Eight-parameter Automated Hematology Analyzers: Comparison of Two Flow Cytometric Systems BENJAMIN DREWINKO, M.D., PH.D., PAMELA BOLLINGER, MT(ASCP), MARY ROUNTREE, MT(ASCP), DENNIS JOHNSTON, PH.D., GAIL CORRIGAN, B. S., WILLARD T. DALTON, M.D., AND JOSE M. TRUJILLO, M.D. The performance of two high-speed 8-parameter automated hematology analyzers (Coulter Counter S-plus and ELT-8) were compared with that of reference instruments (Coulter Counter S Sr and Technicon Autocounter). The precision, linearity, and lack of carry-over of both instruments were superior over that of existing equipment. The especially noteworthy feature of the instruments was their excellent performance in the range of extreme values of both white blood cells and platelets. This enhanced performance, and the fact that all presently known relevant hematologic parameters can be measured by a single instrument on a single sample, make the contribution of these two flow cytometric-based instruments a significant advance in the field of automated laboratory medicine. (Key words: Automated analyzer, cell counts; Platelet count; Hematology; Flow cytometry) Am J Clin Pathol 1982; 78: 738-747 DEVELOPMENT OF INSTRUMENTS for highspeed analysis of cells in a fluid suspension (flow cytometry) has revolutionized the operation of the modern hematology laboratory. 3,5,81012 Improvements in design have yielded instruments that provide greater resolution, superior precision, enhanced accuracy, and a faster analysis rate than currently available automated hematology analyzers. 8,10 "' 14 Additionally, such instruments have the potential to provide increased information on hitherto unsuspected cellular parameters' 14 that may prove of vital and relevant clinical importance in the evaluation of the disease state. 1 , 3 " 1 2 This report provides a side-by-side evaluation of two currently available 8-parameter automated hematology analyzers that employ flow cytometric technology: the Coulter Counter® Model S-Plus, manufactured by Coulter Electronics, Inc. (Hialeah, FL), and the ELT-8,® manufactured by Ortho Instruments (Westwood, MA). Evaluation was conducted in a large oncologic hematology laboratory7 that provided ample opportunity to analyze the performance of such instruments under the usual conditions of either severe myelosuppression or of marked increments of circulating white blood cells and platelets. Received January 27, 1982; received revised manuscript and accepted for publication March 23, 1982. Address reprint requests to Dr. Drewinko: The University of Texas System Cancer Center, M. D. Anderson Hospital and Tumor Institute, Texas Medical Center, Houston, Texas 77030. The University of Texas System Cancer Center, M. D. Anderson Hospital and Tumor Institute, Departments of Laboratory Medicine and Biomathematics, Houston, Texas Materials and Methods Description of Instrumentation S-Plus. The Coulter® Model S-Plus (Coulter Electronics, Hialeah, FL) is an automated eight-parameter hematology analyzer that utilizes either 1 mL of whole blood or 44.7 ^L of capillary blood (diluted in 10 mL of isotonic diluent) to determine in 35-55 seconds the routine seven-parameter hematology count (white blood cell count, WBC; red blood cell count, RBC; hemoglobin concentration, HGB; hematocrit, HCT; mean corpuscular volume, MCV; mean corpuscular hemoglobin, MCH; and mean corpuscular hemoglobin concentration, MCHC), and also an eighth parameter, the platelet count (PLT). The instrument is also capable of analyzing four additional parameters: the RBC volume distribution width; the platelet-crit; the mean platelet volume; and the platelet volume distribution width. Whole blood and predilute samples are introduced into the instrument through separate aspiration pathways but analyzed in the same sensing chambers. The basic Coulter principle of nonoptical one-by-one electronic impedance counting and sizing of particles is utilized to enumerate WBC, RBC, and PLT; the MCV is determined directly from the electrical pulse heights. The hemoglobin concentration is measured by the colorimetric cyanmethemoglobin method. HCT, MCH, and MCHC are not measured directly, but are computed internally from the measured parameters. Different types of cells are discriminated by volumes in two separate chambers (WBC and RBC/PLT) each containing three aperture tubes. Pulses amplitudes generated from cells passing through the apertures are directly proportional to particle volumes and are calibrated to measure cell volumes from 0 to 360 fL. Under these conditions, a leukocyte is any particle with a volume greater than 45 fL that remains in the suspension after the RBCs have been lysed. All cells in the RBC/PLT bath with a 0002-9173/82/1100/0738 $01.30 © American Society of Clinical Pathologists 738 Vol. 78 • No. 5 EVALUATION OF AUTOMATED HEMATOLOGY ANALYZERS volume in the range of 36 to 360 fL are classified as RBCs, while those in the 2 to 20 fL range are classified as PLTs. The raw data of the PLT count from each aperture tube is distributed according to volume by differential gating in a channelyzer. It is then analyzed and smoothed by a log-normal distribution algorithm to cover the PLT range of 0 to 70 fL. If the PLT count is decreased, the instrument wilj continue counting for four additional count cycles until at least 400 particles have been detected and evaluated. If there still is an insufficient number of particles, or if there are statistically discrepant results among the distributions generated by each aperture tube, the result from the raw data is printed along with a code describing the anomaly. If the distribution is not log normally distributed, the result is suppressed and the operator is alerted by a visual signal. For all particles (WBC, RBC and PLT), Poisson coincidence errors in counting are corrected internally by appropriate algorithms. ELT-8. The ELT-8® (Ortho Instruments, Westwood, MA) is an automated hematology analyzer capable of measuring eight hematology parameters (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT) on a 100-microliter sample of whole blood in 60 seconds. The principle utilized by this instrument is that of optical detection of cells in a liquid suspension that scatter light at a low forward angle. The intensity of the scattered light is detected by a photodetector for conversion to electrical pulses: The technology involved is hydrodynamic focusing of a narrow stream of cells through a flow channel 250 j*m in diameter. Focusing is achieved by a liquid laminar sheath flow that is coaxial to the sample flow. This method isolates the cell suspension, confining the particles to the central portion of a liquid jet. A heliumneon laser beam is focused upon this stream of cells forming a minute, cylindrical sensing zone (20 microns in diameter and 7 microns in height) in the center of the sample stream, allowing for an almost single-file array of cells for sensing and detection. WBC are enumerated following lysis of the RBCs. Discrimination between PLT and RBC depends upon three cell variables: volume, refractive index, and "time of flight" through the sensing zone. Each cell generates a pulse whose height and width is proportional to all three discriminating parameters of that cell. A voltage value is assigned to that pulse and compared to an "integral" voltage. A value greater than the integral voltage is classified as an RBC, and a lower value is classified as a PLT: HGB concentration is measured by the colorimetric cyanmethemoglobin method. The HCT is determined from the area under the curve generated by the RBC pulses and the MCV, MCH, and MCHC are computed internally from the measured parameters. The operator is alerted to certain sample abnormalities or 739 instrument malfunctions by means of varied audible alarms or displayed sensor messages. Reagents and Controls S-Plus. The reagents used on the S-Plus were those recommended by the manufacturer. These included Isoton® II (Coulter Diagnostics, Hialeah, FL) a balanced electrolyte solution; Lyse S® II, an agent that lyses red blood cells, removes the cytoplasm from the white blood cells, and is the reactive agent for the HGB concentration determination; and Isoterge® II, a cleaning agent. Strict precautions were taken not to introduce bubbles into the Isoton II in order to prevent high background counts. Reagent consumption per test cycle was 55 mL of Isoton II, and 0.77 mL of Lyse S II. The primary control used during the evaluation was 4C® Plus normal (Coulter Electronics, Hialeah, FL), an eight-parameter control that included fixed human PLTs. Secondary controls for the standard seven parameters consisted of 4C® normal, abnormal low, and abnormal high (Coulter Electronics, Hialeah, FL), and Quanticel™ HP-Human Platelet Reference-normal and low (BHP, Inc., Westchester, PA) for the PLT parameter. ELT-8. The only reagents used during the evaluation were those currently approved for operation of the system and are manufactured by Ortho Instruments (Westwood, MA). These included Salac™, a buffered saline diluent; Lysac™, a red blood cell lysing agent and diluent for WBC; and Cyanac™, a diluent and reactive agent for determination of hemoglobin concentration. We found it advantageous to prefill extra reagent containers with Salac to let particulate matter settle out before installing on the instrument in order to avoid high background counts. It is also important to record the date on the Cyanac reagent container, as the reconstituted stability is 30 days. Reagent consumption per test cycle was 1.6 mL of Cyanac, 0.8 mL of Lysac, and 16.0 mL of Salac. Late during our evaluation, the Salac diluent was replaced with Isolac™ diluent (see Retrofits). The primary controls were CBC-trolR normal (Pfizer Diagnostics, New York, NY) for the seven routine hematology parameters, and Quanticel HP-Human Platelet Reference-normal for the PLT parameter. Secondary seven-parameter controls were Haem-CR-normal and abnormal low (J. T. Baker Diagnostics, Bethlehem, PA) and Quanticel -HP-Human Platelet Reference-low for the PLT parameter. Calibration Both the S-Plus and the ELT-8 were initially calibrated using blood collected in tripotassium EDTA (K3 EDTA) obtained from five hematologically normal donors. Determinations were performed on each whole 740 DREWINK0£7"/1L. blood sample utilizing the following reference methods for each parameter: (1) WBC and RBC: five replicate dilutions measured on a Coulter Counter® Model ZBI (Coulter Electronics, Hialeah, FL); (2) HGB: five replicate dilutions measured on a Gilford spectrophotometer Model 300-N; (3) HCT: five replicate aliquots centrifuged at 10,500 rpm for four minutes and read on a Damon/IEC Micro-Capillary Reader without correction for trapped plasma; and (4) PLT: eight phase microscopy readings of two Unopette® #5855 (Becton-Dickinson, Rutherford, NJ) dilutions (1:100 with 1% ammonium oxalate) with both sides of each chamber read by two technologists. Also at this time, all other automated hematology analyzers available in our clinical hematology laboratory were recalibrated, if necessary, to these reference values. These instruments consisted of three Coulter Counter® Model S Srs, and four AutoCounters™ (Technicon Instruments, Tarrytown, NY). For the purpose of the evaluation, the instruments in the clinical laboratory were used as reference. Quality, Control Before each daily operation, quality control measures were performed for each instrument according to the manufacturer's recommendation (for the test instruments) or to our established laboratory procedures (for the reference instruments). The instruments were considered to be in control when background counts were below recommended levels, and when values for the control materials fell within two standard deviations (SD) of the mean value established in our laboratory for 15 replicate assays over three consecutive days (for the S-Sr, S-Plus and AutoCounter), and 10 replicate assays processed immediately after calibration (for the ELT-8). S-Plus. Checks on instrument function and electronic components were performed daily prior to routine operation and background counts were done. The primary and secondary controls were then processed in duplicate. ELT-8. Quality control measures were performed daily on the ELT-8 in accordance with the manufacturer's recommendations. These included a check on the background count and the processing of replicate samples of the primary controls (or secondary calibrator materials) and the secondary control products. AutoCounter. The background was zeroed before each run. Platelet Reference N® (Technicon Instruments, Tarrytown, NY), a suspension of fixed human platelets at a normal level, was assayed as a standard at the beginning of each run and after every ten patient samples. Later in the study, this product was replaced with Quanticel HP-Human Platelet Reference-normal and low levels. S-Sr. Daily electronic and instrument function checks A.J.C.P. • November 1982 were performed on all three S-Srs according to our established laboratory protocol. Background counts were performed prior to the processing of the control materials. 4C normal was used as a primary control, and Haem-C normal and abnormal low were used as secondary controls. Precision Within-run precision was evaluated for the directly measured parameters, WBC, RBC, HGB, and PLT (and HCT for the ELT-8) at three levels: decreased, (below normal range), normal, and increased (above normal range). On the S-Sr and S-Plus, MCV precision was evaluated at the normal level only, due to the difficulty of obtaining adequate samples at abnormal levels. The SPlus, S-Sr, and AutoCounter were evaluated simultaneously in both the whole blood and predilute mode after making the appropriate dilutions from the same sample of whole blood. Large amounts of blood were required for each study and were obtained from normal donors free of any known hematologic disorders. The exception to this were the leukapheresis blood specimens used for the concentrated WBC studies obtained from two patients who had chronic, myelogenous leukemia. To obtain the values desired in the abnormal ranges, it was necessary to artificially dilute or concentrate the donor blood by separating it into its various components via several centrifugation steps, and then combining the proper volumes of packed RBCs, platelet rich plasma (PRP), and platelet poor plasma (PPP). Normal ranges were evaluated using the donor whole blood sample without additional manipulations. After preparing the blood specimens, 26 consecutive counts were performed on all instruments, and the first count was discarded for statistical analysis. Day-to-day precision of the S-Plus, S-Sr, and AutoCounter was evaluated for the directly measured parameters by calculating the coefficient of variation (CV) of the values obtained over a 30-day period using Counter Check® Plus (Diagnostic Technology, Inc. Great Neck, NY) for the S-Plus, 4C normal for the SSr, and Quanticel-HP normal for the AutoCounter. For the ELT-8, day-to-day precision was evaluated by calculating the CV of values obtained on CBC-trol normal and Quanticel-HP normal over a 20-day period. There were 58 separate values used for the WBC, RBC, HGB, HCT, and MCV parameters, and 48 were included in the calculations of the PLT parameter. Linearity Linearity was evaluated on eleven successive percentile dilutions (5 to 100%) of a given value of each pa- Vol. 78 • No. 5 EVALUATION OF AUTOMATED HEMATOLOGY ANALYZERS rameter at decreased, normal, and elevated levels in the whole blood mode (and in the predilute mode for the S-Sr, S-Plus, and AutoCounter)! Blood for each study was obtained from a single donor, separated into its components through several centrifugation steps, and combined in specific volumes to yield the appropriate dilutions. The only exception was the linearity study at an elevated WBC level, which was performed using a WBC concentrate from a patient with chronic myelogenous leukemia diluted with Isoton II (for the S-Sr, SPlus, and AutoCounter), and from a patient with lymphoma diluted with saline (for the ELT-8). Each dilution was processed in quadruplicate for each parameter, and the average value was plotted against the expected value from the dilution percentile. Carryover Two separate whole blood samples were used in the carry-over studies; one with high parameter values and a second with low parameter values. These samples were prepared utilizing normal donor components and the original samples were either concentrated or diluted to obtain the desired values. Predilute samples were prepared from the high and low specimens for the S-Sr, SPlus, and AutoCounter. A mean value for each counted parameter was obtained for both specimens by running ten replicates. The high and low value specimens were then run interspersed ten times, and these mean values were calculated. Per cent carry-over for each parameter was calculated from this formula: where: LA = x of alternated low value sample; Ls = x of straight replicates of low value sample; H s = x of straight replicates of high value sample; and HA = x of alternated high value sample. RBC Interference on PLT Enumeration Because the AutoCounter, S-Plus, and ELT-8 analyze platelets in the presence of intact or lysed RBCs, it was necessary to evaluate the effect of the RBC concentration on the PLT count. Assays at normal and low platelet levels were performed utilizing blood obtained from a single hematologically normal donor and separated into its various components by several centrifugation and filtration steps: platelet poor RBCs (4 X 103 platelets/ML), PRP, and PPP. Platelet free plasma (PFP) was produced by passing the PPP through a 0.2-fim filter. Eleven tubes were prepared for each trial by varying the amount of RBCs and PFP and adding the same amount of PRP to each tube. The tubes were weighed before and 741 after the addition of each component in order to verify the accuracy of pipetting. Each sample dilution was counted in triplicate; the least-squares fit method was used to plot the platelet count versus the RBC concentration. WBC Interference on HGB Determination To determine the effect of an increased WBC count on the colorimetric determination of HGB, and WBC concentrate was obtained during a therapeutic leukapheresis of a patient with acute lymphocytic leukemia. The concentrate was centrifuged and only the uppermost layer of leukocytes was aspirated to prevent contamination with RBCs. Type specific PPP and washed RBCs obtained from the Blood Bank were used to prepare eleven successive dilutions in which the amount of RBCs was constant and the amount of WBCs and PPP varied. Each dilution was anajyzed in quadruplicate on the SSr, S-Plus, ELT-8, and also on a Coulter® Hemoglobinometer (Coulter Electronics, Hialeah, FL) after centrifugation to remove the leukocytes. The hemoglor bin concentration versus the WBC count was plotted using the least-squares fit method. Retrofits A few inherent problems became apparent during the evaluation of both instruments. For the S-Plus, the RBC concentration artificially decreased the displayed PLT count by an average of 5.2% for every additional 106 RBCs//iL.6 In addition, the PLT counts obtained in the predilute mode were, on the average, 15% lower than those of the whole blood mode. In the case of the ELT8, a continuous downward trend was noticed for the MCV, HCT, and PLT parameters on the control materials. Neither the degree nor the occurrence of the downward trend could be predicted for a routine eighthour work period, let alone after overnight shut-down periods. Patient correlations were run whenever it appeared that a trend had occurred, and reference determinations were performed on all discrepancies. Results confirmed that the calibration of the ELT-8 was continuously drifting downward. Both manufacturers were alerted to take remedial action. For the RBC interference on the S-Plus, improvements consisted of software changes with the installation of a program revision to compensate electronically for the decrease in PLT count as the RBC concentration increased from 0 to 6 X 106/ML. TO correct for the discrepancy between PLT values determined by both aspiration modes, the changes included a new blood sampling valve (BSV), shortening of the bloodcarrying lines from the BSV to the RBC/PLT chamber, 742 Table I. Within-run Precision at Varying Levels on 25 Replicate Counts Reference Instruments Parameter WBC Level HGB HCT PLT Predilute Mode Whole Blood Mode Predilute Mode ELT-8 11.1* 9.2 4.9 2.7 1.8 1.2 1.2 1.6 12.1 J6.7 5.7 ND 3.0 2.7 2.5 2.6 6.3 6.2 3.6 1.9 1.2 1.1 1.8 5.8 5.6 6.8 1.3 2.0 2.1 2.7 2.2 5.3 NDf 3.2 2.0 2.3 1.1 0.9 1.5 1.2 0.9 0.5 '•4 1.8 1.0 1.3 0.7 0.6 0.8 0.7 0.6 1.2 1.5 0.7 1.3 0.8 0.7 0.7 0.8 0.8 0.7 0.6 0.6 0.4 1.8 ND 1.1 1.0 0.9' 0.7 0.4 0.4 0.5 0.5 1.6 2.2 0.8 0.7 0.9 1.0 2.1 1.4 0.4 0.5 ND ND ND ND ND ND ND ND ND ND ND ND 0.8 1.0 0.6 After retrofitting, correlations among results from all instruments were performed on random routine patient specimens processed by the Section of Hematology over an 18-month period. The S-Plus was compared with the S-Sr and the AutoCounter for 429 specimens in the whole blood mode, and for 205 specimens in the predilute mode. The ELT-8 was compared with the S-Plus, the S-Sr, and the AutoCounter for 373 samples. Results from each instrument, considered on a mode pair basis, were plotted as linear regression curves and correlation coefficients were calculated by the Pearson Method. For detailed analysis of the WBC and PLT counts, values were plotted separately at low, normal, and ejevated levels. A correlation analysis between results of specimens obtained from capillary blood (fingerstick) and those from venipuncture were conducted for 25 normal donors for the S-Plus and ELT-8, and also for 25 patient donors for the S-Plus only. Both venous and capillary samples were obtained from each donor, and predilute samples were prepared from the venous specimens for the S-Plus. Duplicate fingerstick specimens were collected utilizing the Hemette II® (Ortho Instruments, 0.9 0.9 0.6 0.6 0.5 0.4 0.8 0.5 ND ND Table 2. Day-to-day Precision at Normal Levels 7.9 7.4 2.6 3.0 1.4 1.9 3.2 2.5 11.9 7.8 5.2 3.3 2.2 2.4 2.2 2.4 7.4 5.0 3.0 2.0 3.1 2.2 2.2 2.1 8.4 6.6 3.4 3.3 2.2 2.7 2.4 1.8 9.0 ND 2.4 3.0 2.2 1.7 ND 1.5 i.3 (X106/ML) (g/dL) 6.3 9.2 13.0 14.8 18.9 (%) 26.9 34.6 43.1 MCV Whole Blood Mode (XIO /ML) 1.98 3.04 4.50 6.10 (fL) 87 94 and carry-over were evaluated again, and the values compared to those obtained before corrections. Correlations on Clinical Specimens S-Plus 3 0.9 1.2 1.8 5.0 8.0 11.0 25.0 43.0 RBC A.J.C.P. • November 1982 DREWINK.O£7"/lL. 3 (X10 A*L) 12 23 43 69 112 326 401 562 Parameters and an adjustment to the RBC diluent dispenser. These improvements brought the predilute PLT count to within 5% or less of the PLT count determined in the whole blood mode. For the ELT-8, the retrofit consisted of certain software changes in the Data Handler module as well as physical product improvements. The physical changes included a new reagent plumbing system, a smaller blocker bar, a V-2 pressure relief valve, and replacement of the photomultiplier tube with a solid state scatter detector. In addition, a new saline diluent, Isolac, that contained a surfactant expressly designed for optical systems replaced the original Salac. Upon retrofitting of both instruments, the calibration, precision, linearity, S-Plus ELT-8 8.9 0.2 2.7 6.9 0.2 2.4 7.2 0.2 2.6 5.00 0.06 1.2 4.63 0.04 0.9 4.52 0.03 0.7 WBC Mean ( 1 0 3 / M U SD CV RBC Mean (10 6 /ML) * Coefficient of variation (%). t ND = not done. Reference Instrument SD CV HGB Mean (g/dL) SD CV 15.0 0.2 1.1 13.1 0.1 0.9 13.7 0.1 0.9 HCT Mean (%) SD CV 41.7 0.7 1.7 40.8 0.4 1.0 42.2 0.4 1.0 1.3 0.8 0.9 93 0.8 0.9 145 15.0 4.4 265 5.9 2.2 355 7.3 2.1 MCV Mean (fl) SD CV PLT Mean(10 3 /ML) SD CV 83 EVALUATION OF AUTOMATED HEMATOLOGY ANALYZERS Vol. 78 • No. 5 743 WBC 100 r 20 40 60 80 0 2 4 6 8 10 3 Expected Value ( X 1 0 / M L ) FIG. 1. Relation between expected and observed values of WBC (at high and low ranges) measured by reference and test instruments. Westwood, MA) for use on the ELT-8 and the Unopette® #5925 (Becton-Dickinson, Rutherford, NJ) for use on the S-Plus. All specimens were analyzed in du- 2 4 6 8 6 Expected Value (x10 //xL) plicate within 4 hours. The capillary specimens collected for the ELT-8 were run in duplicate again after 24 hours. The mean values for venous, capillary, and predilute 4 8 12 16 Expected Value (g/dL) FIG. 2. Relation between expected and observed values of RBC and HGB measured by reference and test instruments. DREWINKO£r/lL. 744 A.J.C.P. • November 1982 PLT High Range 1000 - o 800 • 160 Low Range • S-Plua o ELT-8 • AutoCounter / * ji y t£ *_ co o - 120 /* ^ jio M So ~ 600 « 3 o Jf 80 « > •o 4 0 0 o > e / | > •D > Q> - 40 O 200 1 1 200 400 1 1 600 800 0 Expected Value 1 1 1 40 80 120 (0 a O 160 (X103/ML) FIG. 3. Relation of expected and observed values of PLT (at high and low ranges) measured by reference and test instruments. specimens were plotted as described above, and correlation coefficients were calculated. Results Within-run Precision Table l shows the within-run precision (expressed as CV) at varying levels for each of the directly derived parameters measured by all instruments. At normal levels, both the ELT-8 and the S-PLUS exhibit CVs lower than those of the manufacturer's stated specifications for every parameter. At abnormal levels, the instruments routinely display CVs below 1.5%. Noteworthy are improvements in the precision over that of reference instruments occurring at low concentrations of WBC man- ifested by a 50% decrease in CV. For the PLT parameter, the CV of all instruments were similar, but most important, they were less than 5% for PLT counts down to 50 X 103/>L, and less than 10% for counts below that value. For the S-Plus, a higher CV was exhibited by all parameters when measured in the predilute mode than in the whole blood mode. Day-to-day Precision The ELT-8 and the S-Plus exhibited comparably excellent day-to-day precision for all parameters (Table 2). The most noteworthy difference between these two instruments with respect to the reference instrument occurred for the PLT count. The day-to-day CV for the AutoCounter (4.4%) was twice that of both the ELT-8 (2.1%) and the S-Plus (2.2%). Table 3. Per Cent Carry-over for Direct Parameters Measured by Evaluation and Reference Instruments Reference Instruments S-Plus Mean Value of Specimen Parameter High WBC RBC HGB PLT 30 X 10 3 /ML 4.30 X 10 6 /ML * Per cent. 16 g/dL 950 X 10 3 /ML Low Whole Blood Mode Predilute Mode Whole Blood Mode Predilute Mode ELT-8 0.2 x 103AiL 1.2 X 1 0 % L 4.5 g/dL 5 X 107^L 1.67* 1.62 1.25 0.65 4.22 2.26 1.90 0.66 0.51 0.12 0.32 0.30 1.17 0.71 1.33 0.32 0.0 0.0 0.0 0.0 EVALUATION OF AUTOMATED HEMATOLOGY ANALYZERS Vol. 78 • No. 5 Linearity 745 450 - Figures 1, 2, and 3 display the linear relationship between expected and observed values of the directly derived parameters measured by all instruments. Of particular significance is the enhanced linearity achieved by both newflowinstruments at the low range of WBC and PLT counts rendering them of great value in the routine follow-up of myelosuppressed patients. 350 ^ 250 - 5. m-^ O Carry-over z o o No detectable carry-over was noted for all direct parameters measured by the ELT-8, and a negligible amount was observed for the S-Plus (Table 3). In this respect, it is interesting to note the marked improvement of the S-Plus over the S-Sr in regards to the carry-over in both the whole blood and predilute mode. 200 180 160 140 120 a LU >< _l 0. RBC Interference on PLT Count w a The RBC concentration of the blood sample had virtually no effect on the displayed PLT counts of the ELT8 and a neglible effect on the PLT counts of the AutoCounter. However, there was a notable effect on 50 45 40 35 30 25 • - S-PLUS ° - ELT-8 • -AUTOCOUNTER 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 6 RBC COUNT ( X 1 0 / M L ) FIG. 5. Effect of RBC concentration on displayed PLT count by the S-Plus after retrofit. Points represent the mean of values obtained from six separate S-Plus units. • S-Plus S-Sr o ELT-8 a Hemoglobinometer 10.4 c 3 O u 35 £ a e •o • 30 „• a> >, « a CD '••"it 0 25 0 o" 0 A , **• "-'^ 9.6 . CO 0 r I —•- 9.2 ' 1.0 1 1 2.0 3.0 RBC count (X ! 4.0 5.0 106/ML) FlG. 4. Effect of RBC concentration on displayed PLT count by three different analyzers. I e.o A _o p Z-*- o 6" O i a a 20 0 A 5" ^ - A (3 o • A >- 10.0 • •o • A o —o— a 20 40 60 80 • 100 3 WBC Count ( X 1 0 / M L ) FIG. 6. Effect of WBC concentration on displayed HGB measured by different instruments. DREWINKO ETAL. 746 Table 4. Correlation of S-Plus vs. Reference Instruments Parameter Whole Blood Mode Predilute Mode WBC RBC HGB HCT MCV PLT 0.99* 0.99 0.99 0.99 0.97 0.99 0.99 0.99 0.99 0.98 0.97 0.95 the counts were made in the whole blood or in the predilute mode. Upon retrofitting, this effect was slightly reversed to the extent that the PLT count increased an average of 3.5% for each additional 106 RBC/^L (Fig. 5). WBC Interference on HGB Determination Increments of WBC had no effect on HGB determinations performed by the Coulter Hemoglobinometer, and virtually no effect on the values determined by the ELT-8 (Fig. 6). For the S-Sr, displayed HGB values increased at the rate of 0.03 g/dL for each 5 X 103 WBC, and for the S-Plus, at the rate of 0.05 g/dL per 5 X 103 WBC. ' Pearson correlation coefficients. Table 5. Correlations among S-Plus, ELT-8, and Reference Instruments Parameter S-Plus vs. Reference Instruments ELT-8 vs. Reference Instruments S-Plus vs. ELT-8 WBC RBC HGB HCT MCV PLT 0.99* 0.99 0.99 0.98 0.97 0.98 0.99 0.99 0.99 0.98 0.92 0.98 0.99 0.99 0.98 0.98 0.95 0.98 * Pearson correlation coefficient. the PLT counts of the S-Plus (Fig. 4). The displayed PLT count on the S-Plus decreased an average of 5.2% for every additional 106 RBC/jtL, regardless of whether Table 6. Instrument Correlations of Separate Ranges S-Plus vs. Reference Instruments ELT-8 vs. Reference Instruments S-Plus vs. ELT-8 n n n WBC (XIO 3 /ML) Level <4.0 4.0-10.0 >10.0 0.98* 0.96 0.99 165 158 37 0.98 0.91 0.99 166 157 35 0.99 0.96 0.99 164 161 38 0.90 0.95 0.87 44 236 54 0.91 0.95 0.85 44 237 47 0.86 0.95 0.85 42 258 46 PLT (XIOVML) Level <50 50-400 >400 A.J.C.P. • November 1982 ' Pearson correlation coefficient. Retrofits No significant difference between values of precision, linearity, and carry-over of all directly measured parameters were noted for either instrument after retrofitting. For the S-Plus, corrections resulted in less interference of the RBC on the PLT count and good correlation of PLT counts between the two modes of aspiration. For the ELT-8, improvements gave a superior day-to-day precision of the MCV, HCT, and PLT parameters. Correlations on Clinical Specimens Table 4 presents the correlation coefficient of results for the S-Plus versus the reference instruments for 429 whole blood and 205 predilute specimens. Correlation values for all parameters were excellent with a slight decrease in correlation for PLT seen in the predilute mode (r = 0.95). Correlation of results obtained on the S-Plus, ELT-8, and reference instruments is shown in Table 5 for 373 clinical specimens. Again, all correlation values were extremely good between the instruments, except for a weaker correlation seen for the MCV parameter on the ELT-8 when measured against the S-Sr (r = 0.92). A detailed correlation analysis for the WBC and PLT parameters of the same 373 specimens at low, normal, and elevated levels is expressed in Table 6. Particularly significant is the excellent correlation achieved for WBC at both the low and elevated levels between all instruments; the correlation values for the PLT pa- Table 7. Correlation of Results between Different Collection and Aspiration Technic Patient Specimens Normal Donor Specimens S-Plus ELT-8 S-Plus Parameter Whole Blood vs. Capillary Whole Blood vs. Predilute Whole Blood vs. Capillary Whole Blood vs. Capillary 24 hr Whole Blood vs. Capillary Whole Blood vs. Predilute WBC RBC HGB MCV PLT 0.96* 0.98 0.97 0.97 0.96 0.99 0.97 0.98 0.98 0.97 0.98 0.99 0.99 0.99 0.99 0.99 0.99 0.99 0.99 0.99 0.99 0.95 0.94 0.99 0.99 0.98 0.96 0.95 0.98 0.96 • Pearson correlation coefficient. vol. 78 • No. 5 EVALUATION OF AUTOMATED HEMATOLOGY ANALYZERS rameter among all instruments decreased slightly at the low and elevated levels. Results of the correlation analysis between the mean of capillary specimens and whole blood specimens for both the S-Plus and ELT-8 is shown in Table 7. Excellent correlation values were obtained by both instruments between these two types of specimen collection technics, indicating that the method of obtaining the blood will not affect the accuracy of either instrument. The correlation values between the two aspiration modes on the S-Plus were also very good. Particularly noteworthy were the excellent correlation results obtained by the ELT-8 on the capillary specimens run 24 hours after having been drawn. Discussion The ever increasing demand by clinicians for an expanded range of hematology laboratory tests reported in a timely manner has been met by the development and implementation of sophisticated high-speed multiparameter blood analyzers. However, to be useful in routine practice, results provided by these instruments must be accurate and reproducible over the wide dynamic range of hematology values encountered in clinical practice. This is especially critical at the low range of both WBC and PLT counts where therapeutic decisions are based on relatively small changes of these parameters (i.e., transfusion, cessation, or resumption of chemotherapy, etc.). Also, cost efficiency and timeliness requires that the relevant parameters be analyzed with the least possible number of instruments and with the least number of sample preparative steps. These objectives have been achieved in both flow cytometric systems evaluated in this report, where a breakthrough improvement was achieved in the capability of counting PLTs along with the other seven parameters assayed in the now obsolescent multiphasic hematology analyzers. As determined in our laboratory, precision, Hnearity, and lack of carry-over of both the ELT-8 and S-Plus were superior to that of existing instruments, and slightly better than that reported by both manufacturers and by other investigators evaluating similar prototypes.1,1314 In particular, the precision of the PLT count was markedly superior to that performed manually or with other technology.4,615 This was especially noteworthy for values below 50 X 103/|tL where both precision arid correlation coefficients with existing methods were better than the results reported by other investigators.13 In a study reported elsewhere, we showed that even at levels of 20 X 103//iL, the accuracy of PLT counts determined by the S-Plus were confirmed by counts performed under phase microscopy.2 Thus, in contrast to the suggestions of others,113 it is our practice to report results from the automated count unless additional criteria established in our laboratory to prevent errors in PLT counts2 militate against this procedure. 747 For both instruments, results obtained on whole blood showed an excellent correlation with those determined on capillary blood specimens. This observation was remarkably useful in the case of the PLT count since previous reports4,9 had indicated a highly significant discrepancy between the two sources. The interference of the RBC concentration on the PLT count measured by the S-Plus previously reported by us6 has been corrected mathematically. Unfortunately, this correction has resulted in a slight bias in the opposite direction, creating a constant increase of about 3% more platelets per million RBC. However, this error is considerably less than that inherent to the precision (CV) at any level of the dynamic range of the PLT count as to make it clinically irrelevant. In summary, both flow cytometric hematology analyzers have performed equally well under extreme conditions of hematologic disease. Routine utilization of such instrumentation should have great impact on current large volume clinical hematology laboratory practice. Acknowledgement. Coulter Electronics, Inc. and Ortho Instruments provided instrumentation and reagents to perform this study. The assistance of James J. Conroy, Ortho Instruments is gratefully acknowledged. References 1. Bessman JD: Evaluation of automated whole-blood platelet counts and particle sizing. Am J Clin Pathol 1980; 74:157-162 2. Bollinger P: Accuracy of low platelet counts on the Coulter Counter Model S-Plus. Lab World 1981; 32:64-67 3. Brecher G: The future of automation in hematology. Blood Cells 1980;6:111-114 4. Brecher GB, Scheiderman M, Cronkite EP: The reproducibility and constancy of the platelet counts. Am J Clin Pathol 1953; 23:15-26 5. Brittin GA, Brecher G: Instrumentation and automation in clinical hematology. Progr Hematol 1971; 7:299-341 6. Dalton WT, Bollinger P, Drewinko B: A side-by-side evaluation of four platelet-counting instruments. Am J Clin Pathol 1980; 74:119-134 7. Drewinko B, Wallace B, Flores C, Crawford RW, Trujillo JM: Computerized hematology: Operation of a high volume hematology laboratory. Am J Clin Pathol 1977; 67:64-76 8. Fulwyler MJ: Flow cytometry and cell sorting. Blood Cells 1980; 6:173-184 9. Feusner JH, Behrens JA, Detter JC, Cullen TC: Platelet counts in capillary blood. Am J Clin Pathol 1979; 72:410-414 10. Groner W, Tycko D: Characterizing blood cells by biophysical measurements in flow. Blood cells 1980; 6:141-157 11. Haynes JL: High resolution particle analysis. Its application to platelet counting and suggestions for further application in blood cell analysis. Blood Cells 1980; 6:201-213 12. Kamentsky LA: Objective measurements of information from blood cells. Blood Cells 1980; 6:121 -140 13. Mayer K, Chin B, Magnes J, Thaler HT, Lotspeich C, Baisley A: Automated platelet counters. A comparative evaluation of latest instrumentation. Am J Clin Pathol 1980; 74:135-150 14. Rowan RM, Fraser C, Gray JH, McDonald GA: The Coulter Counter Model S-Plus. The shape of things to come. Clin Lab Haemat 1979; 1:29-40 15. Wertz RK, Keopke JA: A critical analysis of platelet counting methods. Am J Clin Pathol 1977; 68:195-201
© Copyright 2026 Paperzz