Phonak Audéo V-312 (V90/V70/V50/V30) (xS)

Technical Data
Phonak AudГ©o V-312 (V90/V70/V50/V30) (xS)
Receiver-In-Canal (RIC) instrument with size 312 battery
(for fitting range, product details and available options, please see
the Product Information or visit www.phonakpro.com)
RIC instruments can be fitted with a Standard, Power or SuperPower
plus external receiver (xReceiver). The Standard xReceiver (xS) is for
mild to moderately severe hearing loss. Unless otherwise specified,
all data obtained are measured in a closed configuration with a
coupling disc onto a HA-1 coupler (ANSI-S3.7-1995) or an occluded
ear simulator (EN 60711, coupling arrangement according to fig. 4 in
the test standard), and in the Phonak Target measurement settings.
Using pure tone measurements with a digital hearing instrument can
result in a wavy frequency response. This is an artifact resulting from
the use of a narrowband input signal and does not affect the actual
performance with naturally occurring broadband input signals.
2cm3 coupler data
ANSI S3.22-2009
Output sound pressure level
Nominal
111 dB SPL
Ear simulator data
dB SPL
130
EN / IEC 60118 and IEC 60711
120
Maximum
114 dB SPL
HFA
106 dB SPL
Full-on gain
(Input 90 dB SPL)
110
Output sound pressure level
Maximum
122 dB SPL
100
90
1600 Hz
113 dB SPL
80
100
dB SPL
130
Full-on gain
(Input 90 dB SPL)
120
10000
Acoustic gain
110
dB
60
100
1000
Maximum
46 dB
HFA
37 dB
RTG
29 dB
Full-on gain
(Input 50 dB SPL)
90
50
80
100
1000
Reference test gain
(Input 60 dB SPL)
40
10000
Acoustic gain
30
20
Maximum
56 dB
1600 Hz
45 dB
RTG
38 dB
10
100
dB
60
Full-on gain
(Input 50 dB SPL)
1000
10000
Frequency range
<100 Hz - 8800 Hz
Total harmonic distortion
500 Hz
1.5%
Battery current
Quiescent Working
1.1 mA
1.2 mA
Equivalent input noise level
19 dB SPL
50
Reference test gain
(Input 60 dB SPL)
40
30
20
10
1000
10000
<100 Hz - 9200 Hz
Total harmonic distortion
500 Hz
1.5%
Battery current
Quiescent Working
1.1 mA
1.2 mA
Equivalent input noise level
19 dB SPL
800 Hz
2%
1600 Hz
2.5%
Dynamic data
Compression
Attack time
Recovery time
10 ms
50 ms
Input / Output characteristics at 2000 Hz
Dynamic data
Compression
1600 Hz
2%
Attack time
Recovery time
10 ms
50 ms
P out dB
SPL
110
Full-on gain
100
Reference test gain
90
80
70
60
30
40
50
60
70
80
90
100
P in dB SPL
V1.00/2014-06/DHO В© Phonak AG All rights reserved/page 1/3
100
Frequency range
800 Hz
2%
Technical Data
Phonak AudГ©o V-312 (V90/V70/V50/V30) (xP)
The Power xReceiver (xP) is for mild to severe hearing loss.
Ear simulator data
2cm3 coupler data
EN / IEC 60118 and IEC 60711
ANSI S3.22-2009
Output sound pressure level
Output sound pressure level
Maximum
132 dB SPL
1600 Hz
131 dB SPL
Nominal
124 dB SPL
dB SPL
140
Full-on gain
(Input 90 dB SPL)
130
HFA
120 dB SPL
Full-on gain
(Input 90 dB SPL)
130
120
120
110
110
100
100
90
Maximum
127 dB SPL
dB SPL
140
90
100
1000
10000
100
Acoustic gain
Maximum
65 dB
1000
10000
Acoustic gain
1600 Hz
62 dB
RTG
53 dB
dB
70
Full-on gain
(Input 50 dB SPL)
60
Reference test gain
(Input 60 dB SPL)
50
dB
70
Maximum
57 dB
HFA
52 dB
RTG
43 dB
Full-on gain
(Input 50 dB SPL)
60
Reference test gain
(Input 60 dB SPL)
50
40
40
30
30
20
20
1000
100
10000
Frequency range
<100 Hz - 6400 Hz
Total harmonic distortion
500 Hz
1.5%
Battery current
Equivalent input noise level
10000
<100 Hz - 6600 Hz
Total harmonic distortion
500 Hz
1%
Quiescent Working
1.1 mA
1.2 mA
Battery current
Quiescent Working
1.1 mA
1.2 mA
19 dB SPL
Equivalent input noise level
19 dB SPL
800 Hz
1.5%
1600 Hz
1.5%
Dynamic data
Compression
1000
Frequency range
800 Hz
1%
1600 Hz
1%
Dynamic data
Attack time
Recovery time
10 ms
50 ms
Compression
Attack time
Recovery time
10 ms
50 ms
Input / Output characteristics at 2000 Hz
P out dB
SPL
130
Full-on gain
120
Reference test gain
110
100
90
80
30
40
50
60
70
80
90
100
P in dB SPL
V1.00/2014-06/DHO В© Phonak AG All rights reserved/page 2/3
100
Technical Data
Phonak AudГ©o V-312 (V90/V70/V50/V30) (xSP plus)
The SuperPower plus xReceiver (xSP plus) is for moderate to severe
hearing loss.
Ear simulator data
2cm3 coupler data
EN / IEC 60118 and IEC 60711
ANSI S3.22-2009
Output sound pressure level
Output sound pressure level
Maximum
136 dB SPL
1600 Hz
136 dB SPL
dB SPL
140
Full-on gain
(Input 90 dB SPL)
130
dB SPL
140
Nominal
128 dB SPL
Maximum
131 dB SPL
HFA
124 dB SPL
Full-on gain
(Input 90 dB SPL)
130
120
120
110
110
100
100
90
90
100
1000
100
10000
Acoustic gain
Maximum
72 dB
1000
10000
Acoustic gain
1600 Hz
70 dB
RTG
61 dB
Maximum
63 dB
dB
80
Full-on gain
(Input 50 dB SPL)
70
Reference test gain
(Input 60 dB SPL)
60
HFA
58 dB
RTG
47 dB
dB
80
Full-on gain
(Input 50 dB SPL)
70
Reference test gain
(Input 60 dB SPL)
60
50
50
40
40
30
30
1000
100
10000
Frequency range
<100 Hz - 5500 Hz
Total harmonic distortion
500 Hz
2%
Battery current
Equivalent input noise level
10000
<100 Hz - 6000 Hz
Total harmonic distortion
500 Hz
1%
Quiescent Working
1.1 mA
1.2 mA
Battery current
Quiescent Working
1.1 mA
1.2 mA
19 dB SPL
Equivalent input noise level
19 dB SPL
800 Hz
1.5%
1600 Hz
1%
Dynamic data
Compression
1000
Frequency range
800 Hz
1.5%
1600 Hz
0.5%
Dynamic data
Attack time
Recovery time
10 ms
50 ms
Compression
Attack time
Recovery time
10 ms
50 ms
Input / Output characteristics at 2000 Hz
P out dB
SPL
130
Full-on gain
120
Reference test gain
110
100
90
80
30
40
50
60
70
80
90
100
P in dB SPL
V1.00/2014-06/DHO В© Phonak AG All rights reserved/page 3/3