Vapor Pressure and Freezing Point Osmolality Measurements Applied to a Volatile Screen EDWIN J. DRAVIAM, PH.D., EDWARD M. CUSTER, PH.D., AND IRWIN SCHOEN, M.D. This is a report of a rapid and precise screening procedure, developed for the determination of ethanol in serum using osmolality measurements. The osmolality of the patient is determined by freezing point method (freezing point osmometry) and dew point (water vapor pressure osmometry) method. The difference between freezing point osmolality and vapor pressure osmolality (A osm) is due to the presence of volatiles in the serum, because the volatiles are not measured by vapor pressure osmometry. The amount of ethanol (mg/dL) in serum is estimated by multiplying A osm by a factor of 4.2. As. a comparison method, ethanol also is measured by a spectrophotometry alcohol dehydrogenase method. In addition, a significant difference between an osmometric alcohol assayed value and enzymatic spectrophotometric measurement indicates the presence of volatiles, other than ethanol. In addition to ethanol there is a linear relationship between osmolality and isopropanol or methanol when added in vitro to serum. (Key words: Osmolality; Volatiles; Ethanol; Isopropanol; Methanol) Am J Clin Pathol 1984; 82: 706-709 THE ELEVATION of serum osmolality due to the presence of ethanol has been reported.7-8 The measurement of freezing point depression is directly proportional to osmolality. However, ethanol and other water-soluble volatiles (volatile alcohols, ketones, or aldehydes) are not detected when one uses a vapor pressure osmometer in which the measurement is based on dew point depression. When one uses both types of determinations, the difference between freezing point and vapor pressure osmolality is an indication of the presence of watersoluble volatiles. The ingestion of ethanol has been found to be the most common cause of increased serum osmolality in healthy subjects.8 The elevation of serum osmolality due to the presence of other volatiles and nonvolatiles such as isopropanol, methanol, ethylene glycol, and propylene glycol has been reported. 1 ' 5 ' 61012 Using both freezing point depression and dew point measurements in our laboratory, we have established linear relationships between ethanol, isopropanol, and methanol and A osmolality. The osmometric ethanol assay procedure presented in this article may be used as an effective screening procedure to determine whether a gas chromatographic Received November 18, 1983; received revised manuscript and accepted for publication June 1, 1984. Address reprint requests to Dr. Schoen: Clinical Laboratory, Department of Pathology, University of Texas Medical Branch, Galveston, Texas 77550. 706 Department of Pathology, Clinical Chemistry Division, The University of Texas Medical Branch, Galveston, Texas analysis for volatiles is necessary. This procedure is especially suitable for small and medium-sized hospitals, where osmometers are used routinely but riot gas chrorhatography (GC). Additional equipment usually is not required, because many hospitals have both freezing point and vapor pressure osmometers. The most recent osmometers (Wescor Model 5100B®, List $3,195.00, Wescor, Inc., Logan, UT 84321 and \i Osmette® Model 5004, List $3,495.00, Precision Systems, Inc., Sudsbury, MA 01776) are compact in size and require as little as 0.7 and 0.5 square foot of bench space, respectively. The procedure is simpie to perform on a 24-hour basis and does not require special skills. Materials and Methods Apparatus For the freezing point osmometry, an Advanced Cryomatic Osmometer® (Model 3CII, Advanced Instruments, Needham Heights, MA 02194) and a n Osmette Osmometer (Model 5004) were used. A Wescor Model 5100B Vapor Pressure Osmometer was used for water vapor pressure osmometry. Enzymatic ethanol determinations were carried out on DuPont aca® (DuPont Company, Wilmington, DE 19898). Standards Calibration standards and assayed run controls were obtained from Advanced Instruments, Inc., and from Wescor, Inc. Working standards containing 79, 158, 237, and 316 mg of absolute ethanol per deciliter were made in pooled serum and in 0.9% NaCl solution. Isopropanol standards containing 78, 156, 234, and 302 mg/dL in 0.9% NaCl were prepared. Methanol standards containing 16, 32, 79, and 157 mg/dL also were made in 0.9% NaCl. Procedures The Advanced Cryomatic Osmometer was calibrated with the use of 100 and 900 mOsm/kg standards. The Vol. 82 • No. 6 OSMOLALITY MEASUREMENTS APPLIED TO VOLATILE SCREEN Wescor Vapor Pressure Osmometer was calibrated using 300 and 1,000 mOsm/kg standards. The 300 mOsm standard was prepared in our laboratory by dissolving 9.429 g NaCl in 1 L distilled deionized water." Ortho® Normal and Abnormal Unassayed Serum Controls (Ortho Diagnostics, Inc., Raritan, NJ 08869) were used as daily run controls. Clinitrol 290® (Advanced Instruments, Inc.) was employed as an assayed accuracy control on both instruments. A sample size of 300 /xL was used for Advanced Cryomatic Osmometer, and 50 nL was used for /x Osmette Osmometer. The samples and standards were transferred into sample cuvettes with an Eppendorf® pipette (Brinkmann Instruments, Inc., Westbury, NY 11590). In vapor pressure osmometry, an S-fiL sample was delivered using a Drummond® pipette (Drummond Scientific Company, Broomall, PA 19008) on a sample filter paper disc placed in the sample chamber. A 10% decrease in volume of the sample did not change the osmolality of a control serum by freezing point and vapor pressure osmometry, while a 10% increase in volume resulted in 1% decrease in osmolality. The difference between freezing point and vapor pressure osmolality measurements by the two instruments represents delta (A) osmolality. The A osmolality was determined for ethanol, isopropanol, and methanol standards and plotted against the respective alcohol concentration. Osmolality was determined on patient serum specimens with significant alcohol levels. Venous blood was drawn from these patients to completely fill a 10-mL red-top Vacutainer® (Becton-Dickinson and Company, Rutherford, NJ 07070) tube with less than 1 cm of space between the top of the blood and the bottom of the stopper. The clotted tubes were centrifuged for 5 minutes using a Model IEC Spinette® centrifuge (Damon/ IEC Division, Needham Heights, MA 021894). Serum promptly was removed from the erythrocytes, and the ethanol level was measured on Dupont aca, using an alcohol dehydrogenase (EC 1.1.1.1) method. Two B-D Microtainer® tubes (Becton-Dickinson) were filled completely with serum, capped, and kept frozen until the osmolality determination. After thawing the stoppered tubes at room temperature, the contents were mixed thoroughly by inversion ten times before sampling for osmolality determinations. For osmolality determination, serum is reported to be stable for 2 hours at room temperature and for several days in the refrigerator.9 To study the stability of serum ethanol at room temperature, a control serum was spiked with ethanol (130 mg/dL) and kept frozen in aliquots. The aliquots were thawed at room temperature, and the osmolality was determined at various time intervals (0, 0.5, 1, 1.5, 2, 3, 4, 6 hours). The 707 Table 1. Serum Ethanol A Osmolality Storage Stability at Room Temperature (23°C) Time (h) Freezing Point* Osmolality (mOsm) Vapor Pressure* Osmolality (mOsm) 0 0.5 1 1.5 2 3 4 6 303 303 303 303 303 303 303 303 279 279 276 278 278 281 282 282 * Determinations were carried out in duplicates. results are shown in Table 1. The SDs of freezing point and vapor pressure measurements were within the SDs calculated for the respective methods. The osmolality was found to be stable at room temperature for at least six hours for both freezing point and vapor pressure measurements. It has been reported that there was no significant change in ethanol levels when blood samples with anticoagulant were frozen or refrigerated for several months and analyzed by volatilization of reducing substances in blood.4 Results Using Ortho Normal Control Serum, the run-to-run coefficient of variation (CV) for osmolality measurements was determined as 0.94% (276 ± 2.6 mOsm/kg, mean ± SD, N = 20) for the cryoscopic osmometer and 1.1% (276 ± 3.1 mOsm/kg, N = 20) for the vapor pressure osmometer. The instruments were cross-checked with calibrators of the other instrument. The mean of the differences between cryoscopic and vapor pressure measurements was 0.3 mOsm/kg (standard error of the mean = 0.42). The sensitivity of A osmolality measurement near the normal range, calculated as 2 SD of the difference between duplicate cryoscopic and vapor pressure determinations, was 3.9 mOsm/kg or 16 mg/dL of ethanol (N = 20). Two ethanol controls were made by spiking approximately 100 mg/dL and 300 mg/dL of ethanol to serum. The mean ± SD (N = 20) for the controls as determined by osmometry were 100 ± 8 mg/dL and 290 ± 8 mg/ dL, respectively. The run-to-run precision as 2 SD for the A osmolality measurement (N = 20) was 3.8 mOsm/ kg for both the levels. The serum osmolality for 51 random hospitalized patients on the cryoscopic osmometer ranged from 262 to 306, with a mean ± SD of 281 ± 9.6 mOsm/kg. The standard error of the mean (SE*) was 1.3 mOsm/kg. The serum osmolality on the vapor pressure osmometer was 280 ± 10.2 mOsm/kg (SE ; = 1.4). Using a Mest, there was no significant difference between freezing A.J.C.P. • December 1984 DRAVIAM, CUSTER, AND SCHOEN 708 400 n A Osmolality (mOsm)/Kg FIG. 1. Linear regression of A osmolality with standard solutions of ethanol in serum. For 61 patients with positive alcohol levels for the aca (sensitivity = 5 mg/dL), the difference in osmolality between cryoscopic and vapor pressure determinations was compared with the enzymatically assayed ethanol concentrations. The comparison (slope = 1.01; intercept = 7.49; Sj,x = 22.1; r = 0.9797) is shown in Figure 2. All the specimens were negative for acetoacetic acid (<5 mg/dL) and acetone (<50 mg/dL) with Acetest® reagent tablets (Ames Company, Elkhart, IN 46515). Five specimens with high levels of ethanol (>340 mg/dL) were analyzed for the ethanol, methanol, isopropanol, and acetone by gas chromatographic volatile screen. No other alcohol except ethanol was present. The ethanol concentrations by GC were, on the average, 9% higher than the results from osmometry and 13% higher than the aca results. Because enzymatic determinations using aca were linear up to only 300 mg/dL, these specimens were diluted with water. The discrepancy in the aca and GC results may be due to the error in dilution of the specimens for analysis by aca. One of the five specimens was found to contain 30 mg/dL of acetone. This could be equivalent to 5 mOsm/kg.3 point and vapor pressure measurements (P > 0.50). The Discussion A osmolality for these patients ranged from —4 to +6 In order to use osmolality measurements as a screening mOsm/kg. Three of the patients had a A osmolality of procedure, the clinically significant concentrations of 5 mOsm/kg and one patient 6 mOsm/kg, which were the volatile substance need to exceed the sensitivity of greater than the sensitivity (3.9 mOsm/kg) of the method. the A osmolality measurements. In addition to ethanol, This may be due to random error or due to the pressure isopropanol satisfies the above conditions. Although of endogenous volatiles. ethanol increases serum osmolality more than isoproA linear response was indicated when a serum pool panol on w/w basis, it is not possible to tell by osmolality spiked with ethanol corresponding to 0, 79, 158, 237, and 316 mg/dL were plotted against A osmolality (Fig. 1). The concentration of ethanol is obtained by linear regression as follows. 600 Ethanol (mg/dL) = 4.2 (A osm) - 0.04. (standard error y = 1.01 x + 7.49 Sv x = 21.13 of estimate, Sy)l = 2.51; correlation coefficient, r r = 0.9797 >500= 0.9998; N = 5) /) = 61 «to-» i• Dj 3,7 The slope of 4.2 is consistent with previous reports. o \CD With standards made by spiking ethanol to 0.9% NaCl, F fc 4 0 0 a concentrations up to 1,000 mg/dL were found to be O<>/ <>/ linear with A osmolality. About 8% decrease in A ox 300osmolality was observed when the standards were made *•— ID 3 o in saline as opposed to serum. However, this discrepancy raj <o 2 0 0 Sr can be eliminated if the A osmolalities of aqueous 0) r. standards are corrected for volume of water (91%) in the serum.14 100Similar concentrations of isopropanol and methanol in saline were found to be linear with measured osmo100 200 300 400 500 600 lality. Ethanol from Enzymatic Isopropanol (mg/dL) = 5.0 (A osm) - 6.2 (Syx = 6.37; Determinations (mg/dl) r = 0.9989; N = 5). Methanol (mg/dL) = 3.2 (A osm) + 1.6 (Sy.x = 1.50; FIG. 2. Linear regression of enzymatically determined ethanol and ethanol determined from A osmolality measurements. r = 0.9998; N = 5). — men rom • i_ JJS o Vol. 82 • No. 6 OSMOLALITY MEASUREMENTS APPLIED TO VOLATILE SCREEN measurement when a mixture of both alcohols is present. In the emergency room, most of the cases involve intoxication with high blood levels of ethanol. Very few cases of isopropanol intoxication are encountered. Because isopropanol is readily metabolized to acetone, the patients with isopropanol intoxication have significant levels of acetone in blood. The sensitivity of acetone by osmometry is 21 mg/dL (3.9 mOsm/kg). Elevation of A osmolality and a positive acetone may be a practical indicator of the presence of excess isopropanol or acetone (e.g., diabetic coma). In this case a quantitative volatile screen by GC may be necessary to positively identify the alcohols present. The lethal level of methanol (200 mg/dL) is high enough so that a significant A osmolality will be obtained with high levels of methanol. Toxic levels of methanol greater than 400 mg/dL have been reported.5 The sensitivity of osmolality measurements will indicate toxic levels of methanol as it does for ethanol. Ethylene glycol, which is toxic (lethal dose: 100 g), and propylene glycol, which is relatively nontoxic, may attain high blood levels. Propylene glycol has been demonstrated in high enough concentrations to be measurable in the sera of two severely burned patients.1 Additional studies are in progress concerning the use of osmometry in comparison with other methods for detecting nonvolatile alcohols and other volatile aldehydes and ketones of possible clinical significance. The A osmolality due to the presence of volatiles commonly has been determined by finding the difference between the measured osmolality and calculated osmolality using various formulas taken from the literature that are based on analyte (Na, K + , etc.) assays.2,313 These literature presentations do not reveal findings as reliable as those presented in this article. Analytic sensitivity and specificity and correlation coefficients are not as good as achieved in this study using osmolality determinations on the vapor pressure osmometer instead of the calculated osmolality from sodium, glucose, and urea-nitrogen concentrations. In this study no significant change in osmolality was observed by the Wescor instrument when up to 318 mg/dL of ethanol was added to control serum, saline, or water. Similarly, 157 mg/dL of methanol and 302 mg/dL of isopropanol did not change the osmolality measured by vapor pressure. The following procedure is presented for the determination of ethanol in serum. Osmolality measurements are made with cryomatic and vapor pressure osmometers, and the A osmolality is multiplied by a factor of 4.2. The result is reported as follows: Water-soluble volatiles are present equivalent to XXX mg/dL as ethanol (sen- 709 sitivity: ethanol = 16 mg/dL). If only isopropanol or methanol was known to be ingested, the respective linear regression factor could be used for the determination of concentration, and the result will be reported in terms of isopropanol or methanol (sensitivity: isopropanol = 14 mg/dL; methanol = 14 mg/dL). Water-soluble volatiles equivalent to or greater than 16 mg/dL as ethanol are assayed by a more specific enzymatic ethanol method. If the enzymatic result is less than the value obtained from A osmolality by more than 29 mg/dL (3 SD diff; SD2 diff = SD 2 enz + SD 2 A osm; where enz = enzymatic determination; A osm = A osmolality measurement), a volatile screen by gas chromatography is done. If less than 29 mg/dL the presence of significant amounts of (> 1 /6 of lethal level), methanol and isopropanol is ruled out. This volatile screen assay procedure is especially useful in measuring alcohol in the clinically significant range of 50 mg/dL to 600 mg/dL. References 1. Bekeris L, Baker C, Fenton J, Kimball D, Bermes E: Propylene glycol as a cause of an elevated serum osmolality. Am J Clin Pathol 1979; 72:633-636 2. Dorwart WV, Chalmers L: Comparison of methods for calculating serum osmolality from chemical concentrations, and the prognostic value of such calculations. Clin Chem 1975; 21:190-194 3. Glasser L, Sternglanz PD, Combie J, Robinson A: Serum osmolality and its applicability to drug overdose. Am J Clin Pathol 1973; 60:695-699 4. Glendening BL, Waugh TC: The stability of ordinary blood alcohol samples held various periods of time under different conditions. J Forensic Sci 1965: 10:192-200 5. Jacobson D, Bradesen JE, Eide I, Ostborg J: Anion and osmolar gaps in the diagnosis of methanol and ethylene glycol poisoning. Acta Med Scand 1982; 212:17-20 6. McCord WM, Switzer PK, Brill HM, Jr: Isopropyl alcohol intoxication. South Med J 1948; 41:639-642 7. Redetski HM, Koerner TA, Hughes JR, Smith AG: Osmometry in the evaluation of alcohol intoxication. Clin Toxicol 1972; 5:343-363 8. Robinson AG, Loeb JN: Ethanol ingestion Commonest cause of elevated plasma osmolality? N Engl J Med 1971; 284:12531255 9. Schoen I, Fischer C, Winter S, Barnett R: Patient preparation and specimen collection and handling including storage stability, Medical Usefulness Subcommittee of CAP Standards Committee, March 1974 10. Stern EL: Serum osmolality in cases of poisoning (Letter to the Editor). N Engl J Med 1974; 290:1026 11. Tietz NW: Fundamentals of of clinical chemistry. Second edition. Philadelphia, WB Saunders, 1976 12. Vasiliades J, Pollock J, Robinson CA: Pitfalls of the alcohol dehydrogenase procedure for the emergency assay of alcohol: A case study of isopropanol overdose. Clin Chem 1978; 24:383385 13. Weisberg HF: Osmolality-calculated, "delta," and more formulas. Clin Chem 1975;21:1182-1185 14. Wolf AV: Aqueous solutions and body fluids. New York: Hoeber Medical Division, Harper and Row, 1966
© Copyright 2026 Paperzz