GE Krautkramer ROTA - Ultrasonic Tube/Pipe Tester Brochure

GE Krautkramer ROTA Ultrasonic Testing
Machine Series
GE
Oil & Gas
Krautkramer ROTA Systems Product Line
Name
GE supplies the Krautkramer ROTA systems which, depending
on the client’s requirements, can be configured for individual
manual, central or motor adjustment, and are notable for
highly repeatable results. A probe revolution rate of 60008000 rpm provides highest productivity rates.
Version
OD max
TRD flaw
TRD dim
RPM
Conventional ROTA
25
8k8
25
4
4
8000
40
6k8
38,1
4
4
6000
65
90
In addition to defect detection, Krautkramer ROTA systems
can also be used for geometry checking. The systems are well
known for high test accuracy, repeatability of results, and
test speed. Flaw-type defects throughout the material crosssection, and divergences in the dimensions for internal and
external diameter, wall thickness, ovality and eccentricity can
all be detected in a single inspection cycle. A manual probe
fine-adjustment system is installed as standard. It enables
fast and repetitive adjustment. The equipment is fitted with
single or multiple probes, depending on the particular test
application. Single or multi-channel, manual or computercontrolled, parallel scanned inspection electronics systems
are used, depending on the equipment configuration. Many
test standards require a second, independent physical test
procedure. GE fulfils this specification via the integration of
Eddy Current test systems.
130
6k8
65
4
4
4000
8k10
65
6
4
4000
8k10
90
8
2
3000
12k22
90
18
4
3000
8k10
130
6
4
2000
12k22
130
18
4
2000
8k10
45
4
2
4500
Krautkramer Testing Machines
ROTA Z
45 Z
65 Z
8k22
65
6
2
3500
90 Z
8k22
90
6
2
2850
130 Z
8k22
135
6
2
2000
Full Body and Volumetric Inspection
of Tubes and Bars
ROTA S
130 S
16k18
130
16
2
1600
180 S
16k20
180
18
4
1200
280 S
16k24
280
20
4
900
370 S
16k24
370
20
4
650
450 S
16k24
450
20
4
500
GE Krautkramer ROTA Series
450
400
350
300
250
200
150
180
130
135
130
100
90
90
90
56
ROTA 450S
ROTA 370S
28
ROTA 280S
28
38
ROTA 180S
ROTA 45Z
28
ROTA 130S
ROTA 40
18
ROTA 90Z
9
18
ROTA 90
8
13
ROTA 65Z
6
13
ROTA 65
38
ROTA130Z
65
45
25
74
ROTA 130
65
ROTA 25
Hot-processed tubes and bars necessitate special design
features for test and inspection systems, in terms of surface
quality, fouling, coarse manufacturing tolerances, test
throughputs and test speeds. GE meets these requirements
with its Krautkramer ROTA S systems, a special water routing
and cleaning system reliably removes particles from the
probes.
The Krautkramer ROTA S series was specially developed for
hot processed tube with rough surfaces, bend tolerance
greater than standard API specification and including tubes
with upset ends. Unlike the conventional Krautkramer ROTA
systems, the Krautkramer ROTA S features a central probe
distance adjustment system providing extremely fast change
of material dimensions.
370
50
Testing of hot-processed tubes and bars
450
280
0
GE Krautkramer ROTA test systems have their origins
in high-precision testing of tubes and heat-exchanger
tubes for nuclear applications. Krautkramer ROTA 25 and
Krautkramer ROTA 45 Z inspection systems are designed
for this dimension range. The Krautkramer ROTA Z is a
derivate from the standard Krautkramer ROTA line with
central probe adjustment capabilities for fast changeover
times. The Krautkramer ROTA and ROTA Z systems are
conform to the extremely strict inspection requirements for
nuclear specifications and permit not only defect detection
but also complete geometrical checking of the tube crosssection, including internal and external diameter quality
and eccentricity. These systems are in use with virtually all
manufacturers of nuclear piping around the world.
500
Nonferrous tube testing
www.gemeasurement.com
GEIT-60018EN (03/16)
© 2016 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names
mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
Product Information
Krautkramer ROTA 25
Krautkramer ROTA 40
Krautkramer ROTA 65
Product Information
Krautkramer ROTA 45Z
Krautkramer ROTA 65Z
Krautkramer ROTA 90Z
Design
Rotary head with safety interlock and attached motor prepared for mounting on elevating sliding table
Design
Rotary head with safety interlock and attached motor prepared for mounting on elevating sliding table
Speed
400 - 8000 revolutions per minute
Speed
300 - 4500 revolutions per minute
300 - 6000 revolutions per minute
800 - 3500 revolutions per minute
Infinitely adjustable, frequency controlled three-phase motor with PLC control and oil/air lubrication, safety interlocks
Product outer diameter range
Approx. 6 – 25 mm
Signal transmission
Contact less and maintenance-free by means of rotating capacitors
Model 8K-8 with 8 probe holders
and 8 transmission channels
Probe assembly
Approx. 8 – 38 mm
Model 6K-8 with 6 probe holders
and 8 transmission channels
Approx. 13 – 65 mm
Multi section crystal probes
also available on request
Water seal
Sliding ring seals, water basins for backflow of water on inlet and outlet sides
Guidance inside of Rotary head
Rigid, non-rotating guide bushings in inlet and outlet of the rotary head exchangeable for each tube/rod diameter
Up to 25 mm dia. as quick-change
cartridge system
Up to 40 mm dia.
as quick change cartridge system
Through put direction
Inlet direction
Left or right inlet depending on test line layout
Mechanical setup
Replacement of guide bushings and rubber disk water wipers. Individual probe distance and angle adjustment
Incident angle setting
Fine angle adjustment for each probe. Squint angle and acoustic irradiation angle adjustable separately
Water regulation
Manual water control valve with water pressure indicator; water connection on inlet and outlet sides
Flaw detection
e.g. according to EN 10246 / 6 / 7 / 13 / 14 / 17 or according to international specifications
Types of flaws
Depending upon application, the rotary head is equipped with probes for core, longitudinal, transverse, oblique flaws and
delamination and/or dimensional measurement
Water path
Adjustable to application requirements
Testing speed
The UT electronics pulse repetition rate of up to 20 KHz accomodates the high RPM-rate of the rotary head, still permitting statistical noise
reduction for high probability of detection
500 - 2850 revolutions per minute
Infinitely adjustable, frequency controlled three-phase motor with PLC control and oil/air lubrication, safety interlocks
Product outer diameter range
approx. 9 – 45 mm
Signal transmission
Contact less and maintenance-free by means of rotating capacitors
Model 8K-10 with 8 probe holders
and 10 transmission channels
(also available with 12K-22 on request)
Single section crystal probes
800 - 3500 revolutions per minute
Model 8 K-10 with 8 probe holders
and 10 transmission channels
approx. 13 – 65 mm
Model 8K-22 with 8 probe holders
and 22 transmission channels
(8K-10 also available on request)
approx. 18 – 90 mm
Model 8K-22 with 8 probe holders and
22 transmission channels
Probe assembly
Single section crystal probes only. Up to 4 channels per probe holder location standard, more possible on request
Water seal
Sliding ring seals, water basins for backflow of water on inlet and outlet sides
Guidance inside of Rotary head
Rigid, non-rotating guide bushings in inlet and outlet of the rotary head exchangeable for each tube/rod diameter
Up to 35 mm dia. as quick change cartridge
system
Up to 40 mm diameter as quick change
cartridge system
Up to 60 mm dia. as quick change cartridge
system
Through put direction
Inlet direction
Left or right inlet depending on test line layout
Mechanical setup
Replacement of guide bushings and rubber disk water wipers. Separate probe distance and angle adjustment
Product Information Krautkramer ROTA 130S Krautkramer ROTA 180S Krautkramer ROTA 280S
Krautkramer ROTA 370S Krautkramer ROTA 450S
Design
Rotary head with safety interlock and
attached motor prepared for mounting on
elevating sliding table particularly designed
for testing black, hot produced material. AP
tubes with upset ends can also be tested
Rotary head with safety interlock and attached motor prepared for mounting on elevating sliding
table particularly designed for testing black, hot produced material
API tubes with
upset ends can
also be tested
Equipped with protecting flap with safety locking and with flap
for easily cleaning the testing drum and probes. API tubes with
upset ends can also be tested. For reduction of the accumulating dirt inside the testing drum, the Rota can be equipped with
an external tube cleaning device
400 - 1200 revolutions per minute
400 - 900 revolutions per minute
Speed
500 - 1600 revolutions per minute
Product outer
diameter range
Approx. 28 – 130 mm Approx. 38 – 180 mm Approx. 56 – 280 mm
Signal transmission
Contact less and maintenance-free by means of rotating capacitors
250 - 650 revolutions 200 - 500 revolutions
per minute
per minute
Infinitely adjustable, frequency controlled three-phase motor with PLC control and oil/air lubrication, safety interlocks
Model 16K-18 with
16 probe holders
and 18 transmission
channels
Model 16K20 with
16 probe holders
and 20 transmission
channels
Approx. 74 – 370 mm Approx. 90 – 450 mm
Model 16K-24 with 16 transducer manipulators and 24
transmission channels available plus 1 channel (outside) for
temperature compensation of dimensional measurement
Model 16K20 with
16 probe holders
and 20 transmission
channels
Model 16K-18 with
16 probe holders
and 18 transmission
channels
Probe assembly
Variable, single section crystal probes only.
Up to 4 channels per probe holder location
standard, more possible on request
Single and/or multi element crystal probes possible up
to 4 channels per probe location 14 channels available all in all.
Standard assembly: 2 x 4 or 4 x 2 - element transducer for LF,
8 single element transducer for TF, 4 single element transducer
for WT and lamination for combined LF/TF testing. For pure
LF testing 4 x 4 element transducer can be used allowing to
increase the testing speed
Variable, single and multi section crystal
probes possible (up to 4 channels per probe
location standard, more possible on request)
Water seal
Sliding ring seals, water basins for backflow
of water on inlet and outlet sides
The rotating head is internally equipped with sliding ring
seals. Soft rubber disks are used to seal the test drum
with the moving tube against leaking water
Sliding ring seals, water basins for backflow
of water on inlet and outlet sides
Guidance inside
of Rotary head
Rigid, non-rotating guide bushings in inlet and outlet of the rotary head exchangeable for each tube/rod diameter
Up to 35 mm dia.
as quick change
cartridge system
With central manual ratchet for all installed probes, distance display with aid of
adjustment gauge
(approx. 110 to 180 mm) exchangeable for each diameter, soft
rubber disks for the upper range of tube/rod diameter 180 to
275 mm) exchangeable for each diameter
Incident angle setting
Fixed angle (separate fine angle adjustment of each probe), common adaptation of adjustment angle for all probes with central distance
adjustment
Through put direction / Inlet direction
Left or right inlet depending on test line layout and customer’s requirements
Water regulation
Manual water control valve with water pressure indicator; water connection on inlet and outlet sides
Mechanical setup
Flaw detection
e.g. according to EN 10246 / 6 / 7 / 13 / 14 / 17 or according to international specifications
Replacement of guide bushings and rubber disk water wipers; probe distance adjustment with central manual ratchet for all installed probes,
distance display with aid of adjustment gauge
Types of flaws
Depending upon application, the rotary head is equipped with probes for core, longitudinal, transverse, oblique flaws and delamination
and/or dimensional measurement
Incident angle
setting
Water path
Adjustable to application requirements
The UT electronics pulse repetition rate of up to 20 KHz accomodates the high RPM-rate of the rotary head, still permitting statistical noise
reduction for high probability of detection
Fixed angle for W.T. measurement; optional transducer holder
with fixed angle or transducer manipulators (individually
adjustable ± 20°) for transverse flaw testing; adjustable transducer manipulators capable to be individually adjustable
± 20° for longitudinal testing. This allows a specific adjustment
of the transducers to most critical applications
Fixed angle, common adaptation of adjustment angle for all probes with central
distance adjustment; separate fine angle
adjustment of each probe
Testing speed
Fixed angle, common adaptation of
adjustment angle for all probes with
central distance adjustment;
separate fine angle adjustment of
each probe
Water connection
On inlet and outlet sides;
PLC flow control with display
Water circulation with water basins on inlet and outlet
side with filter, flow meter, pump with frequency
converter and PLC flow control by with display
On inlet and outlet sides; PLC flow control
with display
Water purification
Rapid drainage of dirty and contaminated
coupling water
2- or 3-chamber underground tank allows rough purification,
water backflow via band filter to the upper 2-chamber tank
made from stainless steel, coupling water supply to the
Rota from clean water chamber. Water substitution only
due to loss of spraying water. Water exchange one or two
times a year (depending on the tube surface)
Rapid drainage of dirty and contaminated
coupling water
Flaw detection
Depending on individual appication
Types of flaws
Depending upon application, the rotary head is equipped with probes for core, longitudinal, transverse, oblique flaws and delamination
and/or dimensional measurement
Water path
The water path is limited to 40 mm at the maximum test diameter
Testing speed
The UT electronics pulse repetition rate of up to 20 KHz accomodates the high RPM-rate of the rotary head, still permitting statistical noise reduction for high
probability of detection
Tube transportation
The original test bench supplied by GE S&I is equipped with at
least 4 triple centering units. They guarantee an optimal tube
guidance and satisfactory test results