Center for Microscopy and Image Analysis Introduction to Electron Microscopy Andres Kaech Instrumentation Why electron microscopy? Resolution Limit Wavelength/Size Object 1 mm MRI, CT Radio Human eye 100 μm Cells Infrared 10 μm Red blood cells Visible Light microscope Ultraviolet 1 μm Bacteria 100 nm Mycoplasma n o r t c Ele py o c ro s c i m Viruses 10 nm Proteins x,γ-rays 1 nm Electron microscope 0.1 nm Amino acids Atoms Ultrastructure The types of electron microscopes Transmission electron microscope (TEM) Scanning electron microscope (SEM) Hela Cells 2 µm 1 µm The types of electron microscopes Transmission electron microscope (TEM) Electron beam Specimen Scanning electron microscope (SEM) Electron beam ~100 nm Specimen Projection Surface 2 µm 1 µm Why electron microscopy? Rat intestine Light microscopy Electron microscopy Microvilli 1 µm Adherence junction: ß-catenin visualized by immunolabelling using „immunogold“ 10 µm Green…F-actin Yellow…ß-catenin Orange…Nuclei Schwarz & Humbel 2007: Methods in Molecular Biology, vol. 369, Electron Microscopy: Methods and Protocols, Second Edition Edited by: J. Kuo © Humana Press Inc., Totowa, NJ Why electron microscopy? Mouse intestine Actin filaments Glycocalix Junction Microvilli 1 µm Elektronenmikroskopie ETH Zurich Specimens courtesy of Bärbel Stecher, Institute of Microbiology, ETH Zurich Why electron microscopy? Mouse intestine Membrane (lipid bilayer) Actin filaments 500 nm Elektronenmikroskopie ETH Zürich From photons to electrons Light microscopes Photons Photons are substituted with electrons Glass lenses are substituted with electromagnetic and electrostatic lenses Electron microscope Electrons e- From photons to electrons e- Similarities to photons: Wave-particle duality of electrons Optical properties (Diffraction, chromatic abberation, spherical abberation, astigmatism etc.) Resolution depends on aperture and wavelength (Diffraction limited resolution) Abbe’s equation d = 0.61 λ/NA NA = n ⋅ sin α Resolution of electron microscopes The higher the energy of the electrons, the lower the wavelength, the higher the resolution Acceleration voltages of electrons: Transmission electron microscopes (TEM): 40 – 1200 kV Scanning electron microscopes (SEM): 1 – 30 kV Effective instrument resolution TEM: ≈ 0.1 nm Effective instrument resolution SEM: ≈ 1 nm However: Resolution of biological objects is limited by specimen preparation: Practical resolution: > 1 nm Transmission electron microscope vs. Widefield light microscope 2 µm Transmission electron microscope Electron gun Electromagnetic lens TEM grid Widefield light microscope Illumination Condenser lens Specimen Lamp Glass lens Slide Electromagnetic lens Objective lens Glass lens Electromagnetic lens Projector lens Glass lens Phosphorescent screen CCD camera Final image Eye CCD camera Scanning electron microscope vs. Confocal scanning laser microscope 1 µm Scanning electron microscope Confocal scanning laser microscope Laser Illumination Electron gun Photomultiplier (Detector) Photomultiplier Detector Electromagnetic lenses Electromagnetic lens Electromagnetic/ electrostatic lens Lens system “condenser” Beam scanner Objective X-ray, photomultiplier Specimen Glass lenses Mirror Glass lenses Electron microscopes are high vacuum systems Transmission electron microscope (TEM) Scanning electron microscope (SEM) High vacuum Without vacuum: • Electrons would collide with gas molecules • Electron source (tungsten) would blow Vacuum systems Example: Transmission electron microscope Cathode 10-7 - 10-10 mbar IGP Ion getter pump / Oil diffusion pump Specimen holder (object plane) 10-5 - 10-7 mbar IGP Turbo molecular pump TMP 10-0 - 10-2 mbar Rotary pump Viewing screen RP Atmosphere: 1000 mbar Electron source (Electron gun) High voltage (acceleration voltage of e-) Electron source (Electron gun) Thermionic emission (tungsten, LaB6, Schottky emitter) Cold field emission (quantum-mechanical tunneling) Tungsten Thermionic LaB6 Schottky Cold field emission Material W LaB6 ZrO/W W Heating temp. (K) 2700 1800 1800 300 Normalized brightness Required vacuum (Pa) ∆E (eV) Ultra high high Chromatic aberration! Electromagnetic lenses Electromagnetic lens of a transmission electron microscope Electromagnetic lenses Magnetic field depends on current and number of windings e- e- Advantage of a electromagnetic lenses: The focal length can be changed by changing the current: I No movement or exchange of the lens is required for changing focus or magnification! ►Image is inverted and rotated ►Image rotation is corrected in microscopes Electromagnetic lenses Axial astigmatism of electromagnetic lenses … confusion of the image Most relevant aberration in biological electron microscopy x Objective lens “True” image of object Optical axis Object: Source of e(off-axis) z y Focus Reasons: • Inhomegenities of the lens • Contamination of lenses and apertures • Charging of specimen Electromagnetic lenses Axial astigmatism of electromagnetic lenses … confusion of the image Most relevant aberration in biological electron microscopy x “True” image of object Objective lens Optical axis Object: Source of e(off-axis) y Cellulose filter paper in SEM Under focused image elliptic deformation Focus circle of least confusion Over focused image elliptic deformation Electromagnetic lenses Axial astigmatism of electromagnetic lenses … confusion of the image Most relevant aberration in biological electron microscopy x Objective lens Corrector coils “True” image of object Optical axis Object: Source of e(off-axis) y Cellulose filter paper in SEM Focus, corrected astigmatism circle of confusion minimized Electromagnetic lenses Chromatic aberration Due to energy difference of electrons (wavelength) e- (98 kV) e- (100 kV) e- (102 kV) Curvature and distortion of field Spherical aberrations Specimen holders and stages 2 µm Transmission electron microscope Goniometer: x, y, z, r Specimen size: • 3 mm in diameter! • Ca. 100 nm in thickness (electron transparent) Specimen holders and stages 2 µm Specimen on a TEM grid 3 mm Specimen holder Specimen holders and stages 1 µm Scanning electron microscope Viewing chamber = Specimen chamber Gun Objective lens Specimen stub Stub holder Stage Specimen stage (x, y, z, r, tilt) Specimen size: • 100 mm in diameter • 2 cm in z-direction (not electron transparent) Electron – specimen interactions Primary electrons (E0) Backscattered electrons (E=E0) Elastic (higher angle, E=E0) Inelastic (low angle, E=E0-∆E) Unscattered (E=E0) Electron – specimen interactions Inelastic scattering: Primary electrons hit electrons of the specimen atom 2 Energy is transferred from the primary electron to the specimen K 1 L Emission of electrons and radiation M N Electron – specimen interactions Primary electrons Backscattered electrons X-rays Secondary electrons SEM analysis Cathode luminescense Heat Auger electrons Specimen Interaction volume TEM analysis Elastically scattered electrons Inelastically scattered electrons Unscattered electrons Imaging in the transmission electron microscope 2 µm Illumination Condenser lens Specimen Specimen: Electron transparent (very thin: 100 nm) Objective lens Projector lens Final image Image: 2D projection of a volume • CCD camera • Phosphorescent screen • Conventional photosensitive film Imaging in the transmission electron microscope 2 µm Contrast formation in TEM Absorption of electrons Scattering of electrons Phase contrast NOTE: All mechanisms occur at the same time (superposition) Question: Which mechanism is most relevant for biological specimens? Imaging in the transmission electron microscope 2 µm Contrast formation in TEM Biological specimen consist of light elements: Absorption contrast weak Scattering contrast weak “LOW CONTRAST” Phase contrast weak Contrast enhancement required: Treatment with heavy metals (Ur, Pb, Os)! Heavy metals attach differently to different components Imaging in the transmission electron microscope 2 µm Main contrast formation in plastic embedded specimens Scattering of electrons through heavy metals Primary electron beam …Heavy metal ions phospholipids Specimen ribosome Objective aperture Signal Intensity Specimen profile Imaging in the transmission electron microscope 2 µm Thin section of alga stained with heavy metals (Ur, Pb) Electron microscopy ETH Zurich Imaging in the transmission electron microscope 2 µm Thin section of alga without heavy metal staining 1 µm Electron microscopy ETH Zurich Imaging in the transmission electron microscope 2 µm Contrast enhancement by underfocusing Phase differences of diffracted and non-diffracted rays are increased or decreased by changing the focus Underfocus Close to focus Overfocus Fresnel rings Carbon film , ca. 4 nm 100 nm Imaging in the transmission electron microscope 2 µm Contrast enhancement by underfocusing Thin section of a frozen-hydrated apple leaf (“unstained”) 1 µm Electron microscopy ETH Zurich Phase contrast only (H2O vs. biological material) Imaging in the scanning electron microscope 1 µm Scanning electron microscope Illumination Photomultiplier Detector • Photomultiplier • No CCD camera Lens system “condenser” Beam scanner Objective Specimen Specimen: Bulk specimen Imaging in the scanning electron microscope 1 µm Scanning and signal detection Scanning of the specimen Imaging in the scanning electron microscope 1 µm Signal and detection Primary electrons Backscattered electrons X-rays Secondary electrons Heat Auger electrons Specimen SEM analysis Cathode luminescense Interaction volume TEM analysis Elastically scattered electrons Inelastically scattered electrons Unscattered electrons Imaging in the scanning electron microscope 1 µm Signal and detection Different properties of the different signals ► Specific detectors ► Different/specific information Imaging in the scanning electron microscope 1 µm Secondary electron detector Primary electrons +200-500V – Collector voltage +7-12kV HV SE Electrons Photons Photomultiplier Electrons Imaging in the scanning electron microscope 1 µm Contrast formation in SEM (using SE and BSE) Different number of electrons from different spots of the specimen Dependent on location of the detector topography of the specimen acceleration voltage of primary electrons composition of the specimen Imaging in the scanning electron microscope 1 µm Contrast based on SE Virtual light source SE detector (inlens) SE detector Leg of an ant, coated with ca. 10 nm Platinum Electron microscopy ETH Zurich Imaging in the scanning electron microscope 1 µm Contrast based on SE PE PE SE R PE PE PE PE Φ SE SE R R R R Specimen R Primary electrons Secondary electrons Excited volume Imaging in the scanning electron microscope 1 µm Contrast formation in SEM Biological material (light elements): Only few electrons escape from specimen Almost no contrast, similar contrast everywhere on specimen Unsharp image (electrons from “large” volume) Contrast enhancement important & needed: Localization of the signal to the surface Coating of biological specimen with thin heavy metal layer (a few nm) Reducing acceleration voltage Imaging in the scanning electron microscope 1 µm Contrast based on SE: Non-coating vs. coating with heavy metals Uncoated Coated with 4 nm platinum 500 nm Freeze-fractured yeast Electron microscopy ETH Zurich Imaging in the scanning electron microscope 1 µm Contrast formation Uncoated Coated with 4 nm platinum Primary electron beam Primary electron beam Platinum
© Copyright 2026 Paperzz