Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 1 of 8 Percent Lipids Determination References: NOAA Technical Memorandum NOS ORCA 130: Sampling and Analytical Methods of the National Status and Trends Program Mussel Watch Project: 1993-1996 Update. March 1998. 1. Scope and Application Matrices: Tissue sample extractions. Definitions: Refer to Alpha Analytical Quality Manual. This Standard Operating Procedure (SOP) describes the percent lipid determination as adopted by Alpha Analytical. Results derived from this method have been compared, with remarkable accuracy, to nationally published results for both fish and mussel tissue extracts. This method has been developed to determine the percent lipid or total extractable organic (TEO) of any tissue sample. Data derived from this method may be used to assist toxicologists when weighing human health or ecological risk assessments. The data report packages present the documentation of any method modification related to the samples tested. Depending upon the nature of the modification and the extent of intended use, the laboratory may be required to demonstrate that the modifications will produce equivalent results for the matrix. Approval of all method modifications is by one or more of the following laboratory personnel before performing the modification: Area Supervisor, Department Supervisor, Laboratory Director, or Quality Assurance Officer. This method is restricted to use by or under the supervision of experienced analysts. Each analyst must demonstrate the ability to generate acceptable results with this method by performing an initial demonstration of capability, analyzing a proficiency test sample and completing the record of training. After initial demonstration, ongoing demonstration is based on acceptable laboratory performance of at least a quarterly laboratory control sample or acceptable performance from an annual proficiency test sample. A major modification to this procedure requires demonstration of performance. The identification of major method modification requiring performance demonstration is directed by the Quality Assurance Officer and/or Laboratory Director on a case-by-case basis. 2. Summary of Method An aliquot of tissue sample extract is taken, prior to any sample cleanup process;.the aliquot is weighed and the extraction solvent is allowed to evaporate overnight under a fume hood. The aliquot is then re-weighed. Calculations are applied (Section 11) to determine the percent lipid content. If more than one method for organic analysis is being performed, the laboratory Department Manager will select which tissue extract is to be used for percent lipid determination. It is recommended that the method with the largest final volume be chosen. 2.1 Method Modifications from Reference The aliquot of tissue extract was reduced from 20mL to 250-500uL. This eliminates the need to re-concentrate the samples for lipid determination. The intermediate steps of reducing the sample aliquot to total dryness under nitrogen, re-wetting, and re-mixing the sample by adding Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 2 of 8 1mL of DCM, then removing 100uL of sample extract for re-drying for the final lipid weight are replaced by the steps in Section 10.3.3 and 10.3.4. The evaluation of SRMs with known lipid values has proven over time the accuracy of this procedure. 3. Reporting Limits The reporting limit is 0.01% and is largely based on the tolerance of the balance used for the determination. 4. Interferences 4.1 It is important that all weighing be performed at ambient temperatures as hot or chilled items will lose or gain weight with temperature fluctuations. 4.2 The percent lipid determination must be performed on the extract before any cleanup procedures are performed 4.3 All weights must be recorded to four decimal places using an Analytical balance. Small differences in weights can significantly affect the final calculation. 5. Health and Safety The toxicity or carcinogenicity of each reagent and standard used in this method is not fully established; however, each chemical compound should be treated as a potential health hazard. From this viewpoint, exposure to these chemicals must be reduced to the lowest possible level by whatever means available. A reference file of material safety data sheets is available to all personnel involved in the chemical analysis. Additional references to laboratory safety are available in the Chemical Hygiene Plan. All personnel handling environmental samples known to contain or to have been in contact with municipal waste must follow safety practices for handling known disease causative agents. The use of laboratory equipment and chemicals exposes the analyst to several potential hazards. Good laboratory techniques and safety practices shall be followed at all times. All relevant Material Safety Data Sheets (MSDSs) are kept on file. Approved PPE, which includes Safety Glasses, Gloves and Lab Coats, must be worn at all times when handling samples, reagents, chemicals, or when in the vicinity of others handling these items, so that dermal contact is avoided. All standards, reagents and solvents shall be handled under a hood. All flammable solvents must be kept in the flammable storage cabinet, and returned to the cabinet immediately after use. When transporting chemicals, use a secure transporting devise and/or secondary outer container. Chemical storage is properly segregated and adequately ventilated to reduce the possibility of hazardous reactions. Chemical storage in work areas shall be kept to a minimum. Storage on bench tops or other work surfaces, except temporary, is not permitted. All standards and reagents shall be prepared in a hood while using the proper PPE. All additional company safety practices shall be followed at all times as written the Chemical Hygiene Plan. 6. Sample Collection, Preservation, Shipping and Handling 6.1 Sample Collection Not applicable. Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 3 of 8 6.2 Sample Preservation Samples and resulting extracts are stored at -20°C + 10°C. No other preservatives are used. 6.3 Sample Shipping No special sample shipping requirements. 6.4 Sample Handling Samples and resulting extracts shall be stored at -20°C + 10°C. Hold times are not applicable to the percent lipid determination. 7. Equipment and Supplies 7.1 Analytical balance: Capable of weighing to the nearest 0.0001g. The Cahn microbalance is also suitable. 7.2 Aluminum weighing tins 7.3 Fume hood 7.4 Gas-tight Syringes: Various volumes, solvent rinsed with methylene chloride 8. Reagents and Standards 8.1 Sodium sulfate: Granular, anhydrous. Purified by heating to 400°C for 4 hours prior to use. 8.2 Glass wool: Purified by heating to 400°C for 1 hour prior to use. 8.3 Methylene chloride: ACS approved, Pesticide grade. 9. Quality Control The laboratory must maintain records to document the quality of data that is generated. Ongoing data quality checks are compared with established performance criteria to determine if the results of analyses meet the performance characteristics of the method. 9.1 Blank A method blank must be analyzed with every batch. This blank consists of an aluminum weighing tin, with no sample extract added. The result of the method blank must be less than 0.01%. Corrective Action: If the blank result is >0.01%, recalculate the results. If the blank result is still > 0.01%, repeat the tissue extraction for all associated samples and reset the percent lipid determination on the re-extracts. Note: Depending upon the amount of original tissue sample remaining, repeat extractions may not be possible. 9.2 Laboratory Control Sample (LCS) Not applicable to this method. Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 4 of 8 9.3 Initial Calibration Verification (ICV) Not applicable to this method. 9.4 Continuing Calibration Verification (CCV) Not applicable to this method. 9.5 Matrix Spike Not applicable to this method. 9.6 Laboratory Duplicate A Duplicate analysis is performed once per 20 samples, or at least 1 per batch, whichever is more frequent. The RPD between the duplicates must be less than 20% Corrective Action: If the RPD between the duplicates exceeds 20% the analyst shall check the calculations for errors. If no errors are found, repeat the tissue extraction for the batch. Then re-perform the percent lipid determination on the re-extracts. Note: Depending upon the amount of original tissue sample remaining, repeat extractions may not be possible. 9.7 Method-specific Quality Control Samples None. 9.8 Method Sequence • Method Blank • Laboratory Duplicate • Samples 1-19 10. Procedure 10.1 Equipment Set-up The Analytical balance, or Cahn balance, used to make each weighing for this method is checked daily before use with Class 1 weights, in the weight range of use, by a designated analyst, or appointed alternate. These daily checks are documented in the “Balance Calibration Check” logbook. The balance is serviced semiannually by an outside service technician. 10.2 Initial Calibration Not applicable to this method. 10.3 Equipment Operation and Sample Processing 10.3.1 Samples are extracted per the Microscale Solvent Extraction SOP (OP-016) or the Tissue Extraction SOP (OP-018) 10.3.2 Label, weigh and record the weight of an aluminum weighing tin on the Excel spread sheet (see Figure 1), under “weight, t”. Note: Record ALL weights to four decimal places (nearest 0.0001g). NOTE: Record name, date and time as well on Excel sheet 10.3.3 Using a methylene chloride-rinsed 250ul-500ul gas-tight syringe, remove 250ul-500ul of the dried sample extract and gently expel the 250ul-500ul extract into the pre-weighed and labeled aluminum weighing tin from Section 10.3.2, above. Record the volume of the aliquot on the Excel spread sheet under “Sample Aliquot”. 10.3.4 Place the tin under a fume hood overnight to allow for complete solvent evaporation. Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 5 of 8 10.3.5 Re-weigh the tin with the dried extract the following day, and record the weight on the Excel spreadsheet under “weight, g1”. 10.4 Continuing Calibration Not applicable for this method. 10.5 Preventive Maintenance 10.5.1 Analytical and Cahn Balance 11. 10.5.1.1 All balances are calibrated and serviced every six months by an instrument service company. All service records are kept on file. 10.5.1.2 Keep balances clean. Data Evaluation, Calculations and Reporting 11.1 Calculate the Percent Lipids as follows, an example spreadsheet is shown in Figure 1: (W d + a) – W d = Net weight, g (FV / Va) x Net weight x 1000 = TEW TEW / W ext = TEC TEC / 10 = % Lipid where: Wd + a = Weight of the weighing tin and the extract, in grams, as recorded in Section 10.3.6. Wd = Weight of the weighing tin, in grams, as recorded in Section 10.3.2. W ext = Weight of tissue extracted, in grams, wet, (from extraction logbook). FV = Final volume of the extract, in microliters, (i.e., 58000uL = 58.0mL). Va = Volume of the extract aliquot recorded in Section 10.3.3, in microliters (i.e., 5000uL = 5.0mL). TEW = Total Extractable Weight TEC = Total Extractable Concentration 11.2 Procedures for data and record management must adhere to the Quality Systems Manual, other subordinate documents covering record keeping, and the Document Control SOP, (SOP/08-01). All records shall be stored in such a manner as to be safe and accessible for at least 10 years. 11.3 Notebooks: Laboratory notebooks are designed to accommodate the specific analysis. If a sample requires re-analysis or re-extraction for any reason, a notation is made next to the sample entry in that log. Such logbooks shall be archived so as to remain available for at least 10 years. All laboratory notebooks must follow the specifications in the Laboratory Notebook Usage Work Instruction, 108-01, and follow all record keeping and document control practices. Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 6 of 8 11.4 To report data, go to the LIMS and “batch” and “associate” all samples in this analytical batch. 11.5 Enter the results obtained from the Excel spreadsheet. 11.6 Forward the hand recorded raw data from the Excel spreadsheet and the LIMS batch printout to the Department Manager for secondary review. 11.7 All results are reportable without qualification if the QC criteria defined in Section 9 are met. If any QC parameters are not met, all associated samples must be evaluated for re-extraction and/or re-analysis. See section 9 for discussion, including corrective action. 12. Contingencies for Handling Out-of-Control Data or Unacceptable Data Corrective Action for non-compliant data is discussed in Section 9. If a situation arises (such as loss of sample, or the inability for re-extraction or re-analysis) that requires additional action, the Project Manager and/or Laboratory Director are immediately notified and the client is contacted for discussion and/or instruction for any further action. The situation and the agreed upon actions will be discussed in the project narrative that accompanies the data package. 13. Method Performance 13.1 Method Detection Limit Study (MDL) / Limit of Detection Study (LOD) / Limit of Quantitation (LOQ) The laboratory follows the procedure to determine the MDL, LOD, and/or LOQ as outlined in Alpha SOP/08-05. These studies performed by the laboratory are maintained on file for review. 13.2 Demonstration of Capability Studies Refer to Alpha SOP/08-12 for further information regarding IDC/DOC Generation. 13.2.1 Initial (IDC) The analyst must make an initial, one-time, demonstration of the ability to generate acceptable accuracy and precision with this method, prior to the processing of any samples. 13.2.2 Continuing (DOC) The analyst must make a continuing, annual, demonstration of the ability to generate acceptable accuracy and precision with this method. 14. Pollution Prevention and Waste Management All solvent waste generated during lipid analysis must be stored in satellite containers in the preparation laboratory labeled “Organic Solvent”. Once the organic solvent satellite containers are full, they must be emptied into a 55-gallon drum marked “Organic Solvent Waste”. Refer to Alpha’s Chemical Hygiene Plan and Waste Management and Disposal SOP for further pollution prevention and waste management information. Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination 15. ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 7 of 8 Referenced Documents SOP/OP-004 Extraction of Soil, Tissue, Vegetation and Sediment Samples by Pressurized Fluid Extraction WI/108-01 Laboratory Logbook Usage Chemical Hygiene Plan SOP/01-01 Sample Receipt and Login SOP/01-02 Sample Custody and Tracking SOP/08-01 Document Control SOP/08-05 MDL/LOD/LOQ Generation SOP/08-12 IDC/DOC Generation SOP/G-006 Waste Management and Disposal SOP 16. Attachments Figure1: Example Excel Spreadsheet Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015 Alpha Analytical, Inc. Facility: Mansfield Department: Organic Extractions Title: Percent Lipids Determination ID No.:2171 Revision 5 Published Date:3/12/2013 11:38:20 AM Page 8 of 8 Figure 1 Example Excel Spreadsheet Method #: LIPIDS Set up Date: 8/20/02 Analyst: TLB QC Sample ETR# No. BLANKTINB 207077 16 Sample Lipid Final Sample Extract Dish Weight Dilution Volume Aliquot Aliquot D No. (wet) Factor (mL) (uL) (uL) B 16 5.00 5.42 1.09 1.09 59 60 5000 5000 0.00 5000 0.00 5000 0.00 5000 0.00 5000 0.00 5000 0.00 5000 0.00 5000 0.00 5000 (g) weight g1 0.9565 59000 t 0.9565 n1 0.0000 g1 0.9710 60000 t 0.9567 n1 0.0143 g1 0 t n1 0.0000 g1 0 t n1 0.0000 g1 0 t n1 0.0000 g1 0 t n1 0.0000 g1 0 t n1 0.0000 g1 0 t n1 0.0000 g1 0 t n1 0.0000 g1 0 t n1 0.0000 TOTAL TOTAL EXTRACTABLE EXTRACTABLE WEIGHT CONC. % LIPIDS n=net solid Calculations t=tare wt. g=gross solid wt. calculations: 0.00 0.00 0.000 n=g-t TOTAL Ext Wt =((ext. vol in uL)/(aliquot vol in uL))*(n1*1000) TOTAL Ext Conc = (Total Ext Wt)/Sample Wt 171.60 31.66 3.166 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! 0.00 #DIV/0! #DIV/0! %Lipids = Total Ext. Conc. / 10 TEMPLATE: lipid.XLT Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work, the published version of the document should be viewed online. Document Type: SOP-Technical Pre-Qualtrax Document ID: OP-015
© Copyright 2025 Paperzz