on one to help compensation. After checking the compensation reset it to zero before recording the sample. 47 Copying the spectral overlap Once you have calculated your compensation you will need to apply it to all future tubes. This can be done by just writing down the compensation applied to each tube and typing it in for the sample tube, or by using the copy spectral overlap function. For the latter, once you have set the compensation for a control, expand the tube and right click on instrument settings. Select “Copy spectral overlap” then right click on the instrument settings for the experiment and select “Paste spectral overlap”. Repeat this procedure after compensating each control. Having completed the compensation, click Next; the new tube and all subsequent tubes should contain all the compensation and be ready for you to run your samples. There is an example experiment in section 3 where you can try this out. Automatic compensation As the number of fluorescence parameters in an experiment increases, compensation becomes increasingly difficult to set manually. For a six-colour experiment, 30 spectral overlap values need to be adjusted, and for an eight-colour experiment, 56 values need to be adjusted. The process of manually correcting spectral overlap values can take several hours and is very difficult to set accurately. The Compensation Setup feature in BD FACSDiva software is designed to automatically calculate spectral overlap values for an experiment, saving time and eliminating the inaccuracies introduced with manual compensation. We recommend that for any experiment with three or more fluorochromes compensation should be set using the automatic feature. Compensation Setup is designed to work with single-stained controls. These controls can consist of single-stained cells or capture beads. An unstained control is required as well, in a separate tube or in the same tube as the single-stained controls. As mentioned, you can find more details about the process of compensation in section 3. 48 We will now run an example experiment with cells, FITC and PE beads. Open the software (double click the FACSDiva icon on the desk top). There is no password, just hit “Enter” at the password window. *Checklist before starting experiment* Has the LSR been switched on long enough for the lasers to warm up? Is the sheath tank full and waste tank empty? Is the standard configuration of optical filters in place? Is it clean (who used it last? when was it used last?) - Worth giving a few Primes, checking cleanliness with dH20 3.1 The Browser, Worksheet, Inspector and Acquisition windows 1. Browser 2. Cytometer 5. Worksheet 4. Acquisition Dashboard 3. Inspector Fig. 3.1.1 Diva v6 software windows 50 1. The Browser: where you can make new folders and experiments, and access or delete acquired experimental data. Fig. 3.1.2 The Browser window 2. Cytometer: controls all instrument related electronic functions. This window cannot be accessed unless an experiment is open, and the arrow to the left of a tube in the browser is selected (green). Instrument connection status is shown in the footer of the window. There are a number of tabs below the header including Parameters (equivalent to Detectors and Amps in CellQuest) and Compensation; these will be discussed further in later sections. 3. Inspector: Each Diva window has information that can be edited/changed via the “Inspector”, which is accessed by making a window active. 4. Acquisition Dashboard: is the main panel for collecting data. Here you can start acquiring/recording data, and add storage and stopping gates for acquisition. Changes to the number of events displayed and recorded can also be made here. Data recording progress is shown as a coloured bar; when completed, click the NEXT TUBE icon to move on to the next sample. The sample flow rate (events/sec), aborted events and time elapsed is also shown. 5. The Worksheet: where you can make plots and histograms. There are two types of worksheets but we only use the Global type worksheet to acquire data. Data analysis is performed using FlowJo. 51 *Trouble-shooting: Connectivity Problems* The Instrument/Cytometer window displays LSR instrument connection status to the workstation software. If it seems to have trouble connecting (takes too long, or shows “Instrument Disconnected”) try re-starting the computer. If this doesn’t help, try switching off BOTH the computer and LSR, then switching back on. 3.2 Example experiment The rest of this section will take you through the basics of using Diva by running an example experiment using cells and fluorescent beads. 52 Sample tubes: *DAPI will be included in all tubes as a viability dye* Tube 1: unlabelled cells + unlabelled beads (Unlabelled control) Tube 2: unlabelled cells + FITC beads (FITC control) Tube 3: unlabelled cells + PE beads (PE control) Tube 4: unlabelled cells and both beads (the sample) Left Click on Administrator at the top of the tree of folders (to highlight the location). Left Click on the folder icon to create a new folder. Re-name the new folder with your NAME (it should appear at the bottom of the tree under Administrator). Click on the New Experiment icon (brown book) to make a new experiment for today. Click on the New Specimen icon (syringe) to create a new specimen. Rename the specimen as Cells+Beads. Rename the first tubeUnlabelled control. Highlight today’s experiment; in the Inspector window select Experiment tab; check 5log axis box. Highlight Tube_001: Instrument window: Select the Parameters tab We will be using FSC, SSC, 530/30blue (FITC), 575/25blue (PE) and 440/40violet (DAPI). Delete the other parameters. *Renaming things*: often done by right clicking on the item to access a menu. TIP: Include as much information as possible in your experiment (i.e. date, reagents, tube names etc.) as this is saved in the FCS files and is accessible during analysis. The new folder should appear under (in) the folder with your name; re-name today’s folder with your INITIALS, DATE, and LSRIIA, LSRIIB or FORTESSA (e.g. MW010212lSRIIA). If you are using two cell lines, you could make a separate specimen for the second, or if you are looking at samples from several patients/animals each could be a specimen. Give the specimen a short name. If you expand the browser at the level of specimen (click on “+” sign, left of the specimen) you will see a tube icon Tube_001; you may rename this if you wish. The Inspector window contains a number of tabs for a variety of functions, some useful (including re-labelling parameters with names), and some redundant (e.g. acquisition and storage information - which can be changed through the acquisition controls). The Instrument window- further considerations Minimising file size: Data files can be large, so one important thing to remember to do before saving data is to delete any unused parameters/colours (click the bullet on the left of the parameter name and click the Delete button on the bottom left) - you can select several by clicking the first one, holding shift and clicking the bottom bullet point. Owing to a bug in Diva: it is necessary to leave at least two parameters from the blue laser in the Instrument window. 53 Instrument Window: In the Parameters Tab set parameters to log or linear as appropriate. Instrument Window: In the Parameters tab: Tick the W box for both FSC and SSC to permit doublet discrimination. Instrument Window: Threshold tab: set FSC and SSC to 5000 (we will be keeping it low since we are running beads as well as cells). Inspector Window (highlight Experiment) Label the fluorescence channels with the fluorochromes you are using. Draw the following plots and histograms. Plot 1: FSC-A vs. DAPI Plot 2: FSC-A vs. SSC-A Plot 3 & 4: FSC-A vs. SSCW and/or FSC-A vs. FSC-W Plot 5: FITC vs. PE Histogram1:FITC Histogram 2: PE Under the Parameters tab: generally, FSC/SSC will be linear and fluorescence will be log. Digital data are stored as linear so you can actually change between LOG/LIN after acquisition (however, if you wish to do this you must ensure that all events are on scale for both log and lin displays because you can’t adjust the voltages after acquisition). Exceptions to using LOG for fluorescence include DNA and other measurements where differences between +/- are quite small; when doing BrdU analysis it is also sometimes worth trying a linear scale for better resolution. Doublet Discrimination: A, H and W can be measured for all parameters. In general, use A as it represents the total fluorescence. W is analogous to time spent passing through the laser beam so we can use this to discriminate against cells stuck together. W from either scatter parameter can be used (or both if cells are sticky and extra stringency is required). Doublet discrimination is particularly important in DNA analysis where a different strategy should be used; please see a member of staff about this. Threshold values: These are channel number values used to tell the computer what to ignore as noise/debris (i.e. events below a certain channel number will not be displayed or recorded). We tend to use both FSC and SSC. In analysis these can be set relatively high (e.g. 10-15K). Use of threshold aids processing time as fewer events are analysed. For samples with a lot of debris this is particularly useful. In the Inspector window, labels tab: 530/30 blue: label with FITC 575/25 blue: PE 440/40 UV: DAPI Plots: Using the dot plot icon, drag and let go, or click on the workspace to get a standard size plot. Parameters displayed can be changed by clicking on the name of the parameter in the plot axis (drop down menu). 54 Working with Plots, Regions and Gates There are 3 types of plots available on the LSR, the icons used to select these are circled in red below. These will allow you to draw dot plots, contour plots and histograms respectively. Fig. 3.2.1 Icons at top of the worksheet window used for drawing plots (red) and regions (green). Plot drawing icons (left to right) Dot plot, Contour plot, Histogram. Gate Drawing icons (left to right): Autopolygon, Snap-To gate, polygon, Rectangle, Quadrant, Interval, Autointerval, Snap-To interval. Plots Dot plots and contour plots are both 2 dimensional displays allowing you to compare 2 different parameters on the same events. For example, this will allow you to see which FITC positive events are also APC positive. An example of a dot plot is shown in fig. 3.2.3.B. Dot plots represent each event as a dot, where a population of events all have similar staining they will form a distinct cluster of dots. A contour plot shows the population as a set of contour lines. Lines are drawn as a percentage of the total events; the area between each line has the same percentage of the total events so that the bigger the population is, the closer together the lines will be. This is particularly useful to visualize the distributions of populations that are very close together, or overlapping. 55 A1 A2 B1 B2 Figure 3.2.2 A1 shows a sample where two populations (P2 and P3) of interest are identified by drawing gates on a dot plot. In this case, the gates are drawn so that they are touching. A2 shows the same data presented on a contour plot: it can be observed that there is likely to be slight overlap of P2 and P3. When the gates are re-drawn with some separation, as shown in B1, the separation of P2 and P3 is increased. Histograms are single parameter presentations of data and should be used with careful consideration as they have some significant limitations, meaning that they are not suitable for all types of experiments. One commonly observed error is using histograms to gate positive events for analysis whilst not taking into account the contribution of autofluorescence to the signal. This is particularly common for people doing single colour experiments such as FITC and PE, as cells often have “greenish” autofluorescence. Fig. 3.2.3 demonstrates clearly the danger in gating off a histogram. 56 A B Fig. 3.2.3 A common error of using a histogram instead of a 2-parameter plot to observe GFP: A shows an interval region on a histogram used to select events that are GFP+ve, which seems to be around 18% of the total events. B is the same population of cells shown on a 2-parameter plot, with regions drawn to show that, of the 18% selected by the interval region in A, around 15% are autofluorescent, and only 2.5% are truly GFP+ve. Regions and Gates To select a population of cells for further analysis or to obtain statistics on the population a “region” can be drawn around it. There are a number of tools for drawing regions, the icons for these were shown in fig. 3.2.1. Region Name Used on plot Use: type(s) Autopolygon Dot/contour To automatically draw a polygon around a population. Snap-To gate Dot/contour To automatically draw a polygon a population. The shape of the polygon will adjust as the population changes. Polygon Dot/contour To manually draw a polygon region around a population. Rectangle Dot/contour To manually draw a rectangle or square region around a population. Quadrant Dot/contour To divide the plot into “quarters” (not necessarily of equal size), each of which is a region. Arms may be hinged (so that not 90 degrees). Interval Histogram To manually select a region on a histogram. Autointerval Histogram To automatically select a region on a histogram. Snap-To interval Histogram To automatically select a region on a histogram, the region will adjust as the population changes. Fig. 3.2.4 Types of regions/gating tools available for specifying populations of interest. 57 Combining Regions Once you have drawn a region it will be given a name, e.g. P1 (meaning population 1), you can then select to show only the cells within this region on a new plot. To do this right click on the new plot and select Show Populations→P1. If you draw a new region on the new P1 gated plot, this region will become P2 and will be a “daughter” of P1 (the parent). Regions drawn on the same plot will be siblings and not linked to each other. You can view the organisation of your regions by opening the population Hierarchy window. Do this by right clicking on the margin of any plot and selecting view → population hierarchy. The word “gate” is often used to describe a region. However, a gate can also be a combination of parent and daughter regions and refers to which events would be displayed on a plot gated on Pn (when Show Populations→Pn was used). Where a population has no parent (e.g. P1) the single region defining it can be referred to as a gate because every event within it would be displayed. Where the region has a parent (e.g.P2) only cells that fell within the daughter region and all parent regions would be displayed. *More Points To consider about gating* Regions drawn on negative populations (excluding scatter and viability gates) need to be pushed to the very edge or slightly off the axis to avoid excluding negative events that are crushed up against the axis. Drawing regions very close together may result in unexpectedly overlapping populations even if the regions themselves were not overlapping. This is caused by slight statistical variations in the positioning of the events on the plot. It is useful to obtain a general picture of how well separated the populations are by looking at a contour plot Data follows a biological distribution that does not always fit with rigid gating. Be consistent and logical in how you draw regions, to minimise subjective bias for inter-experiment comparison. Quadrants/ rectangles may not necessarily fit plots with four compensated populations (-/-, +/-, -/+ and +/+), particularly when there is a range of antigen expression involved, using these straight line regions can mean part of a population ends up in the wrong region. It’s worth considering using a polygon or altering the quadrant so that the sections are different sizes/shapes (this is thoroughly discussed in the Data Analysis section of Howard Shapiro’s Practical Flow Cytometry 4th Ed. 58 On LSR: hit RUN in LO Remove tube of H20 Install Tube 1 onto SIP (UNLABELLED CONTROL). *TIP* Quickly CHECK your sample before running: Flick to mix, make sure NO CLUMPS ARE VISIBLE! Pipette if necessary. Acquisition Controls: Click Acquire. Show 5000 events on plots. *TIP* Monitor the threshold rate in the Acquisition Status window to be sure that your sample is running and hasn’t clogged. Adjust scatter PMTs to see both cells and bead population on the scatter plot. Scatter: We want to be able to see the whole population of interest on our scatter plot, such that we don’t lose the edges, and so that we can just see the debris, keeping it in the bottom end of the scale. Where this is actually placed is subjective, and dependent on what you are doing. For samples where several cell populations are present such as in whole blood, if you were interested in looking at lymphocytes, you may choose to increase scatter to bring these out further on the plot, but this will push the granulocytes off. If you were interested in apoptosis, you may choose to increase scatter to expand the area of interest- where the apoptotic bodies with lower FSC may fall. Select a gating tool: Draw a gate (P1) around the DAPI negative populations. This will include cells and beads. Population Hierarchy: Open by right clicking on any plot and selecting show population hierarchy. Right click on the scatter plot-->Show Population P1. Select Gating tool: Draw a large second gate (P2) to include the cells and beads. Show P2 on the FSC-A vs. SSC-W plot and draw a gate (P3) around the tight singlet population of lowest SSC-W. Show P3 on the rest of the plots and histograms. At this point we can start running the samples and adjusting the voltages for the detectors. Remove the tube of water and place the machine into RUN. Make sure the speed is LO. Under Acquisition Controls choose to display 5000 events (so the plots refresh quickly). The scatter plot will now only display events from P1. Notice that the software organises this second gate as a daughter of P1 because P1 was shown first on the plot before P2 was drawn. Notice that the singlet population is tighter on the FSC-A vs. SSC-W than vs. FSC-W. As mentioned previously, sometimes even after gating singlets on FSC-A vs. SSC-W there are still some clumps present that can be seen on the FSC-A vs. FSC-W plot. 59 Fig. 3.2.5 An example of the FACSDiva Worksheet showing the plots created in the example experiment. This is data generated by running the unlabelled control. 60 Once the scatter looks ok and the basic gates have been drawn, the fluorescence plots will look cleaner due to the exclusion of dead and dying cells, and we can adjust the PMT voltages. 10 5 10 <780/60 blue 5-A>: PE-Cy7 <780/60 blue 5-A>: PE-Cy7 10 5 10 4 27.2 10 10 37.8 10 3 10 2 0 4 3 2 0 0.83 0 A 2 3 0.53 8.55 26.5 4 10 10 10 <675/20 blue 4-A>: PerCP 2 0 5 10 Ungated B 4000 10 3000 SSC-A 95.1 2000 3 2000 1000 48.5 2 FSC-W <450/40 violet1-A>: DAPI 10 5 10 4000 3000 4 4 Singlets 10 5 10 3 10 10 10 <675/20 blue 4-A>: PerCP 1000 99.2 0 0 0 C 1000 Ungated 2000 FSC-A 3000 4000 0 1000 D Live 2000 FSC-A 3000 4000 E 0 0 1000 2000 FSC-A 3000 4000 Scatter Fig. 3.2.6 The importance of correct gating: A shows fluorescence data from all events acquired in a sample from a two-colour experiment. B shows the fluorescence of the same sample following gating to exclude events that are dead cells and debris. The gating strategy was as follows: the DAPI negative population was selected first (C); this live population was then shown on a scatter plot (D) and gated to exclude debris and obvious clumps. The live+scatter gated events were then further gated to exclude doublets by excluding events with a high FSC-W signal (E). The difference between the two plots is very clear especially when looking at the double negative population. 61 Baseline PMT Voltages The optimum voltage set up is one where the signal from positively stained populations can be best separated from negative/unstained populations, with negative populations being positioned above the electronic noise and at the same time allowing dim populations to be distinguishable above background. BD published an Application Note (Establishing Optimum Baseline PMT Gains to Maximise Resolution on BD Biosciences Digital Flow Cytometers) which can be found on their website. This protocol basically involves running fluorescent beads and plotting %CV (a measure of peak distribution defined as standard deviation of the peak divided by the mean channel number of the peak, times 100, reflecting how “fat” a peak is) at different PMT voltages to find the minimum voltage where fluorescence %CVs are at their tightest (values start to plateau) and hence the minimum voltage for optimal resolution and sensitivity. http://www.bdbiosciences.com/pharmingen/products/display_product.php?keyID=126 In practice, one finds that when voltages are set such that positive signals and negative signals are separated and voltages are balanced against other fluorochromes used (so that spill-over can be compensated out) they will be above the baseline anyway. 62 Compensation UNLABELLED CONTROL: In general as we go up the spectrum to the red end we need to increase the voltage (fewer photoelectrons produced so need to amplify more) Adjust PMT voltages: Place the signal from the negative cells in the first two logs of your plots. Click Restart in the Acquisition controls. Check that the separation for PE is the best it can be (play with voltage). Run PE control tube: Check fluorescence for PE. Do the same with FITC control but ensure FITC voltage> PE voltage. *TIPS AND TRICKS* By convention, we are used to running controls starting with the shortest emission wavelength fluorochrome (‘greenish’). It is actually often more efficient to start at the other end- with the ‘far red’, long wavelength fluorochromes. These have the least energy and often require a higher PMT voltage. Once these are set (+ve’s and – ve’s as well separated as possible), then you can work your way back towards FITC. Working the other way you will often find you have to readjust FITC, PE etc. FITC +ve PE+ve Fig. 3.2.7 The PE control is shown on the left, on a FITC vs. PE plot. In this particular example very little compensation will be needed to remove PE out of FITC. To the right is the FITC control on bi-exponential display; considerable spill over into the PE channel can be seen. 63 Acquisition Controls: Set a Stopping Gate on P3 for data acquisition. Record 20K events of the unlabelled control. Click Next icon for a new tube; re-label as FITC control. Record 20K events. Repeat for PE. Recording data: Once you have checked that the single colour controls look ok, you are ready to record data. It is a good idea to set a Stopping gate to stop acquisition when the desired number of relevant events has been seen. The most basic level is to set this to P3 (for live, single cells). Remember if you have a small volume to not let the machine run out of sample/fluid. This will introduce air into the system (e.g. if you are running near to the bottom of your sample, you need to be QUICK in hitting Record to save, and getting that tube OFF. Ideally, single colour controls for Fluorescence Compensation have both a negative and a bright positive population, which are well separated. How many cells do I need to record? To be able to draw any statistically significant conclusions, we need to look at the actual final population of interest. In experiments such as immuno-phenotyping, the final population may be very rare and the last in line of a number of hierarchically structured gates. Rare events follow a Poisson distribution, so we use the following formula to calculate CV: CV= (1/√n) x 100 (CV = standard deviation, n = cell number) We can see that for a CV<10 (ideally CV=1) we need at least 10K cells. This is especially relevant with rare populations and it is the actual number of these that is important – it should be at least 100 relevant events (i.e. after correcting for background). This can be tricky as in theory both control and test samples should be the same size! Right click on any plot and open the Statistics Window. Right click on the statistics view itself to edit information displayed in three different tabs, Header (information about the experiment to be displayedmost of which can be un-ticked), Population (gates can be de-selected if statistics not needed) and Statistics (a table of all parameters with possible statistics which can be selected for display). 64 Edit the Statistics window. Load the data from the FITC control by selecting the arrow to the left of the tube. Draw gates around the positive and negative populations. Select a fluorescence plot and turn on the biexponential display; this is the plot we will use for compensation. Compensate the FITC out of the PE channel. Remove all the options under the Header tab. Under the populations tab show only P4 and P5. Under the Statistics tab tick the box for Median and remove any other statistics. On a 2-parameter plot of FITC vs. PE, gate both the negative (this will be P4) and positive (this will be P5) populations. The FITC+ population will also appear to be slightly PE+, i.e. curving upwards away from the xaxis. Turn on bi-exponential display (by left clicking on the plot once). In the Inspector window check the boxes to show bi-exponential display for both x and y-axes. First, we want to remove the percentage of FITC signal that is leaking into PE channel, i.e. PE- %FITC. On the LSR this will be 575/25blue -530/30blue. The Compensation tab can be found in the Instrument window. The Compensation Matrix allows full, inter-laser compensation of all parameters against each other. The matrix is set up with the first fluorochrome (shortest wavelength PMT, equivalent to “FL1”) to be subtracted from all the other parameters, then the second and so forth down the column. Use of the bi-exponential display while adjusting compensation is advised to properly observe compensated parameters. Compensation can be added by typing a value or using the slide bar in the right column until the single colour control no longer appears positive in the second channel (median fluorescence values of the positive and negative populations are equal). *Tips and Tricks* When compensating a single colour control one can think: What is in the tube is in the middle column, and the channels that you are removing that fluorochrome from are in the left column. 65 Fig 3.2.8 The workspace above shows FITC compensated out of the PE channel (575/26blue – 530/30blue) using median values. 16.7% compensation is required to bring the median PE fluorescence of the FITC population (P5) to match those of the negatives (P4). 66 Right click on the FITC tube. Select Copy spectral overlap. Right click on the PE tube. Select Paste with zeros. Load up the next control: PE control. Compensate PE out of FITC. On the fluorescence compensation plot, place PE on the x axis, and FITC on the y-axis. If we were using more colours (not counting DAPI as it does not leak into FITC or PE channel and we are selecting cells which are DAPI negative), one would change the y-axis parameter to the next channel and repeat the process of finding the medians etc, for all the other colours. *Putting the control colour on the x-axis and changing the y-axis as you compensate x out is methodical but everyone has different ways of doing this, you may find a way that you prefer instead. When all controls have been systematically compensated and the spectral overlap copied and pasted, the bi-exponential display can be switched off (not critical but helps with processing if you are collecting huge numbers of cells). Draw gates for the FITC and PE single positive populations. Using the PE control data, draw the FITC+ gate; then use the FITC control to draw the PE+ gate. For more information on this see the section on FMOs in the fluorochromes section. Click Next, and re-label the new tube. Apply compensation and record sample. Gate the cell population. Show that these are FITC and PE negative. Also gate the smaller bead population. Show these are FITC+ and PE+ 67 Automatic Compensation Compensation cannot be calculated if PMT voltage settings are changed while recording compensation tubes. All tubes must be recorded with the same PMT voltages. The optimum voltage set up should be made before creating the compensation controls. To do that, you should create some dot-plots on the global sheet that allow you to see the spillover between the detectors. If you need to re-adjust the PMT voltage after creating the compensation controls you will need to record all compensation tubes again. Create the compensation controls Go to Experiment > Compensation Setup > Create Compensation Controls A list of the parameters that will be used as compensation controls is created (Fig. 3.2.9). Check if you have in the list all the desired parameters (use options Add or Delete if necessary) Fig. 3.2.9 Leave the Include separate unstained control tube/well checkbox selected (highlighted with the red box) when you are running unstained sample as one of your compensation controls. Deselect the checkbox when you are including unstained sample in each of your stained control tubes or wells. When you click OK, a new specimen named Compensation added to Browser, Controls (Fig. 3.2.10) the open experiment with tubes for each in is the specified parameter. A normal worksheet is created for each single colour control, along with an Fig. 3.2.10 unstained control worksheet if the checkbox was selected. The experiment’s cytometer settings are copied to the controls, all compensation values are set to zero, and the Enable Compensation checkbox is deselected. 68 Install the first stained tube onto the cytometer and start acquisition. Adjust the pre-defined P1 gate to your population (Fig. 3.2.11) and press record. When recording is finished install the next stained control tube onto the SIP. After you create appropriate compensation controls, you need to verify gates, before calculate compensation. Adjust the gate around the fluorescence-positive population on the histogram (gate P2 on Fig. 3.2.11). Fig. 3.2.11 Fluorescence-negative populations can be defined using the Unstained Control Tube. If negative populations had not been already defined, you could define them by creating an Interval Gate around the negative population for each Stained Control Tube (Gate P3 on Fig. 3.2.11). Generate the compensation matrix After data has been recorded and gates have been adjusted, you are ready to calculate compensation. Go to Experiment > Compensation Setup > Calculate Compensation (Fig.3.2.12) The software calculates the overlap as the median fluorescence intensity (MFI) of the positive stained control minus the MFI of the negative stained control for each control in all channels. If there is a gated unstained population in the Unstained Control tube and a gated unstained population in the Stained Control tube, the population in the Stained Control tube will be used in the calculation. Fig. 3.2.12 69 If the compensation calculation is successful, a dialog appears prompting you for the name for the compensation setup. Enter a name, and click: o Link & Save—Links cytometer settings to the experiment and save the setup to the compensation setup catalog. o Apply Only—Applies the cytometer settings to the experiment without saving the settings to the compensation setup catalog. o Cancel—Dismisses the dialog without saving the setup. In the link and save option you cannot change either the fluorescence PMT voltages or the compensation values in a tube without data. You can only adjust the scatter PMT voltages. After recording your sample the compensation values are available for manual adjustments. In the apply only option all the PMT voltages and compensation values can be changed before recording the sample. However, do not forget that if you change the fluorescence voltages, the compensation values are no longer valid. You have to record again all the compensation controls with the same voltages. Record Samples Select Tube_1 o To see the dot plots created in the beginning you have to return to the Global Worksheet, by clicking the icon highlighted in red on Fig. 3.2.11. o Or you can create a new specimen and record your samples. Acquire and record your samples. o To view the compensation the option Enable Compensation should be ticked (in the Compensation tab in the Cytometer Window) in all the sample tubes. 70 Exporting data from Diva as listmode (.fcs) files When you have finished running all your controls and samples, the data needs to be exported from the software as flow cytometry standard, or listmode files compatible with other flow software. FCS3.0 files are compatible with newer analysis software such as FlowJo. FCS2.0 are an older file type, which cannot use 18-bit data, and so has lower resolution. Fluorescence data is exported in linear or log amplified depending how it was collected but you cannot change from one to the other post collection. Additionally, any compensation added at the time of collection cannot be removed. Select the experiment from the Browser. Right click--> Export--> FCS files Choose save path: Save the Experiment in D:\BDExport\FCS You can hold down the Shift key and use left click to select multiple experiments. In the window that appears check that fcs 3.0 is selected. This only exports the raw data files; if you wish to keep the gates/plots etc that you can re-import back into Diva, you will need to export the whole experiment as a .xml file. Please ask a member of staff to help you with this. All data is saved in the BDExport folder on the D: drive of the computer. The FACS Lab staff create a new folder each month where data will be exported so the save path should look something like this: D:\BDExport\FCS\2012\Aug 2012 Experiment completed. You can analyse your results in FlowJo or other analysis programmes. *You should not need to change anything; a new folder with your name will be created automatically, there is no need to type in your name and doing so creates unnecessary folders.* Diva will place all your experiments from one month together in your folder (named as it appears in the Browser window). Data backup is currently performed by the FACSlab and IS. 71 Working with templates If you will be running the same experiment every time you use the LSR, or would like to periodically run an experiment with the same antibodies, you may find it useful to export the experiment as a template. Creating a template allows you to save: Plots used to look at samples (scatter, fluorescence etc). Gates and gating strategies. “Statistics View” display settings. Fluorescence parameter labels (e.g. 530/30blue re-labelled to CD4-FITC will show up as CD4-FITC 530/30blue on any plot axes). Stopping gate settings (e.g. record 10K live cells). Compensation previously applied using single colour controls. Instrument settings such as PMT voltages. For example, in the FACS Lab we commonly use DNA analysis templates from which we can immediately open up an experiment displaying a scatter plot, PI-A vs PI-H (doublet discrimination plot), PI-A DNA histogram, PI set to Linear, and stopping gate set to record 10K of scatter gated single events. Once you have created an experiment using your template, you can tailor it specifically for that day’s experiment. For example, the PMT voltages may not be quite right for the cells you are running today (e.g. the scatter is slightly different because you are running a different type of cell, or the staining has not worked quite as well so you need to increase a fluorescence PMT voltage); bear in mind that compensation will need to be changed if you change fluorescence PMT voltages. You can also add parameters to your experiment if you have stained to look at an extra marker, and so forth. *Important!* Compensation should always be checked and not simply assumed to be correct! Compensation can generally only be saved and re-used if you are using the same batch of antibody (you should perform the experiment a few times including the single colour controls to check that you get consistent results with your staining). If you get a new batch of antibody, you will need to check with single colour controls that the compensation is still accurate. Changing any of the fluorescence PMT voltages will affect the compensation! 72 Creating a template After you have set up and run the experiment for the first time, you should have an experiment (which you should have re-named with your initials, the date, and LSRIIA/LSRIIB/Fortessa A/ Fortessa B) within YOUR FOLDER in the Browser (as described in section 3.2). *Saving a template* Ensuring first that the experiment is open: Right click on your experiment (book icon) in the Browser; select from the menu: Export Experiment Template Users This opens a Wizard: Rename your template; include your name or initials. Change the template type to: “Users”. Click Finish. The location of the template (e.g. should you want to delete it) on the D drive is: D:\BDExport\Templates\ Experiment \ Users * Using the template* Click once on your folder to highlight the location to create an experiment. Select the Experiment tab New Experiment (OR: right click your folder) This opens a window showing all saved templates; choose yours from the list. Remember to re-name the experiment within the Browser. B A Fig. 3.2.9 Using a saved template to create a new experiment. A Creating a new experiment in selected folder (in this case, the FACSlab folder). B Selecting template: select Users tab and choose from list of saved templates. 73 Flow cytometers are complex instruments that require a precisely focused flow of cells through a three dimensional space (the flow cell). Perfect function of fluidics is required for the cells to correctly arrive at the laser interrogation point(s), and perfect alignment of excitation and detection optics is required to capture the fluorescence emitted by cells at these interrogation points with the greatest sensitivity, resolution and optimal signal to noise ratio. 4.1 Routine Checks In the FACS Lab, we routinely run checks to assess the performance of the cytometers. Broadly these can be divided into alignment checks and sensitivity checks. Brightly fluorescent beads, excited by each laser wavelength, are used to assure us that the cytometer is aligned; i.e. that the laser is hitting the stream at the right place and the fluorescence detection optics give the maximum signal. At a given voltage, the bead peak will fall in a certain channel so we can plot this over days, weeks and months to assess machine performance. As well as alignment, we have to assess sensitivity; this is the cytometers ability to resolve small differences in fluorescence, especially low levels of fluorescence. To do this we use beads that are a mixture of levels of fluorescence (typically we use 8-peak beads i.e. a solution that contains 8 different levels of fluorescence). Again, at given voltages we know the best separation that can be achieved and this is monitored over time. Fig. 4.1 8-peak beads used in quality control assurance. 74 Deterioration in either alignment or sensitivity will indicate that intervention is needed and this is done either by members of the FACS Lab or by company field service engineers. 4.2 Laser Delay, Area Scaling and Window Extension Fluorescent beads also help us to set laser delay, area scaling factor and window extension. These are three parameters that are important but should not be altered by users. The lasers in the LSRs are spatially separated i.e. particles travel sequentially through the beams. For example, in the LSRIIA the order is 488nm, 405nm, 355nm, 633nm. The electronics need to be able to match signals to each laser; to do this, the distance between each must be precisely known. We use fluorescent beads to check this. As alignment in the LSR is fixed, this should not alter. However, if the flow cell becomes dirty, particles take more time to pass between lasers and the timing therefore becomes out of tolerance. Cleaning usually cures this. If you notice a problem, you should alert any member of the FACS Lab. The area scaling factor is used to ensure that area and height measurements are in the same magnitude. The H signals generated by the electronics are fixed i.e. they will have a value somewhere between 0 and 16,384 (214). A signals are the sum of all the sampled height signals within the window gate and for a typical pulse is in the range of 0 to 262,144 (218), so a scaling factor has to be applied to the A signal. In practice, this will make very little difference for most cell types and the value has been set to cover the range seen in the Lab. The window extension represents the amount of time during which a pulse from the particle is sampled. Increasing the window extension extends the detection time to allow for a more complete sampling of the signal pulse. However, if the window extension is too large, more noise is included in the signal pulse, and the peak CV is increased. 75 Alternatively, if the window extension is too small, the pulse may not be completely measured. On the LSR, this should not be adjusted by the user. These parameters are routinely checked and adjusted if necessary by FACS Lab staff. For more information please visit Practical Flow Cytometry” (Needs registration): http://www.beckmancoulter.com/wsrportal/wsr/research-and-discovery/products-andservices/flow-cytometry/practical-flow-cytometry/index.htm 76 One of the basic requirements for flow cytometry is a single cell suspension free from clumps. For cultured cell lines, there are two types: cells grown in suspension (e.g. B and T cell lines) and adherent cells which may grow on feeder layers, coated surfaces or straight onto plastic. Cells can also be obtained from whole blood or tissue samples. 5.1 Harvesting cells Preparing suspension cells basically involves taking an aliquot, pelleting by centrifugation, washing and staining. Preparing cells from whole blood is similar to working with cells in suspension, however a lysis step is usually required to remove red blood cells as there are so many of them (problematic if you are trying to look at rarer events if not lysed, and also very autofluorescent). They are also difficult to distinguish from white blood cells using scatter. Adherent cells should be harvested as for tissue culture. Typically this is with trypsin (this is cell type dependent, fragile cells may also require gentler enzymes; there are many commercially available cell dissociation solutions that you may wish to try). Cells should be carefully pipetted to break up clumps, and filtered through a suitable sized mesh (a cell strainer is the easiest way to do this). Take care also to not over digest as this may cause damage and clumping. *Surface Antigens* If you are looking at surface antigens in adherent cells, be aware that some antigens may be digested when harvesting cells for analysis, so you may have to test alternative enzymes. If the antigen is sensitive to all enzymatic/chelating treatments then you may have no choice but to mechanically retrieve your cells by scraping with a rubber policeman or cell scraper. This is quite a harsh method that often damages cells and produces clumps that should be broken up by pipetting or passing through a 21 gauge needle followed by filtering through mesh. Be aware, however, that you may still lose some of your cell surface markers due to damage. 77 5.2 Resuspending cells for analysis Suspension cells can be resuspended in PBS; (use Ca2+/Mg2+ free) these should not clump unless there are a lot of dead cells (released DNA is sticky, treat with DNAse if problematic). Adherent cells are often fine in PBS also, but sometimes a chelating agent, EDTA, can be added (up to 5uM) to help inhibit the reforming of metal dependent cell:cell interactions, if the cell type you are using has a tendency to clump. Foetal calf serum may be added if viability is a problem but it must be kept to a minimum (less than 2%), as excess of protein (which is sticky) will clog the machine and can also act as a lens, affecting the scatter signals. The way that you retrieve and resuspend your cells for analysis is really cell type dependent, and it is always a good idea to talk to other members of your lab who may have used the cell line you are using. We can help with the retrieval of cells from various tissues/other 3-D structures (e.g. embryoid bodies) or even paraffin sections. The FACS Lab staff are happy to offer assistance with protocols to meet any sample preparation requirements you may have and also to offer advice if you are having problems preparing your cells. 78 ADC: analogue to digital converter, electronic component that digitises an analogue input signal (in this case the electrical signal received from the PMTs). Digitisation is required for interfacing with the cytometer software. analog: in cytometry, generally describes older electronics that measure signals as a continuous spectrum, retaining information in the electronics while manipulations such as gating are performed; ultimately the signal is digitised to interact with the software. area measurement: integral of the area under the curve of an electrical pulse/signal, describes the total fluorescence from one cell. area scaling: use of a scaling factor which is applied to an electrical signal in order to make use of full range of the 18-bit data scale available in the electronics. autofluorescence: any fluorescence, generally due to cellular organelles and cyclic structure molecules other than due to the fluorochrome of interest. bandpass filter: optical filter which collects a narrow window of light. baseline PMT voltage: minimum voltage that can be applied to a PMT before bottom end sensitivity, and resolution is compromised. benchtop: generally used to refer to analysers. bi-exponential display (also logicle display): type of graphical display that has log scaling at the top end, and linear around zero, negative events can be visualized instead of piling up on the axis. Clean (FACSClean): bleach cleaning agent manufactured by BD. compensation: the mathematical subtraction (applied through the software by the operator, but performed by hardware components) to remove a particular fluorochrome's emission that can be detected in channels other than its own. comp beads (compensation beads):a type of antibody capture bead made of polystyrene, manufactured by BD; available as Anti-Rat or Anti-mouse Ig, kappa. They bind to rat or mouse light chain Ig and can be used instead of cells. Other commercially produced capture beads are also available that can be used to aid compensation. configuration: used to describe the optical system in a cytometer - the number of lasers 79 (and their wavelengths) number of detectors (and which optical filters) it is equipped with. contour probability plot: graphical display with contour levels drawn expressing percentages of events falling at different levels, particularly useful for visualising the distribution of events to determine how well-separated populations from closely drawn gates will be. core stream: stream of sample injected into the flow cell in the middle of the sheath; undergoes hydrodynamic focussing. CV: coefficient of variation; a measure of the dispersion of a data set. Commonly used in flow cytometry to evaluate quality of DNA staining and to determine data file size (number of target events required) in rare event analysis. dead time: processing time during which a cytometer cannot handle information coming from the next cell passing through the interrogation point(s). detector: see also photomultiplier tube. dichroic mirror: mirror which reflects specified wavelengths and may have long or short pass function (reflecting light longer or shorter than specification). digital: in flow, generally describes a cytometer with electronics that immediately convert the analogue signal (from the PMTs) into binary form before further manipulations are performed. This system is less vulnerable to noise, and cytometer dead time is also reduced. digitise: conversion of information to digital form - that is, binary code (0s and 1s). Digital data has increased resolution if the bit size is increased (e.g. 10110011= 8 bit, 110101001110= 10 bit). directly conjugated antibody: antibody directly linked to a fluorochrome, no secondary antibody is required. doublet discrimination: gating of single events for analysis, typically using pulse width measurements. event: particle of interest above threshold value passing through the primary laser and 80 triggering measurement FACSFlow: BD brand sheath fluid for analysers; contains mild cleaning agents. fcs file: flow cytometry standard, or “listmode” type of file. flow cell: region in cytometer where hydrodynamic focusing occurs, a cuvette-type structure in the LSRII. forward scatter (see scatter): related to refractive index of the cytoplasm as well as cell size. Fluorescence Minus One (FMO or F-1): all colours MINUS one colour. Particularly important for multi-colour flow; used to draw gates while taking into account nonspecific combinatory interactions of all fluorochromes used in an experiment. Each colour used should have both a single colour control (for compensation) and an FMO (for gating). gate: a logical combination of regions. height measurement: the height of the electrical pulse/signal, usually attributed to the brightest spot of the particle. histogram: single parameter graphical representation of data created by placing numbers of cells (y-axis) in channels (x-axis) representing a range of (increasing) values. hydrodynamic focusing: fluid focusing by the addition of a sample into a faster flowing sheath stream; used to constrict the core and promote a single celled stream that passes through the interrogation points. interrogation point: the laser intercept point, where fluorochromes from a particle of interest become excited, emitting fluorescence which is then collected for analysis. laminar flow: stable fluid flow in layers with no mixing. laser delay: amount of time required a for particle to pass between each laser (due to spatial separation of lasers) - allows hardware to determine which signal was generated by which laser, on the basis of timing. 81 lookup table (LUT): data structure stored in computer memory – used in cytometers to convert values to log. longpass filter: transmits light longer than specified wavelength. marker: equivalent of a gate, which can be drawn onto a histogram to define a region. median fluorescence technique: a means to calculate correct compensation, by adding compensation while observing median statistics for negative and positive populations in channels that spill-over is being compensated out of, until the median values are equal. noise: signals other than those coming from fluorescence light given off by particle of interest. octagon (PMTs): octagonal arrangement accomodating up to 8 detectors. optical bench: the optical layout of the cytometer, a bench area comprising the lasers, emission, collection and detection optics. parameter: attribute that can be measured by the cytometer. peak: pulse height measurement. photomultiplier tube, PMT: device with a specially coated surface designed to collect photons and from them generate an amplified electrical signal (electron flow). population: 1) cells defined by a gated region, 2) cluster of cells apparent by eye, on a plot. precision: the accuracy of a cytometer, ability to make the same measurement on a particle. consistently. primary antibody: antibody directed against antigen of interest, requiring a secondary fluorochrome linked antibody. primary laser: laser used to trigger the cytometer electronics when a cell passes through. prime, priming: function on LSR that prepares the flow cell by emptying and slowly refilling it, ensuring that it is free of air bubbles. 82 pulse width measurement: see also width measurement. region: windows drawn by the user in the software to highlight cells of interest. resolution: a cytometer’s ability to distinguish between slightly different levels of signal, is improved by increased bit-size digitisation generated by better ADCs; closely tied in with sensitivity and precision. Rinse (FACSRinse): detergent cleaning agent manufactured by BD. sample injection port (SIP): assembly on LSRII where sample tube is installed. sample injection probe (also “sample probe”): metal tube on SIP which takes up sample from facs tube. scatter: light scattered by a particle, usually measured from the primary laser. May be scattered forward (in the direction of the laser) or to the side (90° to direction of laser in the same plane). secondary antibody: fluorochrome linked antibody directed against species that primary antibody was generated in, allows primary antibody to be detected and also has amplifying effect. sensitivity: a cytometer's ability to detect a dim positive signal over background/noise, or the threshold set. sheath: pressurised fluid that the sample is injected into; the sheath then carries the sample to the flow cell for interrogation. In the FACS Lab, we use PBS as sheath fluid. sheath line: tubing carrying sheath from the tank to the flow cell. sheath filter: filter membrane that ensures PBS is free of particles. shortpass filter: optical filter transmitting shorter than specified wavelength. side scatter (see scatter): related to granularity or complexity of a particle, side scatter increases in dead or dying cells; also changes when cells are fixed. signal to noise ratio: ratio that compares a desirable signal (e.g. music) to undesirable signals (such as background noise). The higher the value, the less obtrusive the 83 background noise is. spectral overlap: emissions by a fluorochrome detected in ranges other than around its peak. Emission curves are generally not Gaussian (bell-shaped), and usually trail off to the lower energy end of the spectrum (often referred to as a "red-tail of emission") thus overlapping more from this side. spill-over: also known as spectral-overlap. tandem (tandem conjugate): pair of fluorochromes that have been chemically linked together in such a way that they are spatially very close. This allows the "electron donor", when excited, to transfer its excited state electrons to its partner (the recipient) by resonance energy transfer. The recipient uses this energy to emit fluorescence in its particular emission wavelength. This has been a means to create new fluorochromes by combining ones that already exist. Tandem conjugates are sensitive to light and oxidation and their quality should be monitored carefully. trigon (PMTs): arrangement of three PMTs in collection optics. threshold: channel number that needs to be exceeded in order for the electronics to start measuring a pulse (i.e. to recognize that there is a cell passing the laser). viability dye: dye that enters dying/dead cells which have compromised membranes, allowing discrimination between live and dead populations. width measurement: calculated value that is proportional to the amount of time a cell takes to pass through a laser. window extension: extension of the window of time in which a pulse is being sampled. Too small a windows extension value may result in loss of information, but too large means the electronics may still be sampling the first cell when the next cell arrives. 84 Symptom Cause Solution Threshold setup incorrectly. No events displayed. Sample injection port clogged. Pressure leak crack in FACS tube. Run button not green. Pressure leak – instrument not pressurised. Threshold should be set around 10-15k on FSC. Try these procedures (in order!!): 1-Pipette/filter – if sample has visible clumps. 2-Backflush (press Prime) the line. If clog persists: 3-Run the cleaning protocol. 4.Apply a slight vacuum with a syringe. If none of these steps work please ask a member of FACS lab staff for help. Transfer sample to a new tube. Instrument not pressurised, check sheath tank lid properly fastened and air line connected. Lift pressure release valve to check if tank is pressurising. Crack in FACS tube, transfer sample to new tube. Air in sheath filter - purge air (follow instructions described previously). DIVA not connecting to cytometer. Interface communication failed. Restart computer, if this doesn’t solve it, restart BOTH instrument and computer. If the problem persists ask a member of the FACS lab staff for help. Software glitch. Re-export data and count files again. Always check number of files!! FCS files does not contain uncompensated data. Data exported as FCS2 not FCS3. Re-export in FCS3. Check FCS2 not selected. Loss of compensation in exported data even if exported in FCS3. Software glitch. Try re-opening the experiment, re-export files. Ask a member of staff if problem persists. Unexpected loss of signal, unusual compensation. (e.g. PerCP and APC) Time delay issues between lasers. No fluorescence signal, or lower than usual. Incorrect filter in front of PMT. Not all .fcs files exported. 85 If problem occurs while using tandem conjugates, check the dyes have not degraded; ask FACS staff. Ask a member of the FACS lab staff for help. Check the correct filter is in place and change if necessary – ask a member of staff to demonstrate.
© Copyright 2026 Paperzz