An Evaluation of the Vita-Stat Automatic Blood Pressure Measuring Device B. FRANK POLK, M.D., M. S c , BERNARD ROSNER, P H . D . , R U D Y FEUDO, M. E D . , AND MARTIN VANDENBURGH, B.A. Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 SUMMARY Four Vita-Stat (VS) automatic, coin-operated, blood pressure measuring derices were evaluated for accuracy and precision. Under field conditions, 342 adults had two blood pressure measurements with both the VS device and with the Random-Zero (RZ) device. Two of the VS machines gave significantly higher systolic (SBP) and diastolic blood pressure (DBP) values, compared to the RZ values. The mean difference between these two units and the RZ device were clinically important (14.0 and 6.9 mm Hg SBP; 7.5 and 6.8 mm Hg DBP), and resulted hi the misclassiflcation of 23% of normotensives as hypertensives. We observed a significant order effect on SBP with the VS device, with a mean decrement of 4.9 mm Hg between the first and second SBP values, compared to 0.6 mm Hg with the RZ device. Even after adjusting for this bias, all four VS devices gave significantly more variable SBP readings than the RZ unit; two of the four also gave more variable DBP values. These data suggest that the VS device is neither accurate nor precise enough at the present time to be recommended for widespread use. These findings also raise questions concerning the monitoring of performance of this and similar devices in the field. (Hypertension 2: 221-227, 1980) KEYWORDS • blood pressure measuring device • Vita-Stat device R • Random-Zero device our knowledge.1 We have evaluated four of these machines, under field conditions, for accuracy and precision. ELIABLE automatic blood pressure measuring devices may have an important role in the detection of elevated blood pressure and perhaps in improving compliance with therapy among individuals with known hypertension. Recently, a coin-operated device called the Vita-Stat Computer (Vita-Stat Medical Services, Tierra Verde, Florida) has been distributed to many pharmacies, department stores, banks, shopping centers, and hotels across the United States. This device is fully automatic and uses the standard auscultation (Korotkoff sound) principle of measurement. Despite the claim of the company that "Independent research studies by leading medical authorities have proven the accuracy of the Vita-Stat computer," only one such study has been published, to Patients and Methods During December, 1978, adult customers (ages 20 and over) in one pharmacy, one department store, and two shopping centers were invited to have their blood pressures measured. A list of 61 locations in northern New England was provided by the company, which was not aware of the evaluation program. These four locations were chosen from the list because they were in the greater Boston area, and because it was estimated that their customers would be reasonably representative of the general adult population of New England, and would be sufficiently numerous so that approximately 100 persons could be screened at each site in a short period of time. The Vita-Stat device (VS) measures systolic and diastolic blood pressures and is based on the auscultation principle. The customer places his left arm in a From the Channing Laboratory and Departments of Medicine and Social and Preventive Medicine, Harvard Medical School; and the Peter Bent Brigham Hospital Division of the Affiliated Hospitals Center, Inc., Boston, Massachusetts. Address for reprints: B. Frank Polk, M.D., Channing Laboratory, 180 Longwood Avenue, Boston, Massachusetts 02115. Received August 9, 1979; revision accepted October 30, 1979. 221 222 HYPERTENSION Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 looped cuff so that when the cuff closes automatically it will be placed starting at the elbow. A microphone is mounted in the cuff such that it is near the brachial artery when the arm is inserted. When the start button is pressed, the cuff tightens snugly by use of a slip clutch and inflates to 160 mm Hg. If Korotkoff sounds are sensed at that pressure, inflation continues in 20 mm Hg increments until Korotkoff sounds are no longer sensed or a maximum pressure of 220 mm Hg is reached. The cuff then deflates in 4-6 mm Hg decrements with a brief pause after each decrement to determine the presence or absence of Korotkoff sounds. Systolic blood pressure (SBP) is registered at the appearance of Korotkoff sounds and diastolic blood pressure (DBP) at their disappearance. There is an artifact rejection system built into the machine so that extraneous noise and false sounds are filtered out or rejected. The blood pressure values recorded by VS were compared to those obtained by a trained observer using the modified Random-Zero device (RZ) (Hawksley, Sussex, England). This device is similar in principle to the standard mercury sphygmomanometer. A mechanical device interposed between the mercury reservoir and manometer column, however, draws variable amounts of mercury into an expandable chamber. This modification has the effect of obscuring the true zero level of mercury until the end of the reading, thus minimizing observer bias.1 Labarthe et al.' have shown good agreement between overall mean SBP and DBP as determined by the RZ device compared to a standard mercury sphygmomanometer (Baumanometer 300), although RZ values tended to be slightly lower. All data were collected by one screener who had extensive training and experience in blood pressure screening programs. Training and standardization included orientation to physiologic aspects of systemic arterial blood pressure, to the techniques of its indirect measurement, and to familiarization with the modified RZ. This screener was trained and tested by performing multiple readings with a double stethoscope on several subjects, and was standardized by a technique based on audio cassette tapes described by Rose.4 A brief questionnaire was administered before the blood pressures were obtained. Data collected included age, sex, race, history of hypertension, and antihypertensive treatment. When the participant was sitting, four blood pressure measurements on the left arm were taken: two with the VS and two with the RZ. For the RZ measurements, the appropriate cuff size was chosen on the basis of arm circumference. The devices were used in an alternating sequence, and the order of the sequence (VS-VS-RZ-RZ vs RZ-RZ-VS-VS) was determined by using a table of random numbers. The 50 cents required for each VS measurement was supplied by the investigators. The RZ device was calibrated between sites. A paired t test was used to compare mean blood pressures obtained with the VS and RZ machines.' VOL 2, No 2, MARCH-APRIL 1980 This method also was used to compare the order effect (first reading minus second reading) for the two devices. The measure of variability chosen was the mean absolute difference between replicates after correcting for the above order effect = 8x (5y) = | XH — X21 - Ax | ( | Yu - Y2I - Ay | ), where X,,, X21 and Y u , Y21 are the first and second readings for the VS and RZ devices respectively, and Ax (Ay) = %n XJI (Yn — Y21). For each location, Ax and Ay were computed separately. A paired t test was then used to compare 8\ and 5y based on the statistic 8x— 8y = I X,, - X21 - Ax | - | Y,, - YM - Ay | . Results Patient Characteristics A total of 412 individuals were interviewed: 106, 101, 105, and 100 from locations A-D respectively. Individuals were excluded from analysis if they were missing one or more systolic or diastolic readings on either machine. There were no missing RZ readings, but 70 persons (17.0%) were excluded from analysis because of missing VS readings (8, 17, 7, and 38 respectively). In each instance of missing values, the VS read out blank readings and returned the coins. A third measurement was not attempted in these cases. Characteristics of the 342 persons for whom blood pressure data were complete are presented in table 1. The study population was predominantly white (92.4%), had a median age of 57 years, and was approximately half men and half women. Of interest, nearly half (47.4%) had been told in the past by a physician that their blood pressures were elevated, over one-third (36.4%) had at one time been treated for hypertension, and 29.0% were taking antihypertensive medication(s) at the time of screening. Accuracy The mean systolic and diastolic blood pressures are presented by location in table 2. At two locations there were negligible, nonsignificant differences between VS and RZ blood pressure readings. Two of the VS machines, however, gave notably higher mean SBP (14.0 and 6.9 mm Hg) and DBP (7.6 and 6.8 mm Hg) measurements compared to those obtained with the RZ. For these two units, the differences were significant, with p values < 0.001. The frequency distributions of these differences are given in table 3. At locations C and D, the mean of two systolic readings with the VS was > 20 mm Hg higher than the RZ in 28.6% and 9.7% of participants respectively; the mean of two diastolic readings with the VS was > 10 mm Hg higher than the RZ in 40.8% and 37.1% of participants respectively. The potential effects on sensitivity and specificity if the VS readings were used to diagnose high blood presure are shown in table 4. We assumed that the "correct" diagnosis is given by the RZ and that high blood pressure is defined (arbitrarily) as SBP 160 mm Hg and/or DBP 95 mm Hg, because these are the AUTOMATIC BLOOD PRESSURE MEASUREMENT/Po/fc and Rosner 223 TABLE 1. Characteristics of S4£ Participants: Age, Sex, Race, and Hypertension Status n % 20-29 30-39 40^9 50-59 60-69 70-79 80-89 39 39 54 66 100 38 6 11.4 11.4 15.8 19.3 29.2 11.1 1.8 Sex distribution (n = 342) Men Women 164 178 48.0 52.0 Race distribution (n = 342) White Black Other 316 23 3 92.4 6.7 0.9 History of hypertension (n - 340) Yee No 161 179 47.4 52.6 Previously taken antihypertensive Yes medication (n = 338) io 123 215 36.4 63.6 Currently taking antihypertensive Yes medication (n = 338) io 98 240 29.0 71.0 Characteristic Age distribution (n = 342) Overall - 53.0 «= 16.1 median = 57 range = 20-89 X 8D Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 TABLE 2. Mean Blood Pressures and Differences Between Vita-Stat and Random-Zero Device Readings at Four Different Locations Location Random-Zero Vita-Stat Difference* t statistic p value Systolic blood pressure (mm Hg) n A B C D 98 84 98 62 X 142.5 134.1 147.9 135.4 SD X 8D S 8D 21.0 22.5 20.3 16.7 142.0 133.6 133.9 128.5 18.1 23.2 18.3 19.0 0.5 0.5 14.0 6.9 11.2 12.1 11.7 13.6 0.475 0.405 11.909 3.982 n.s. n.s. < 0.001 < 0.001 8D t statistic -0.650 0 7.871 5.825 p value n.s. n.s. < 0.001 < 0.001 Diastolic blood pressure (mm Hg) n S A 98 B C D 84 83.9 79.7 89.0 87.3 98 62 SD 12.6 13.1 12.0 13.3 i 85.2 79.7 81.4 80.4 8D if 11.3 12.0 9.9 10.9 -1.3 0.0 7.6 6.8 6.4 8.6 9.5 9.2 *A paired means t test was used to analyze differences between the means of two machines. cutoff points chosen for the definition of definite hypertension by the U. S. Public Health Service" as well as by the distributors of the VS. Overall, 23.3% of the normotensive participants were incorrectly classified as hypertensive, and 16.4% of the hypertensive participants were incorrectly classified as normotensive. Precision As shown in table 5, we observed a significant and noteworthy order effect on SBP values with the VS. That is, there was a significant decrement between the first and second VS SBP readings (mean difference = 4.88 mm Hg;/? < 0.001) compared to a small and nonsignificant decrement between the first and second RZ SBP readings (mean difference = 0.58 mm Hg; p = not significant). The difference in order effect between the VS and RZ was about twice as great when the VS recordings are first and second (overall mean difference = 5.72 mm Hg) as compared to when the RZ recordings are first and second (overall mean difference = 2.83 mm Hg). The order effect on systolic readings, however, is significantly larger for VS in HYPERTENSION 224 VOL 2, No 2, MARCH-APRIL 1980 TABLE 3. Frequency Distribution of Differences Between Mean Values of the VitaStat Replicates and RandomZero Replicates Mean of 2 VS readings minus mean of 2 RZ readings* B A n n % D C n % % n % 2 3.2 Systolic blood pressure (mm Hg) <-20 >-20, < - 1 0 >-10, < 0 > 0, < 10 > 10, < 20 3 3.1 4 4.8 1 1.0 14 14.3 28.6 36.7 15.3 15 0 0.0 3 4.8 4 4.1 9 33 22 14.5 32.3 35.5 4 4.8 33.7 32.7 28.6 20 10 17.9 25.0 35.7 11.9 6 9.7 28 36 15 2 >20 21 30 2.0 98 32 28 84 98 62 Diastolic blood pressure (mm Hg) I Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 >-15, >-10, > - 5, > 0, > 5, > 10, <-10 <- 5 0 < 5 < < 10 < 15 2 2.0 3 3.6 3 3.0 1 1.6 6 6.1 7 8.3 2 2.0 1 1.6 19 13 4.1 4 6.5 6 6.1 8. 22 13 15.5 16.7 28.6 15.5 4 10 19.4 31.6 24.5 10.2 6 6.1 8 9.5 20 0 0.0 2 2.4 20 22.4 21.4 20.4 20.4 12.9 19.4 21.0 21.0 16.1 31 24 >15 14 24 21 84 98 12 13 13 10 98 62 •VS = Vita-Stat device; RZ = Random-Zero device. TABLE 4. Sensitivity and Specificity of Diagnosis of Hypertension Made by the Vita-Stat Device, Using Criteria of Mean SBP > 160 mm Hg and/or DBP > 95 mm Hg Mean RZ <160 — (RZ normotensive) Mean RZ ^ , . and/or (RZ hypertensive) •^ y o Location A B C D All Mean VS _ „_ Mean VS n r , and/or n Specificity < 95 95+ 60 10 70 85.7 62 67 92.5 5 60.2 83 50 33 39 16 55 70.9 211 64 275 76.7 160+ 160+ Mean VS < 95 nr. Mean VS „? 95+, and/or n Sensitivity 5 23 28 82.1 4 13 17 76.5 93.3 1 14 15 1 85.7 6 7 11 56 67 83.6 Abbreviations: SBP = systolic blood pressure; DBP =• diastolic blood pressure; RZ = Random-Zero device; VS = Vita-Stat device. both instances. The sequence of recordings began with the VS in 52% of participants. There was not an important order effect on DBP measurements with either device. Finally, the variability of each VS machine is compared to the RZ in table 6. At all 4 locations, SBP measurements on the VS were significantly more variable than the RZ counterparts. Two of the VS machines also gave significantly more variable DBP measurements compared to the RZ. Discussion Accuracy As one very important performance characteristic of automatic blood pressure measuring devices, AUTOMATIC BLOOD PRESSURE MEASUREMENT/Po/Jt and Rosner 225 TABLE 5. Comparison of Difference Between First and Second Blood Pressure Readings with VitaStat and Random-Zero Devices Diastolic blood pressure I statistic p value SD Systolic blood pressure t statistic p value SD Location n X A B C D 98 4.50 4.49 5.97 4.27 4.88 10.38 12.07 12.34 11.09 11.53 1.20 0.00 0.57 0.39 0.58 8.04 7.33 8.02 7.05 7.70 1.48 0.00 0.70 0.44 1.39 3.30 4.49 5.40 3.89 4.30 12.77 15.06 15.30 13.33 14.22 2.56 2.73 3.49 2.30 5.59 Vita-Stat All 84 98 62 342 4.29 3.41 4.79 3.03 7.83 0.59 0.38 1.77 1.16 0.98 5.28 6.48 8.89 11.38 8.05 1.11 0.54 1.97 0.80 2.25 0.65 -0.10 0.29 0.36 0.31 4.47 4.83 4.31 5.13 4.65 1.44 -0.19 0.67 0.55 1.19 n.s. n.s. n.s. n.s. n.s. 7.15 8.10 9.66 12.74 9.34 -0.08 0.54 1.52 0.50 1.33 n.s. n.s. n.s. n.s. n.s. < 0.001 0.001 < 0.001 0.004 < 0.001 n.s. n.s. n.s. n.s. 0.024 Random-Zero A B C D All 98 84 98 62 342 Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 A 98 B C D 84 98 All 342 62 n.s. n.s. n.s. n.s. n.s. Vita-Stat—Random-Zero -0.06 0.012 0.48 0.008 1.48 < 0.001 0.81 0.025 0.67 < 0.001 TABLE 6. Variability of SBP and DBP Measurements of the Vita-Stat Machine Compared to the RandomZero Device (A = VS-RZ) Location Mean A A 2.33 2.51 2.55 2.65 Systolic blood pressure (mm Hg) 7.97 98 10.06 84 9.44 98 8.19 62 2.89 2.29 2.67 2.55 0.29 0.81 3.05 4.83 Diastolic blood pressure (mm Hg) 4.20 98 84 5.53 6.77 98 62 8.89 0.68 1.34 4.46 4.28 B C D A B C D SD measurements of SBP and DBP should closely approximate readings obtained by adequately trained observers using mercury sphygmomanometers. 8 ' 7 Evaluation of accuracy should be performed under field conditions and on multiple units of the device. Two of the four automatic units evaluated in this study produced mean SBP and DBP readings that were significantly and importantly higher than paired n t statistic p value 0.005 0.025 0.009 0.013 TIM. n.s. < 0.001 < 0.001 measurements obtained with the RZ. These mean differences did not result from a few extreme discrepancies; these two devices yielded a mean SBP or DBP that was > 10 mm Hg higher in approximately 60% and 40% respectively of the paired comparisons. It is unlikely that the discrepancies resulted from systematically lower measurements determined by the trained observer, since the RZ minimizes observer 226 HYPERTENSION Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 bias,2 and since there was close agreement between the RZ and VS in two locations. Inadequate maintenance may have contributed to the inaccuracies observed, but one of the faulty VS was serviced unexpectedly in the middle of the evaluation process, and it was observed that the subsequent VS readings remained generally higher than the paired readings. The number of VS devices (four) evaluated in this study was small. The locations were not chosen randomly, but rather on the basis of convenience and estimated representativeness of their customer populations. Thus, it is possible that our findings are not representative. However, they are comparable to those of Berkson et al.1 who compared blood pressure values determined by eight randomly chosen VS machines with those obtained by two trained screeners. The number of participants (10-34) at each site was small, but three of the eight machines gave significantly higher systolic readings than those of the screeners, and two gave significantly higher diastolic readings. They observed that the average machinehuman difference in SBP varied from -5.7 to 12 mm Hg, and concluded that the VS SBP determinations " . . . may be unduly sensitive to malfunction or maladjustment . . . " These investigators also found that machine-human agreement on blood pressure classification was somewhat lower than the humanhuman agreement. We compared the VS to one highly trained screener. It would be of great interest to compare the interobserver variability between two human observers — with and without special training — to the variability between two VS machines. Precision There was a striking order effect in SBP measurements with all four VS devices; that is, the first VS SBP reading was significantly and notably higher than the second (approximately 5 mm Hg higher, on average) on all four machines. This was especially true if the first measurement in the sequence of four was obtained on the VS device. In contrast, there were small and generally nonsignificant order effects with the RZ SBP readings, and with either device with regard to DBP readings. Order effects in this direction have been reported previously*' * and are assumed to be due to adaptation. The notably greater order effect on SBP here observed with the VS device may be related to greater apprehension experienced while placing one's arm into an entrapping machine than that experienced with the cuff of a sphygmomanometer that allows mobility. While it was not done in this study, future evaluations should record whether or not participants have used the device previously. It should be remembered that this order effect occurred while a screener was present; it is at least possible that blood pressures might be even more elevated when the average person engages the machine alone. The importance of this phenomenon increases when one realizes that most individuals are not likely to buy two VOL 2, No 2, MARCH-APRIL 1980 consecutive measurements within a short time and, indeed, are not advised to do so by the literature found on the machine. The major consequence of this order effect is on false positive misclassification. Our estimate of 23.3% of normotensives misclassified as hypertensive was based on the mean of two VS measurements, and thus are conservative. Even after accounting for this order effect, the variability of SBP values in the VS readings was consistently and significantly greater than the variability in the RZ measurements. As shown in table 6, the absolute differences in mean replicate BP readings after subtracting the bias (order effect) from the VS and RZ values are significantly greater for all four VS machines than their RZ counterparts with respect to SBP and for two machines with respect to DBP. Implications The consequences of spuriously elevated blood pressure measurements depend on the actions taken by the individuals in question. If measurements are made for screening purposes, i.e., evaluation of persons with unknown blood pressure status, then misclassification as hypertensive may result in unwarranted anxiety in individuals without sustained elevations in blood pressure, and misclassification as normotensive may cause a false sense of security and delay accurate diagnosis and proper treatment. While a false positive misclassification of hypertension may be preferable to a false negative misclassification, it clearly is better to minimize all classification errors. Over-diagnosis is not only likely to cause apprehension, but is likely to result in unnecessary visits to physician's offices for confirmation or, more likely, refutation of the diagnosis of hypertension. Although the data are sparse and conflicting, there have been suggestions that labelling persons as hypertensive may have adverse psychological and behavioral consequences.10 These consequences presumably are not dependent on accurate diagnosis but simply on the label itself. If measurements are made for the evaluation of blood pressure status in persons known to be hypertensive, then the result is misassessment of blood pressure control. Despite instructions that accompany the machine, some individuals may decide to change their antihypertensive medications on their own initiative, with the possibility of adverse consequences. At best, spuriously high readings are likely to cause undue anxiety and unnecessary visits to the physician. Ironically, this automatic blood pressure measuring device may be cost-ineffective by increasing the number of physician visits. There is, perhaps, a more important question raised by these data: who is responsible for monitoring the performance of these commercial machines with respect to validity and repeatability? How and how often is performance to be monitored? The data presented here suggest that the maintenance and/or calibration provided by the distributor are inadequate. AUTOMATIC BLOOD PRESSURE MEASUREMENT/i»o/Jt and Rosner Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 If performance has to be monitored closely by regulatory governmental agencies, the devices may not be cost-effective, even if accurate. Our observation that two of four machines tested in the field were inaccurate implies that all machines would have to be monitored carefully and continuously. A possible alternative would be to restrict distribution of devices to a few locations where they could be carefully and efficiently monitored. The economics of visits to the physician for blood pressure determinations make a fully automatic blood pressure measuring machine highly attractive. Such a device could have an important role in the detection and management of high blood pressure, and might improve compliance with therapy among some individuals just as self-recording of blood pressures has been shown to do. In evaluating the effect of selfrecording of blood pressure, Johnson et al. 11 demonstrated a reduction in blood pressure among hypertensive patients who admitted to difficulty remembering to take their pills. We agree with a recent editorial in Lancet12 that these devices are to be encouraged " . . . provided that the measurement is reasonably accurate, that the instructions given are sensible, and that doctors cooperate by checking the blood pressure themselves . . . " The data presented here, however, suggest that the Vita-Stat device currently is not "reasonably" accurate, and its use is not to be encouraged at this time. There is a need for further evaluation of more specimens of this device. 227 References 1. Berkson DM, Whipple IT, Shireman L, Brown MC, Raynor W, Shekelle RB: Evaluation of an automated blood pressure measuring device intended for general public use. Am J Public Health 69: 473, 1979 2. Wright BM, Dore CF: A random-zero sphygmomanometer. Lancet 1: 337, 1970 3. LaBarthe DR, Hawkins CM, Remington, RD: Evaluation of performance of selected devices for measuring blood pressure. Am J Cardiol 32: 546, 1973 4. Rose GA: Standardization of observers in blood pressure measurement. Lancet 1: 673, 1965 5. Snedecor G, Cochran WG: Statistical Methods, 6th ed. Ames, Iowa, Iowa State University Press, 1967, p 93 6. U.S. Department of Health, Education and Welfare. PHS. Advance Data from Vital and Health Statistics. No. 1, Washington, D.C., U.S. Government Printing Office, October, 1976 7. Feinlieb M, LaBarthe D, Shekelle R, Kuller, L: Criteria for evaluation of automated blood pressure measuring devices for use in hypertensive screening programs. Circulation 49:6, 1974 8. Wilber JA, Millward D, Baldwin A, Capron B, Silverman D, James LM, Wolbert T, McCombs NJ: Atlanta community high blood pressure program methods of community hypertension screening. Circ Res 30 (suppl II): 11-101, 1972 9. Hypertension Detection and Follow-Up Program Cooperative Group: Variability of blood pressure and the results of screening in the Hypertension Detection and Follow-Up Program. J Chronic Dis 3 1 : 651, 1978 10. Haynes RB, Sackett DL, Taylor DW, Gibson ES, Johnson AL: Increased absenteeism from work after detection and labeling of hypertensive patients. N Engl J Med 299: 741, 1978 11. Johnson AL, Taylor DW, Sackett DL, Dunnett CW, Shimizu AG: Self-recording of blood pressure in the management of hypertension. Can Med Assoc J 119: 1034, 1978 12. Editorial: Who should measure the blood pressure? Lancet 1: 137, 1979 An evaluation of the Vita-Stat automatic blood pressure measuring device. B F Polk, B Rosner, R Feudo and M Vandenburgh Hypertension. 1980;2:221-227 doi: 10.1161/01.HYP.2.2.221 Downloaded from http://hyper.ahajournals.org/ by guest on June 14, 2017 Hypertension is published by the American Heart Association, 7272 Greenville Avenue, Dallas, TX 75231 Copyright © 1980 American Heart Association, Inc. All rights reserved. Print ISSN: 0194-911X. Online ISSN: 1524-4563 The online version of this article, along with updated information and services, is located on the World Wide Web at: http://hyper.ahajournals.org/content/2/2/221 Permissions: Requests for permissions to reproduce figures, tables, or portions of articles originally published in Hypertension can be obtained via RightsLink, a service of the Copyright Clearance Center, not the Editorial Office. Once the online version of the published article for which permission is being requested is located, click Request Permissions in the middle column of the Web page under Services. Further information about this process is available in the Permissions and Rights Question and Answer document. Reprints: Information about reprints can be found online at: http://www.lww.com/reprints Subscriptions: Information about subscribing to Hypertension is online at: http://hyper.ahajournals.org//subscriptions/
© Copyright 2026 Paperzz