OP-FTIR: Product Description and Specifications

OP-FTIR: Open Path Spectrometer for Air Quality Monitoring
PRODUCT DESCRIPTION
The OP-FTIR is a low-cost, compact,
ruggedized, Open-Path Fourier Transform
Infrared (OP-FTIR) spectrometer system. The
system is designed to work in the informationrich 7 to 14 micron region with 8 cm-1 spectral
resolution. This resolution is specifically
tailored for identifying and quantifying heavy
industrial solvent molecules with comparable
selectivity and better sensitivity than a highresolution (Δσ<1 cm-1) spectrophotometer.
The new OP-FTIR system combines low
manufacturing cost with high sensitivity in a
rugged, fieldable instrument. The sensitivity
increase offered by our low spectral resolution
design permits the use of an uncooled
pyroelectric detector in place of the more
traditional cryogenically-cooled mercury
cadmium telluride detector.
Algorithms
tailored for use with low resolution spectra
maintain system specificity without sacrificing
overall sensitivity.
System bulk and
complexity are further reduced through the
use of a novel encoder-based metrology
system that replaces the Helium-Neon laser
interferometer used in most FTIR systems
The
system includes
algorithms
to
automatically compute concentration x path
length products for multiple molecular species
from the measured spectra, allowing for
autonomous, stand-alone operation. A low
cost plastic press-molded retroreflector array
for use in place of the typical hollow glass
corner cube retroreflector array is available as
an option.
OPTRA is a large scale provider of chemical
agent detection sensors for the military
marketplace and a small scale provider of
innovative environmental solutions.
The OP-FTIR Spectrometer Module
ADVANTAGES
•
Broad range of target gasses
•
Long Open Path Range: 100m
•
High signal-to-noise
•
Uncooled detector
•
Light weight: <7kg including telescope
•
On-board black-body calibration
•
Discrimination algorithms included
•
Full technical support; customization
available
APPLICATIONS
•
Industrial monitoring
•
Air quality monitoring
•
Threat detection
•
HVAC monitoring
•
In situ process monitoring
OP-FTIR
TECHNICAL SPECIFICATIONS
The standard OP-FTIR Spectrometer Module includes a 150mm telescope and a modulator
assembly which includes the interferometer, detector, IR source and reference system. Optional
accessories include a Portable PC and a Retroreflector Array, which can be built to customer’s
specifications.
Functional Specifications
SPECIFICATION
VALUE
Spectral Range
714 to 1428 cm-1
Spectral Resolution
8 cm-1
Integration time (single scan)
1.25 s
Open path range
1 to 100 m
Operational Specifications
SPECIFICATION
VALUE
Temperature range
20 to 30 ºC
Spectral Range
Physical Specifications
800
SPECIFICATION
Size: Spectrometer module
PC
VALUE
26.7 x 21.6 x 18.4 cm
40.9 x 28.2 x 23.1 cm
Retroreflector array
per customer spec
Weight: Spectrometer module
6.35 kg
700
600
500
400
300
200
100
PC
12.7 kg
0
600
800
1000
1200
1400
1600
1800
2000
optical frequency (cm-1)
CO2 Laser Spectrum
Optional Portable PC and Retroreflector Array
900
910
920
930
940
950
960
970
980
990 1000
optical frequency (cm-1)
All specifications are subject to change without
notice.
OPTRA, Inc.
461 Boston St. Topsfield, MA 01983
978-887-6600 FAX: 978-887-0022
[email protected] or http://:www.optra.com
OP-FTIR.doc
10/25/04