Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Appli cat i on N ote 1 0 3 4 2 Automated Determination of Gasoline Range Organics (GRO) in Water Via Valve-and-Loop Headspace GC Key Words Petroleum, hydrocarbons, water, flame ionization detector, headspace, EPA Method 5021A, EPA Method 8015D Goal The purpose of this application note is to optimize the analytical set-up for the determination of gasoline range organics in water. Introduction Oil spills of varying magnitudes can be a serious source of contamination of the surrounding environment, in particular for bodies of water. Water pollution can pose a serious threat to the affected ecosystem and the encircling area habitability. For this reason, it is important to have an accurate and effective system for screening and evaluating possible contaminations. Markers of possible water contaminations are gasoline range organics (GRO) which correspond to the range of alkanes from C6 to C10 and covering a boiling point range of approximately 60 °C – 170 °C1. EPA method 5021A2 is applicable to a wide range of organic compounds, including GRO, that have sufficiently high volatility to be effectively removed from soil/sediment, solid waste, aqueous, and water-miscible liquid samples using an equilibrium headspace procedure. Similar to EPA method 8015D3, a simple and reliable GC/FID system could then be used for the quantification of volatile organics or as a screening technique prior to headspace GC/MS quantitative analysis. The headspace analysis can be performed with a chromatographic system composed of a Thermo Scientific™ TriPlus™ 300 Headspace autosampler and a Thermo Scientific™ TRACE™ 1310 GC controlled by the Thermo Scientific™ Dionex™ Chromeleon™ 7.2 Chromatography Data System (CDS) software. This system allows sample analysis and data processing to occur in an automated, reproducible, and reliable manner. Equipment A TriPlus 300 Headspace autosampler and the TRACE 1310 GC are used for the analysis. The data are collected and processed with Chromeleon 7.2 CDS. The system is equipped with one Instant Connect Split/Splitless (SSL) injector with a dedicated headspace liner (P/N 453A1335) and an Instant Connect Flame Ionization Detector (FID). A Thermo Scientific™ TRACE™ TR-1 column was used with the following dimensions: 30 m × 0.32 mm × 3 µm (PN 260A395P). The standards used for calibration are acquired from Restek™ (PVOC/GRO mix (Wisconsin) PN 30095). The calibration curves are made with seven levels at the following concentrations for each analyte: 10 ppb, 50 ppb, 100 ppb, 1 ppm, 5 ppm, 10 ppm, and 100 ppm. 2 Analysis Results and Discussion The sample analyzed is 5 mL of water placed into a 20 mL vial. Vials are tightly closed with magnetic caps (Thermo Scientific™ La-Pha-Pack™ Aluminum Crimp Seals, PN 18091307). The TriPlus 300 Headspace sampling system parameters are configured as follows: Table 1 reports the calibration results for all the compounds tested. All the calibration curves show outstanding linearity in the tested range. • Equilibration time of the sample is 25 min with vial shaking set to high Table 1. Calibration results for all compounds. Number of Points Coeff. of Determination MTBE 7 0.99999 • Transfer line temperature set to 95 °C Benzene 7 0.99975 • Pressurization mode is set to Pressure; pressure is set to 1 bar and the pressure equilibration time to 0.2 min Toluene 7 0.99980 Ethylbenzene 7 0.99986 m-Xylene/p-Xylene 7 0.99989 • Loop size is 1 mL o-Xylene 7 0.99992 • Injection mode is Standard, and injection time is set to 0.5 min 1,3,5-Trimethylbenzene 7 0.99997 1,2,4-Trimethylbenzene 7 0.99998 Naphthalene 7 0.99987 • Oven temperature set to 85 °C • Manifold temperature set to 95 °C • Loop filling mode is set to Pressure; pressure is set to 0.5 bar with an equilibration time of 0.1 min • Sample line is purged after injection for 1 min at 100 mL/min • Vial venting is set to On The TRACE 1310 GC parameters are configured as follows: • Starting temperature at 50 °C, hold 1 min, then raise at 15 °C/min up to 220 °C, hold 2 min. • SSL inlet temperature is set to 150 °C, split mode with split flow 60 mL/min with a split ratio of 20:1 • Carrier used is helium, with a constant flow of 3 mL/min • FID temperature is set to 300 °C Peak Name MTBE 90 80 70 60 Area 40 215.4 200.0 200.0 187.5 187.5 175.0 175.0 162.5 162.5 150.0 150.0 137.5 137.5 125.0 125.0 112.5 112.5 100.0 100.0 87.5 30 75.0 75.0 62.5 62.5 20 50.0 50.0 37.5 37.5 25.0 25.0 12.5 0 -5.6 12.5 0 -5.6 -11 10 0 -7 -4 0 215.4 20 40 60 Amount 80 100 121 Ethylbenzene -11 426 0 20 40 60 Amount 80 100 120 128 m-Xylene/p-Xylene 190.2 200.0 400 187.5 375 175.0 350 162.5 325 150.0 300 137.5 137.5 275 125.0 125.0 250 112.5 225 100.0 Area 112.5 100.0 12.5 0 -5.6 -11 0 20 40 60 Amount 80 100 120 128 1,3,5-Trimethylbenzene 190.2 25.0 12.5 -19 0 20 40 60 Amount 80 100 120 134 1,2,4-Trimethylbenzene 0 -8.7 -37 -20 68.5 65.0 150.0 55.0 137.5 137.5 50.0 125.0 125.0 45.0 112.5 112.5 40.0 100.0 100.0 35.0 87.5 75.0 75.0 62.5 62.5 50.0 50.0 37.5 37.5 25.0 25.0 12.5 12.5 0 -8.7 -37 -20 0 -8.7 -37 -20 40 60 80 Amount 100 120 140 150 Figure 1. Calibration curves of some of the target analytes. 0 20 40 60 80 Amount 100 120 140 150 Naphthalene 60.0 Area Area 162.5 150.0 Area 175.0 20 o-Xylene 37.5 162.5 0 120128 87.5 175.0 87.5 100 50.0 25 0 -16 80 62.5 50 25.0 60 Amount 75.0 75 37.5 190.2 200 100 50.0 40 150.0 125 62.5 20 162.5 150 75.0 0 175.0 175 87.5 3 Toluene 87.5 Area Area 50 Area Benzene 215.4 Area 100 30.0 25.0 20.0 15.0 10.0 5.0 0 0 20 40 60 80 Amount 100 120 140 150 -4.8 -68 -50 0 50 Amount 100 150 184 4 100 90 80 m-Xylene/p-Xylene 70 60 50 Ethylbenzene o-Xylene 40 pA 30 20 10 0 -10 -20 -30 -42 7 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 8 Minutes Figure 2. Separation of the couple ethylbenzene m/p-xylene at a resolution of 1.83. Table 2. Repeatability results of each GRO compound at 1 ppm in water. Peak Area is expressed in pA*sec. MTBE Benzene Toluene Ethylbenzene m/p-Xylene o-Xylene 1,3,5-Trimethylbenzene 1,2,4-Trimethylbenzene Naphthalene 1 57.3628 55.8949 59.9575 59.6926 126.4131 74.4254 66.2386 77.0404 56.238 2 57.0797 55.4711 59.4999 59.3961 125.7007 74.0777 65.9278 76.7329 56.1561 3 57.5383 56.4514 60.8489 60.9327 128.8599 75.7043 67.8679 78.6368 56.355 4 57.6155 56.1589 60.3285 60.3256 127.5803 75.0778 67.0354 77.8603 56.5539 5 57.102 59.9814 126.7492 74.5956 66.5793 77.2604 55.9639 6 57.4345 56.2741 60.466 60.4806 127.7841 75.1562 67.2012 77.7268 56.0468 7 58.2604 57.1632 61.5558 61.641 130.1047 76.5622 68.5266 79.4365 56.7669 8 58.7969 57.4769 61.8347 61.8393 130.5417 76.8693 68.6879 79.5756 57.3098 9 57.3306 57.3574 61.3496 61.0969 128.8684 75.6615 67.4593 78.0111 55.9104 55.7743 59.9651 Standard Deviation 0.5641 0.7267 0.8032 0.8457 1.6559 0.9457 0.9607 1.0069 0.4498 Relative Standard Deviation 0.98% 1.29% 1.32% 1.40% 1.29% 1.26% 1.43% 1.29% 0.80% Figure 2 shows good separation of the couple ethylbenzene m/p-xylene, whose resolution is 1.83. Repeatability has been evaluated for each compound and the results are reported in Table 2. 5 100 90 80 m-Xylene/p-Xylene 70 1,3,5-Trimethylbenzene Ethylbenzene 60 o-Xylene 50 pA 1,2,4-Trimethylbenzene Toluene Benzene 40 Naphthalene 30 MTBE 20 10 0 -10 0 1 2 3 4 5 6 Minutes 7 8 9 10 11 12 12.5 Figure 3. PVOC/GRO mix (Wisconsin) standard in water at 1 ppm. 11 10.5 10 9.5 9 8.5 pA Toluene 8 1,3,5-Trimethylbenzene 7.5 m-Xylene/p-Xylene MTBE 7 6.5 o-Xylene 6 Naphthalene 1,2,4-Trimethylbenzene Ethylbenzene Benzene 5.5 5 4.5 4 0 1 2 3 4 5 6 7 8 9 10 11 12 12.5 Minutes Figure 4. Chromatogram of tap water spiked with gasoline. Blank samples have been analyzed after each sequence of at least 10 samples, and they showed a clear baseline. Figure 3 shows a chromatogram of the Wisconsin standard mix at the 1 ppm calibration point, and Figure 4 shows the chromatogram of a tap water sample spiked with gasoline. References As the results confirm, the system comprised of the TriPlus 300 Headspace autosampler, the TRACE 1310 GC, and Chromeleon CDS is a reliable and automated solution for Gasoline Range Organics in water. With its 120-vial sample tray and the large 18-vial incubation oven overlap capacity, the system guarantees excellent throughput. The inertness of the entire sample path and high temperature capability eliminate any carryover effect, ensuring the highest sample integrity and results consistency. The optional barcode capabilities of the autosampler and the accurate auditing capabilities of Chromeleon CDS also ensure the highest data quality and traceability. 1. “Interlaboratory Study of Three Methods for Analyzing Petroleum Hydrocarbons in Soils,” API. Publication Number 4599, American Petroleum Institute, March 1994. 2. EPA Method 5021A. 3. EPA Method 8015D. www.thermofisher.com ©2016 Thermo Fisher Scientific Inc. All rights reserved. Restek is a trademark of Restek Corp. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 South Africa +27 11 570 1840 Spain +34 914 845 965 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 AN10342_E 08/16S Appli cat i on N ote 1 0 3 4 2 Conclusion
© Copyright 2026 Paperzz