Trias®tsa Abutment - Servo

Indication
Trias®titanium-shock-absorber
The Trias®tsa abutment contains an integrated damping element inside the post. Thus it guarantees optimal biomechanical properties and enables an even force distribution to prosthesis, bones and inserted implants.
In this way prosthesis fractures can be considerably reduced. The transverse resilience and the 17° cone of the
Trias tsa post allow a correction of divergence of up to 38°. The Trias ®tsa abutment is designed as support for complete denture
®
prosthesis in the lower or upper jaw. It is available for implant diameter of 3.8 mm. As for a prosthetic bridge provision (either
Procedure
System Elements
implant / implant or implant / natural abutment tooth), the Trias®tsa post is alternatively available for fixed dentures.
Artikel-Nr.
857381tra
H1 for fixed prosthesis
857382tra
H2 for complete denture prosthesis
857383tra
H3 for complete denture prosthesis
857384tra
H4 for complete denture prosthesis
Trias®tsa post for implant diameter of 3.8 mm (or larger)
Distance height of 1 mm for fixed dentures (bridges)
Distance height of 2 / 3 / 4 mm for complete denture prosthesis
857000tra
Trias ®tsa female
857001tra
Trias ®tsa model analog (model pin with crown)
857002tra
Trias ®tsa impression coping or auxiliary modelling element
857003tra
Trias ®tsa socket wrench – short
857004tra
Trias ®tsa socket wrench – long
The prosthesis can be placed both directly into the mouth of the patient and indirectly by means of impression taking and model making in
the dental lab. For that, impression posts and model analog of the Trias® implant system can be used.
The Trias®tsa post is screwed into the implants using the socket wrench
(short or long) and the torque ratchet with torque of 30 Ncm for fixed denture
prosthesis (bridges) and of 25 Ncm for complete denture prosthesis.
Procedure
Trias®titanium-shock-absorber
The Trias®tsa female with circle snap system is placed onto the post in such way
that the internal hexagon of the Trias®tsa female matchs the post’s outer hexagon
and – using light pressure – clicks into place.
To avoid that the used cold-curing polymer flows into inverse conical spaces, rubber
dam rings are fitted over the posts and pulled to below the resilience grooves. In this
way the complete dentures remain removable.
When using an already existing complete denture prosthesis, this will be ground free
at the implant positions. Then add one bead of cold-curing polymer into each present
cavity and place the prosthesis.
After completion of polymerisation, the prosthesis including the Trias®tsa females
can be removed from the posts. Now the rubber dam rings can be removed using
forceps. If needed, excess cold-curing polymer is removed.
Implant dimensions on immediate loading and immediate restoration of 2-4 interforaminally inserted implants
The Trias®tsa abutment allows an immediate restoration including an immediate loading of implants in compliance with the relevant guide-
lines. This relates, for example, to the immediate loading and immediate restoration of two implants placed in region 33/43. A diameter of at least
Status 2010/07/14
3.8 mm and a length of 16 mm should be given. Additionally the primary implant stability shall be at least 35 Ncm.
Distribution:
Servo-Dental GmbH & Co. KG
Rohrstraße 30
D-58093 Hagen-Halden
Phone: +49 (0) 2331 - 9591-0
Fax: +49 (0) 2331 - 9591-25
[email protected] | www.servo-dental.de
Manufacturer: BoneCare GmbH, Steingasse 1, D-86150 Augsburg