无图英文版小论文

Establishment of a visual and rapid detection method for influenza A
virus based on Au@Pt nanozyme
Yinzhen Zhaoa, Yuran Wangb, Fang Lic, Min Lic, Panpan Shid, Ruiting Gaoe, Hongzhi Lif,
Xiyu Shangf, Yunlong Wanga,g,*
a
Luoyang Vocational and Technical College, Luoyang 471099, Henan, China
b
Henan Collaborative Innovation Center of Molecular Diagnosis and Laboratory Medicine, Henan Medical
University, Xinxiang 453003, Henan, China
c
Zhengzhou Ninth People's Hospital, Zhengzhou 450000, Henan, China
d
Zhengzhou Technical College, Zhengzhou 450010, Henan, China
e
Henan General Hospital, Zhengzhou 450002, Henan, China
f
Zhengzhou Sixth People's Hospital, Zhengzhou 450000, Henan, China
g
Henan Bioengineering Technology Research Center, Zhengzhou 450000, Henan, China
* Corresponding author. E-mail address: [email protected]
Highlights
• Au@Pt nanozyme-based visual assay for influenza A virus completes within 30 min
• Detection limit of 0.20 ng/mL, approximately 4-fold lower than a colloidal gold strip
• No cross-reactivity with influenza B virus or SARS-CoV-2 nucleoproteins
•
Intra- and inter-assay CVs below 10% after 37 °C accelerated storage
Abstract
We report a visual assay for influenza A virus (IAV) that couples the peroxidase-mimicking catalytic
color reaction of gold-platinum nanoparticles (Au@Pt NPs) with magnetic bead-based target
enrichment. Au@Pt NPs were synthesized by chemical reduction, and a double-antibody sandwich
format was established in which magnetic beads capture the viral nucleoprotein and Au@Pt
NP-labeled antibodies generate the signal. The resulting magnetic bead-antigen-Au@Pt complex
catalyzes the oxidation of 3,3',5,5'-tetramethylbenzidine (TMB) by hydrogen peroxide, yielding a
characteristic blue-green product. The limit of detection was 0.20 ng/mL, approximately four times
lower than that of a commercial colloidal gold strip (0.78 ng/mL). Intra- and inter-assay coefficients
of variation were both below 10.00% (n = 10). The assay recognized influenza A virus nucleoprotein
only and showed no cross-reactivity with influenza B virus or SARS-CoV-2 nucleoproteins, and its
performance was unchanged after accelerated storage at 37 °C for 5 days (P > 0.05). Because the
procedure takes about 30 min and the result can be read by eye, this platform provides a practical
route to point-of-care testing for influenza A.
Keywords: Au@Pt nanoparticles; carboxylated magnetic beads; influenza A virus; immunoassay
1 Introduction
Influenza is an acute respiratory infection caused by single-stranded, negative-sense RNA viruses of
the family Orthomyxoviridae, whose causative agents are characterized by high genetic variability
and efficient transmission [1,2]. The World Health Organization (WHO) estimates that seasonal
influenza accounts for about one billion infections, three to five million cases of severe illness and
290,000–650,000 deaths worldwide each year [3,4]. Influenza A virus (IAV) is the main epidemic
subtype: antigenic drift in the surface glycoproteins hemagglutinin (HA) and neuraminidase (NA)
1
allows the virus to escape pre-existing immunity, while antigenic shift and reassortment can generate
variants with pandemic potential [5-7]. Within two months of its recognition, the 2009 pandemic
H1N1 virus had been reported in more than 70 countries, illustrating how quickly such a variant can
spread [8]. Reassortment between avian and mammalian viruses can also breach species barriers
[9,10], and the continuing zoonotic transmission of avian influenza viruses therefore remains a serious
public-health concern [11].
In China, influenza-like illness (ILI) accounts for approximately 2–8% of outpatient and emergency
department visits, and influenza A viruses predominate among laboratory-confirmed cases, although
both proportions vary with season and year [12]. Early treatment with neuraminidase inhibitors such
as oseltamivir reduces the risk of pneumonia, prolonged hospitalization and death in hospitalized
patients [13], which underlines the clinical value of a rapid diagnosis.
Here we combine the signal-amplifying color reaction of gold-platinum nanoparticles (Au@Pt NPs)
with the efficient magnetic separation provided by carboxylated magnetic beads to build a sensitive,
visual antigen assay for IAV. The platform requires no specialized instrumentation and is therefore
suited to point-of-care testing (POCT) in primary care and resource-limited settings.
2 Materials and methods
2.1 Materials
2.1.1 Main reagents
Monoclonal antibodies against the IAV nucleoprotein (clones 5C4-4E5, 5C4-4F5, 5F4-4G5 and
5F4-4G6) were provided by the Henan Bioengineering Technology Research Center. Carboxylated
magnetic
beads
(2.8
μm)
were
purchased
from
Thermo
Fisher
Scientific.
2-(N-Morpholino)ethanesulfonic acid (MES), 1-(3-dimethylaminopropyl)-3-ethylcarbodiimide
hydrochloride (EDC) and N-hydroxysuccinimide (NHS) were from Shanghai Aladdin Biochemical
Technology Co., Ltd. Chloroauric acid (HAuCl₄), sodium chloroplatinate hexahydrate
(Na₂PtCl₆·6H₂O), trisodium citrate dihydrate (Na₃C₆H₅O₇·2H₂O), potassium carbonate (K₂CO₃) and
PEG-20000 were from Sinopharm Chemical Reagent Co., Ltd. Ascorbic acid (AA) was from
Sigma-Aldrich (Shanghai) Trading Co., Ltd. A commercial influenza A/B antigen detection kit (InTec
Products, Xiamen) was used for comparison.
2.1.2 Reference materials
Influenza A virus nucleoprotein served as the sensitivity reference, and influenza B virus
nucleoprotein and SARS-CoV-2 nucleocapsid protein as specificity references.
2.1.3 Instruments
Particle size was measured with a laser particle size analyzer (Malvern Panalytical). Absorbance was
read on a TU-1810 spectrophotometer (Beijing Purkinje General Instrument Co., Ltd.) and on a
full-wavelength microplate reader (Thermo Fisher Scientific). Heating and stirring were performed
with a DHT-type heating mantle (Shandong Juancheng Hualu Electric Heating Instrument Co., Ltd.).
Other equipment included a centrifuge (Hunan Xiangyi Laboratory Instrument Development Co.,
Ltd.), a magnetic separation rack (Anhui Baisha Biotechnology Co., Ltd.), a vortex mixer (Changzhou
Zhongjie Laboratory Instrument Manufacturing Co., Ltd.), a rotary mixer (Shanghai Damu Industrial
Co., Ltd.) and a Milli-Q water purification system (Merck).
2.2 Preparation and characterization of Au@Pt nanoparticles
2
2.2.1 Preparation of colloidal gold seeds
Colloidal gold seeds were prepared by heating 100 mL of 0.01% (w/v) HAuCl₄ solution to boiling,
adding 0.5, 1.0 or 2.0 mL of 1% trisodium citrate solution and boiling for a further 12 min. The
suspension was cooled to room temperature and made up to the original volume with water.
2.2.2 Preparation and characterization of Au@Pt core–shell nanoparticles (Au@Pt NPs)
Water (30 mL) was placed in a three-necked flask, and 41, 82 or 123 μL of 10% sodium
chloroplatinate solution (Au:Pt molar ratios of 1:1, 1:2 and 1:3, respectively) was added; the mixture
was stirred magnetically for 5 min. The colloidal gold seed solution (30 mL) was then added and
stirring was continued for another 5 min. The mixture was heated to 100 °C under vigorous stirring
and 15, 30 or 45 μL of 1 mol/L AA (Pt:AA molar ratios of 1:1, 1:2 and 1:3, respectively) was added
rapidly. After 15 min the suspension was cooled to room temperature, adjusted to 30 mL with
ultrapure water and characterized by dynamic light scattering (DLS).
2.3 Preparation of Au@Pt–antibody conjugates and immunomagnetic beads
2.3.1 Preparation of Au@Pt labels
Au@Pt NPs (300 μL) were mixed thoroughly with an equal volume of 1 mmol/L K₂CO₃ solution, and
400 μL of 0.01 mg/mL anti-IAV monoclonal antibody 5C4-4E5 was added. After incubation at room
temperature for 20 min, 10 μL each of 10% BSA and 10% PEG-20000 was added and incubation was
continued for a further 20 min to block unreacted sites and improve stability. The conjugate was
collected by centrifugation (10,000 rpm, 10 min) and resuspended in 400 μL of storage buffer.
2.3.2 Preparation of immunomagnetic beads
Carboxylated magnetic beads (1 mg) were transferred to a 1.5 mL tube and washed twice with 500 μL
of 25 mmol/L MES buffer (pH 5.0) under magnetic separation. The beads were activated with 100 μL
of freshly prepared EDC/NHS solution (50 mg/mL each) for 30 min in the dark with rotation, washed
twice with MES buffer and resuspended in 500 μL of the same buffer. Anti-IAV monoclonal antibody
5F4-4G5 (30 μL, 1 mg/mL) was added and the beads were rotated for 1 h. Unbound antibody was
removed magnetically, and the beads were blocked with 500 μL of 50 mmol/L Tris-HCl (pH 7.4) for
15 min and then twice with 500 μL of PBS containing 1% BSA and 1% Tween-20 for 2 h each. The
immunomagnetic beads were finally resuspended in 1 mL of PBS containing 1% BSA and stored at 4
°C.
2.4 Establishment and optimization of the visual detection method
2.4.1 Establishment of the detection method
Immunomagnetic beads (50 μL) were mixed with 1 mL of sample in a 1.5 mL tube and incubated for
10 min with rotation. After magnetic separation and three washes with PBS containing Tween-20, 50
μL of Au@Pt-labeled antibody was added, the mixture was incubated for 10 min and the beads were
separated and washed again. Chromogenic substrate (200 μL) was added; positive samples developed
a distinct blue-green color, whereas negative samples retained the original color of the beads. The
reaction was stopped with 100 μL of stop solution and the absorbance was read at 450 and 630 nm
(A₄₅₀ and A₆₃₀) on a microplate reader.
2.4.2 Screening of antibody pairs
Four anti-nucleoprotein monoclonal antibodies (5C4-4E5, 5C4-4F5, 5F4-4G5 and 5F4-4G6) were
screened in a 4 × 4 orthogonal matrix in which each antibody served both as the capture reagent on
3
magnetic beads and as the Au@Pt-labeled detector. Pairs were ranked by the ratio of the absorbance
of a positive sample to that of a negative control (P/N).
2.4.3 Screening of Au@Pt NP particle size
With the optimal pair fixed, Au@Pt NPs of different sizes were conjugated under identical conditions
and the P/N ratios of the resulting labels were compared as described above.
2.4.4 Optimization of Au@Pt NP labeling conditions
The pH of the labeling reaction was adjusted with 100, 200, 300 or 400 μL of 1 mmol/L K₂CO₃
solution, and the optimal volume was identified from the P/N ratio. With this volume fixed, the
amount of antibody was varied (1, 2, 3 or 4 μg) and optimized in the same way.
2.4.5 Optimization of carboxylated magnetic bead coupling conditions
Coupling was carried out in 25 mmol/L MES buffer at pH 4.0, 5.0 or 6.0, and the optimal pH was
selected from the P/N ratio. The amount of antibody used for coupling was then varied (10, 20, 30, 40
or 50 μg) and optimized in the same way.
2.4.6 Optimization of incubation time and reagent volumes
With all other parameters fixed, the first incubation was tested at 5, 10 and 15 min while the second
was held at 10 min; the second incubation was then varied over the same range with the first fixed at
10 min, and the combination giving the highest P/N ratio was adopted. The volumes of
immunomagnetic beads and Au@Pt label (20, 50 and 80 μL) were then examined in a checkerboard
design, and the optimal combination was again selected on the basis of the P/N ratio.
2.4.7 Optimization of washing cycles
The number of washing cycles after the second incubation (0, 1, 2, 3 or 4) was evaluated with ten
replicates per condition, and the optimal number was selected from both the reproducibility (CV) and
the P/N ratio (Section 3.8).
2.5 Statistical analysis
Data are presented as mean ± standard deviation (SD). Group means were compared by the t-test,
with P < 0.05 considered statistically significant. Analyses were performed in SPSS 28.0 (IBM Corp.,
Armonk, NY, USA).
3 Results
3.1 Characterization of Au@Pt NPs
Dynamic light scattering (Fig. 1) showed that, for a given seed, the diameter of the Au@Pt NPs
increased as more platinum precursor was added, and that, for a given amount of precursor, the final
diameter increased with the size of the gold seed.
Fig. 1. Size distribution characteristics of Au@Pt NPs synthesized with precursor at gradient concentrations. A,
B and C are Au@Pt NPs prepared using 22.58 nm colloidal gold as seed, with Au:Pt molar ratios of 1:1, 1:2 and
1:3, respectively. D, E and F are Au@Pt NPs prepared using 42.17 nm colloidal gold as seed, with Au:Pt molar
ratios of 1:1, 1:2 and 1:3, respectively. G, H and I are Au@Pt NPs prepared using 66.21 nm colloidal gold as
seed, with Au:Pt molar ratios of 1:1, 1:2 and 1:3, respectively.
4
3.2 Screening of paired antibodies
As shown in Fig. 2, beads coated with 5F4-4G5 and Au@Pt NPs labeled with 5C4-4E5 gave the
lowest background for negative samples and the strongest response for positive samples, and hence
the highest P/N ratio. This pair was used in all subsequent experiments.
Fig. 2. Results of matched monoclonal antibody pair screening for influenza A virus. (A) Correlation between
the amount of antibody conjugated to carboxyl magnetic beads, the amount of antibody labeled with Au@Pt
NPs, and the absorbance value of the negative sample (N). (B) Correlation between the amount of antibody
conjugated to carboxyl magnetic beads, the amount of antibody labeled with Au@Pt NPs, and the absorbance
value of the positive sample (P). (C) Correlation between the amounts of antibody conjugated to carboxyl
magnetic beads and antibody labeled with Au@Pt NPs, and the positive-to-negative ratio (P/N).
3.3 Screening of Au@Pt NP particle sizes
The P/N ratio decreased steadily as the platinum content increased (Fig. 3). The best discrimination
was obtained with Au@Pt NPs of 31.10 nm grown on 22.58 nm gold seeds at an Au:Pt ratio of 1:1,
which combined a low background with the strongest positive signal.
Fig. 3. Particle size distribution of various Au@Pt nanoparticles.
3.4 Optimization of labeling conditions for Au@Pt NPs
K₂CO₃ adjusts the surface charge and colloidal stability of Au@Pt NPs and therefore affects antibody
conjugation. Between 100 and 400 μL of K₂CO₃ the absorbance of negative samples was essentially
unchanged, whereas that of positive samples rose and then fell, the P/N ratio being maximal at 200 μL
(Fig. 4A). The absorbance of both positive and negative samples increased with the amount of
antibody, but the P/N ratio peaked at 3 μg (Fig. 4B). Labelling was therefore performed with 200 μL
of K₂CO₃ and 3 μg of antibody.
Fig. 4. Optimization of Au@Pt NP labeling conditions. (A) Relationship between the amount of K₂CO₃ and the
absorbance values of positive (P) and negative (N) samples labeled with Au@Pt NPs, and the P/N ratio. (B)
Relationship between the amount of antibody and the absorbance values of positive (P) and negative (N)
samples labeled with Au@Pt NPs, and the P/N ratio.
3.5 Optimization of coupling conditions for carboxylated magnetic beads
The highest P/N ratio was obtained at pH 4.0 (Fig. 5A), consistent with more efficient EDC activation
of carboxyl groups in weakly acidic media. The positive signal increased and then decreased as the
amount of antibody used for coupling increased, whereas the negative signal changed little; 20 μg per
mg of beads gave the best discrimination (Fig. 5B).
5
Fig. 5. Optimization of carboxyl magnetic bead–antibody coupling conditions. (A) Absorbance values of
samples at coupling buffers with different pH levels. (B) Absorbance values of samples with various amounts of
antibody coupled to carboxyl magnetic beads.
3.6 Optimization of incubation time
The absorbance of positive samples increased with incubation time but leveled off after 10 min,
indicating that the reaction had approached equilibrium (Fig. 6); the P/N ratio was also maximal at 10
min. Both incubation steps were therefore set to 10 min.
Fig. 6. Optimization of antigen–antibody incubation times during sample detection. (A) Absorbance values at
different incubation times for carboxyl magnetic bead-conjugated antibody and sample antigen. (B) Absorbance
values at different incubation times for Au@Pt NP-labeled antibody and sample antigen.
3.7 Optimization of the amounts of immunomagnetic beads and Au@Pt labels added
Increasing the amounts of immunomagnetic beads and Au@Pt label raised both P and N, but the P/N
ratio was maximal when 50 μL of each was used (Fig. 7); this combination was adopted as optimal.
Fig. 7. Volume optimization of carboxyl magnetic bead-conjugated antibody and Au@Pt NP-labeled antibody.
3.8 Optimization of washing cycles
The absorbance of negative samples fell as the number of washes increased, indicating progressive
removal of non-specific material (Fig. 8). Excessive washing, however, also weakened the positive
signal, particularly beyond three washes, whereas in the unwashed group residual free label lowered
the positive signal and made it less reproducible.
On the basis of the P/N ratio, one or two washes removed most of the background without loss of
sensitivity, and a single wash gave the best overall performance and was the most convenient for
routine use. The optimal number of washes may need to be re-established when the assay is
transferred to an automated platform.
Fig. 8. Optimization of washing times.
3.9 Evaluation of detection performance
No color developed in the blank, whereas the signal increased with the concentration of IAV
nucleoprotein (Fig. 9A). A visible color difference was still discernible at 0.20 ng/mL, which was
therefore taken as the limit of detection (LOD). The assay recognized IAV nucleoprotein but not
influenza B virus or SARS-CoV-2 nucleocapsid protein (Fig. 9B). Under the same conditions the
commercial colloidal gold strip gave an LOD of 0.78 ng/mL, so the new method was about four times
more sensitive (Fig. 9C).
6
Fig. 9. Performance evaluation of the carboxyl magnetic bead-conjugated antibody and Au@Pt NP-labeled
antibody assay. (A) Results of the lowest detection limit; wells 1–12 correspond to 0, 100, 50, 25, 12.5, 6.25,
3.13, 1.56, 0.78, 0.39, 0.20 and 0.10 ng/mL, respectively. (B) Results of specificity detection; 1: negative
control sample (PBST); 2: seasonal influenza A virus nucleoprotein (H1N1); 3: influenza B virus nucleoprotein;
4: SARS-CoV-2 nucleocapsid protein. (C) Analytical sensitivity of the colloidal gold test strip; wells 1–9
correspond to 0, 50, 25, 12.5, 6.25, 3.13, 1.56, 0.78 and 0.39 ng/mL, respectively.
Table 1 summarizes the precision data. For the negative sample and for positive samples at 10, 100
and 1000 ng/mL, the intra- and inter-assay CVs were all below 10%, which meets the usual
acceptance criterion for immunoassays.
Accelerated stability was assessed over six consecutive days (day 0 to day 5) at 37 °C using the
negative sample and positive samples at 10 and 100 ng/mL. Results were compared with day 0 by the
t-test, and P > 0.05 was taken to indicate the absence of a significant change.
No concentration group changed significantly during the observation period (Tables 2–4). The P
values were 0.332 for the 10 ng/mL sample and 0.826 for the 100 ng/mL sample, and the negative
control showed neither abnormal color development nor non-specific reactions, which further
supports the stability and interference resistance of the system.
Overall, the reagent was stable under accelerated conditions. Extending the observation period and
testing clinical specimens would be the logical next step.
Table 1.
Precision test results of the assay.
Sample
concentration
(ng/mL)
Mean
SD
CV (%)
1000
1.23
0.09
7.04
100
0.90
0.06
10
0.21
0
0.04
Within-run precision
Between-run precision
Mean
SD
1.32
0.11
8.31
6.76
0.87
0.06
6.36
0.02
8.65
0.20
0.02
8.09
0.003
6.64
0.04
0.003
7.50
CV (%)
CV, coefficient of variation; SD, standard deviation.
Table 2.
Accelerated stability of the negative sample.
Day
Replicate 1
Replicate 2
Replicate 3
Replicate 4
Mean
Result
0
0.02660
0.02340
0.02860
0.02660
0.0263
Negative
1
0.02730
0.02520
0.02940
0.02630
0.02705
Negative
2
0.02740
0.03040
0.02880
0.02670
0.028325
Negative
3
0.03140
0.03070
0.02780
0.03220
0.030525
Negative
4
0.04040
0.04240
0.04380
0.04670
0.043325
Negative
5
0.04790
0.04110
0.04270
0.04280
0.043625
Negative
P (t-test)
Table 3.
0.332
Accelerated stability of the 10 ng/mL positive sample.
Day
Replicate 1
Replicate 2
Replicate 3
Replicate 4
0
0.3520
0.3364
0.2952
0.3300
0.3284
7.3
1
0.3452
0.3242
0.3004
0.3195
0.322325
5.71
2
0.3352
0.3242
0.3104
0.3095
0.31983
3.83
3
0.3153
0.3044
0.3235
0.2957
0.309725
3.94
4
0.2931
0.2845
0.3352
0.3075
0.305075
7.28
7
Mean
CV (%)
5
0.32780
0.3011
0.28390
0.29920
P (t-test)
Table 4.
0.303
6.02
0.332
Accelerated stability of the 100 ng/mL positive sample.
Day
Replicate 1
Replicate 2
Replicate 3
Replicate 4
Mean
CV (%)
0
1.178
1.265
1.1731
1.362
1.244525
7.15
1
1.197
1.126
1.1362
1.335
1.19855
8.03
2
1.187
1.216
1.1362
1.265
1.20105
4.49
3
1.1555
1.1309
1.1627
1.2597
1.1772
4.81
4
1.1535
1.3095
1.0276
1.1975
1.172025
9.94
5
1.1841
1.2372
1.1794
1.1547
1.18885
2.92
P (t-test)
0.826
CV, coefficient of variation.
4 Discussion
Bimetallic core-shell Au@Pt nanoparticles with a platinum monolayer shell combine high
enzyme-like catalytic activity with a low platinum content and are widely used to catalyze the
oxidation of chromogenic substrates such as TMB in biosensing [14,15]. Higareda et al. [16] showed
that the electrocatalytic performance of Au@Pt nanoparticles depends on the Au/Pt atomic ratio,
which provides a rationale for tuning activity through composition. Guided by this principle, we
synthesized Au@Pt NPs by seed-mediated growth and controlled the shell through the ratio of seed
size to platinum precursor. The preparation grown on 22.58 nm gold seeds at an Au:Pt ratio of 1:1
(31.10 nm) gave the highest peroxidase-like activity in our assay, and the systematic comparison of
seed size and composition reported here should assist the design of low-platinum catalytic labels.
Quantitative RT-PCR detects IAV with a sensitivity of the order of 10³ copies/mL [17], but it requires
dedicated instrumentation, trained staff and careful sample processing, which limits its use in primary
care [18,19]. Colloidal gold lateral-flow strips are simple and give an immediate visual result, but
their sensitivity is insufficient for samples with low viral loads [20,21]. The present method addresses
both limitations: the Au@Pt label amplifies the signal from small amounts of antigen through efficient
catalysis, while the magnetic beads capture the virus specifically and remove matrix components.
Together these features improve both the sensitivity and the specificity of the assay.
The assay is complete within 30 min and the result can be read by eye, so neither specialized
equipment nor a trained operator is required. This makes the method suitable for primary care,
resource-limited settings and on-site screening, and the format is compatible with automated liquid
handling for higher throughput. These features are relevant both to public-health surveillance and to
the control of seasonal and emerging influenza.
5 Conclusions
We have established a visual, rapid and sensitive assay for IAV that combines the peroxidase-like
activity of Au@Pt nanozymes with magnetic enrichment. The LOD is 0.20 ng/mL, specificity is high,
the CV is below 10%, and performance is retained after accelerated storage. The assay takes about 30
min, the result is visible to the naked eye and the same reaction can be quantified in a microplate
reader, which makes the method suitable for point-of-care use in primary care and resource-limited
settings.
8
Author contributions
Yinzhen Zhao: Conceptualization; Investigation; Methodology; Writing – original draft.
Yuran Wang: Conceptualization; Data curation; Investigation; Writing – original draft.
Fang Li: Data curation; Investigation; Methodology; Supervision.
Min Li: Data curation; Investigation; Methodology.
Panpan Shi: Data curation; Investigation; Methodology.
Ruiting Gao: Data curation; Methodology; Resources.
Hongzhi Li: Data curation; Methodology; Resources.
Xiyu Shang: Data curation; Methodology; Resources.
Yunlong Wang: Conceptualization; Supervision; Funding acquisition; Writing – review & editing.
Declaration of competing interests
The authors declare that they have no known competing financial interests or personal relationships
that could have appeared to influence the work reported in this paper.
Funding
This work was supported by the Henan Provincial Science and Technology Research Project (Grant
No. 242102310263).
Data availability
The data supporting the findings of this study are available from the corresponding author upon
reasonable request.
References
[1] Vemula SV, Zhao J, Liu J, Wang X, Biswas S, Hewlett I. Current approaches for diagnosis of influenza
virus infections in humans. Viruses 2016;8(4):96. https://doi.org/10.3390/v8040096.
[2] Carnell GW, Ferrara F, Grehan K, Thompson CP, Temperton NJ. Pseudotype-based neutralization assays for
influenza: a systematic analysis. Front Immunol 2015;6:161. https://doi.org/10.3389/fimmu.2015.00161.
[3]
World
Health
Organization.
Influenza
(seasonal).
Fact
sheet,
2025.
https://www.who.int/news-room/fact-sheets/detail/influenza-(seasonal) (accessed 28 February 2025).
[4] Yong J, Lu S, Lu C, Huang R. The development history, structural composition, and functions of influenza
viruses and the progress of influenza virus inhibitors in clinics and clinical trials. Mini Rev Med Chem
2025;25(3):196-207. https://doi.org/10.2174/0113895575316416240724043949.
[5] Kim H, Webster RG, Webby RJ. Influenza virus: dealing with a drifting and shifting pathogen. Viral
Immunol 2018;31(2):174-83. https://doi.org/10.1089/vim.2017.0141.
[6] Chen J, Wang J, Zhang J, et al. Advances in development and application of influenza vaccines. Front
Immunol 2021;12:711997. https://doi.org/10.3389/fimmu.2021.711997.
[7] Bedford T, Riley S, Barr IG, et al. Global circulation patterns of seasonal influenza viruses vary with
antigenic drift. Nature 2015;523(7559):217-20. https://doi.org/10.1038/nature14460.
[8] Dawood FS, Jain S, Finelli L, et al. Emergence of a novel swine-origin influenza A (H1N1) virus in humans.
N Engl J Med 2009;360(25):2605-15. https://doi.org/10.1056/NEJMoa0903810.
9
[9] Reid AH, Fanning TG, Janczewski TA, et al. Characterization of the 1918 "Spanish" influenza virus
neuraminidase
gene.
Proc
Natl
Acad
Sci
USA
2000;97(12):6785-90.
https://doi.org/10.1073/pnas.100140097.
[10] Kumar B, Asha K, Khanna M, et al. The emerging influenza virus threat: status and new prospects for its
therapy and control. Arch Virol 2018;163(4):831-44. https://doi.org/10.1007/s00705-018-3708-y.
[11] Kang M, Wang LF, Sun BW, et al. Zoonotic infections by avian influenza virus: changing global
epidemiology,
investigation,
and
control.
Lancet
Infect
Dis
2024;24(8):e522-31.
https://doi.org/10.1016/S1473-3099(24)00234-2.
[12] Yu H, Alonso WJ, Feng L, et al. Characterization of regional influenza seasonality patterns in China and
implications for vaccination strategies: spatio-temporal modeling of surveillance data. PLoS Med
2013;10(11):e1001552. https://doi.org/10.1371/journal.pmed.1001552.
[13] Koul PA, Mir H, Shah TH, et al. Combination therapy of nitazoxanide with oseltamivir compared with
oseltamivir in hospitalized patients with seasonal influenza. Lung India 2024;41(1):55-9.
https://doi.org/10.4103/lungindia.lungindia_711_21.
[14] Gao Z, Xu M, Lu M, et al. Urchin-like (gold core)@(platinum shell) nanohybrids: a highly efficient
peroxidase-mimetic system for in situ amplified colorimetric immunoassay. Biosens Bioelectron
2015;70:194-201. https://doi.org/10.1016/j.bios.2015.03.039.
[15] Kim DM, Yoo SM. Colorimetric systems for the detection of bacterial contamination: strategy and
applications. Biosensors 2022;12(7):532. https://doi.org/10.3390/bios12070532.
[16] Higareda A, Kumar-Krishnan S, Garcia-Ruiz AF, et al. Synthesis of Au@Pt core-shell nanoparticles as
efficient
electrocatalyst
for
methanol
electro-oxidation.
Nanomaterials
2019;9(11):1644.
https://doi.org/10.3390/nano9111644.
[17] van Elden LJR, Nijhuis M, Schipper P, Schuurman R, van Loon AM. Simultaneous detection of influenza
viruses A and B using real-time quantitative PCR. J Clin Microbiol 2001;39(1):196-200.
https://doi.org/10.1128/JCM.39.1.196-200.2001.
[18] Merckx J, Wali R, Schiller I, et al. Diagnostic accuracy of novel and traditional rapid tests for influenza
infection compared with reverse transcriptase polymerase chain reaction: a systematic review and
meta-analysis. Ann Intern Med 2017;167(6):394-409. https://doi.org/10.7326/M17-0848.
[19] Espy MJ, Uhl JR, Sloan LM, et al. Real-time PCR in clinical microbiology: applications for routine
laboratory
testing.
Clin
Microbiol
Rev
2006;19(1):165-256.
https://doi.org/10.1128/CMR.19.1.165-256.2006.
[20] Posthuma-Trumpie GA, Korf J, van Amerongen A. Lateral flow (immuno)assay: its strengths, weaknesses,
opportunities and threats. A literature survey. Anal Bioanal Chem 2009;393(2):569-82.
https://doi.org/10.1007/s00216-008-2287-2.
[21] Hsiao WWW, Fadhilah G, Lee CC, et al. Nanomaterial-based biosensors for avian influenza virus: a new
way forward. Talanta 2023;265:124892. https://doi.org/10.1016/j.talanta.2023.124892.
10