2.2.36. Ionic concentration (ion-selective electrodes) EUROPEAN PHARMACOPOEIA 5.0 01/2005:20236 Table 2.2.36.-1. - Values of k at different temperatures 2.2.36. POTENTIOMETRIC DETERMINATION OF IONIC CONCENTRATION USING ION-SELECTIVE ELECTRODES Temperature (°C) k 20 0.0582 25 0.0592 30 0.0602 If the electrode system is used frequently, check regularly the repeatability and the stability of responses, and the Ideally, the potential E of an ion-selective electrode varies linearly with the logarithm of the activity ai of a given ion, as linearity of the calibration curve or the calculation algorithm in the range of concentrations of the test solution ; if not, expressed by the Nernst equation : carry out the test before each set of measurements. The response of the electrode may be regarded as linear if the slope S of the calibration curve is approximately equal to k/zi, per unit of pCi. E0 = R = part of the constant potential due to the apparatus used, gas constant, T = absolute temperature, F zi = Faraday’s number, = charge number of the ion including its sign. At a constant ionic strength, the following holds : Ci = molar concentration of the ion, f = the activity coefficient k = If : S , and = slope of the calibration curve of the electrode, the following holds : and for : . The potentiometric determination of the ion concentration is carried out by measuring the potential difference between two suitable electrodes immersed in the solution to be examined ; the indicator electrode is selective for the ion to be determined and the other is a reference electrode. Apparatus. Use a voltmeter allowing measurements to the nearest 0.1 millivolt and whose input impedance is at least one hundred times greater than that of the electrodes used. METHOD I (DIRECT CALIBRATION) Measure at least three times in succession the potential of at least three reference solutions spanning the expected concentration of the test solution. Calculate the calibration curve, or plot on a chart the mean potential E obtained against the concentration of the ion to be determined expressed as − log Ci or pCi. Prepare the test solution as prescribed in the monograph ; measure the potential three times and, from the mean potential, calculate the concentration of the ion to be determined using the calibration curve. METHOD II (MULTIPLE STANDARD ADDITIONS) Prepare the test solution as prescribed in the monograph. Measure the potential at equilibrium ET of a volume VT of this solution of unknown concentration CT of the ion to be determined. Make at least three consecutive additions of a volume VS negligible compared to VT (VS ≤ 0.01VT) of a reference solution of a concentration CS known to be within the linear part of the calibration curve. After each addition, measure the potential and calculate the difference of potential ∆E between the measured potential and ET. ∆E is related to the concentration of the ion to be determined by the equation : or Ion-selective electrodes may be primary electrodes with a crystal or non-crystal membrane or with a rigid matrix (for example, glass electrodes), or electrodes with charged VT (positive or negative) or uncharged mobile carriers, or sensitised electrodes (enzymatic-substrate electrodes, CT gas-indicator electrodes). The reference electrode is generally a silver–silver chloride electrode or a calomel electrode, with V S suitable junction liquids producing no interference. CS Procedure. Carry out each measurement at a temperature constant to ± 0.5 °C, taking into account the variation of the slope of the electrode with temperature (see Table 2.2.36.-1). S Adjust the ionic strength and possibly the pH of the solution to be analysed using the buffer reagent described in the monograph and equilibrate the electrode by immersing it in the solution to be analysed, under slow and uniform stirring, until a constant reading is obtained. General Notices (1) apply to all monographs and other texts = volume of the test solution, = concentration of the ion to be determined in the test solution, = added volume of the reference solution, = concentration of the ion to be determined in the reference solution, = slope of the electrode determined experimentally, at constant temperature, by measuring the difference between the potentials obtained with two reference solutions whose concentrations differ by a factor of ten and are situated within the range where the calibration curve is linear. 55 2.2.37. X-ray fluorescence spectrometry EUROPEAN PHARMACOPOEIA 5.0 Plot on a graph (y-axis) against VS (x-axis) and extrapolate the line obtained until it intersects the x-axis. At the intersection, the concentration CT of the test solution in the ion to be determined is given by the equation : To determine the concentration of an element in a sample, it is necessary to measure the net impulse rate produced by one or several standard preparations containing known amounts of this element in given matrices and to calculate or measure the mass absorption coefficient of the matrix of the sample to be analysed. Calibration. From a calibration solution or a series of dilutions of the element to be analysed in various matrices, determine the slope of the calibration curve b0 from the following equation : METHOD III (SINGLE STANDARD ADDITION) To a volume VT of the test solution prepared as prescribed in the monograph, add a volume VS of a reference solution containing an amount of the ion to be determined known to give a response situated in the linear part of the calibration curve. Prepare a blank solution in the same conditions. µM Measure at least three times the potentials of the test solution and the blank solution, before and after adding the reference solution. Calculate the concentration CT of the ion to be analysed using the following equation and making the necessary corrections for the blank : C VT = volume of the test solution or the blank, CT = concentration of the ion to be determined in the test solution, VS = added volume of the reference solution, CS = concentration of the ion to be determined in the reference solution, ∆E = difference between the average potentials measured before and after adding VS, S slope of the electrode determined experimentally, at constant temperature, by measuring the difference between the potentials obtained from two reference solutions whose concentrations differ by a factor of ten and are situated within the range where the calibration curve is linear. = absorption coefficient of the matrix M, calculated or measured, = net impulse rate, = concentration of the element to be assayed in the standard preparation. Mass absorption coefficent of the matrix of the sample. If the empirical formula of the sample to be analysed is known, calculate its mass absorption coefficient from the known elemental composition and the tabulated elemental mass absorption coefficients. If the elemental composition is unknown, determine the mass absorption coefficient of the sample matrix by measuring the intensity of the scattered X-radiation IU (Compton scattering) from the following equation : µMP = mass absorption coefficient of the sample, IU = scattered X-radiation. Determination of the net pulse rate of the element to be determined in the sample. Calculate the net impulse rate of the element to be determined from the measured intensity of the fluorescence line and the intensity of the background line(s), allowing for any tube contaminants 01/2005:20237 present. Calculation of the trace content. If the concentration of the element is in the linear part of the calibration curve, it can FLUORESCENCE be calculated using the following equation : (3) 2.2.37. X-RAY SPECTROMETRY Wavelength dispersive X-ray fluorescence spectrometry is a procedure that uses the measurement of the intensity of the fluorescent radiation emitted by an element having an atomic number between 11 and 92 excited by a continuous primary X-ray radiation. The intensity of the fluorescence produced by a given element depends on the concentration of this element in the sample but also on the absorption by the matrix of the incident and fluorescent radiation. At trace levels, where the calibration curve is linear, the intensity of the fluorescent radiation emitted by an element in a given matrix, at a given wavelength, is proportional to the concentration of this element and inversely proportional to the mass absorption coefficient of the matrix at this wavelength. Method. Set and use the instrument in accordance with the instructions given by the manufacturer. Liquid samples are placed directly in the instrument ; solid samples are first compressed into pellets, sometimes after mixing with a suitable binder. f = dilution factor. 01/2005:20238 2.2.38. CONDUCTIVITY The conductivity of a solution (κ) is, by definition, the reciprocal of resistivity (ρ). Resistivity is defined as the quotient of the electric field and the density of the current. The resistance R (Ω) of a conductor of cross-section S (cm2) and length L (cm) is given by the expression : (3) G. Andermann &M.W. Kemp, Analytical Chemistry 30 1306 (1958). Z.H. Kalman & L. Heller, Analytical Chemistry 34 946 (1962). R.C. Reynolds, Jr., The American Mineralogist 46 1133 (1963). R.O. Müller, Spectrochimica Acta 20 143 (1964). R.O. Müller, Spectrochemische Analyse mit Röntgenfluoreszenz, R. Oldenburg München-Wien (1967). 56 See the information section on general monographs (cover pages)
© Copyright 2026 Paperzz